Abstract/OtherAbstract:
PURPOSE: The aim of this study was to evaluate the appropriateness of using computed tomography (CT) to cone-beam CT (CBCT) deformable image registration (DIR) for the application of calculating the "dose of the day" received by a head and neck patient.
METHODS: NiftyReg is an open-source registration package implemented in our institution. The affine registration uses a Block Matching-based approach, while the deformable registration is a GPU implementation of the popular B-spline Free Form Deformation algorithm. Two independent tests were performed to assess the suitability of our registrations methodology for "dose of the day" calculations in a deformed CT. A geometric evaluation was performed to assess the ability of the DIR method to map identical structures between the CT and CBCT datasets. Features delineated in the planning CT were warped and compared with features manually drawn on the CBCT. The authors computed the dice similarity coefficient (DSC), distance transformation, and centre of mass distance between features. A dosimetric evaluation was performed to evaluate the clinical significance of the registrations errors in the application proposed and to identify the limitations of the approximations used. Dose calculations for the same intensity-modulated radiation therapy plan on the deformed CT and replan CT were compared. Dose distributions were compared in terms of dose differences (DD), gamma analysis, target coverage, and dose volume histograms (DVHs). Doses calculated in a rigidly aligned CT and directly in an extended CBCT were also evaluated.
RESULTS: A mean value of 0.850 in DSC was achieved in overlap between manually delineated and warped features, with the distance between surfaces being less than 2 mm on over 90% of the pixels. Deformable registration was clearly superior to rigid registration in mapping identical structures between the two datasets. The dose recalculated in the deformed CT is a good match to the dose calculated on a replan CT. The DD is smaller than 2% of the prescribed dose on 90% of the body's voxels and it passes a 2% and 2 mm gamma-test on over 95% of the voxels. Target coverage similarity was assessed in terms of the 95%-isodose volumes. A mean value of 0.962 was obtained for the DSC, while the distance between surfaces is less than 2 mm in 95.4% of the pixels. The method proposed provided adequate dose estimation, closer to the gold standard than the other two approaches. Differences in DVH curves were mainly due to differences in the OARs definition (manual vs warped) and not due to differences in dose estimation (dose calculated in replan CT vs dose calculated in deformed CT).
CONCLUSIONS: Deforming a planning CT to match a daily CBCT provides the tools needed for the calculation of the "dose of the day" without the need to acquire a new CT. The initial clinical application of our method will be weekly offline calculations of the "dose of the day," and use this information to inform adaptive radiotherapy (ART). The work here presented is a first step into a full implementation of a "dose-driven" online ART.
Authors:
Catarina Veiga; Jamie McClelland; Syed Moinuddin; Ana Lourenço; Kate Ricketts; James Annkah; Marc Modat; Sébastien Ourselin; Derek D'Souza; Gary Royle
Related Documents :
24824977 - Evaluation of the extent of ground-glass opacity on high-resolution ct in patients with...
15690407 - Evidence based medicine in inborn errors of metabolism: is there any and how to find it.
24351257 - Cerebellar lesions in children with tuberous sclerosis complex.
24147467 - Esophagography after pneumomediastinum without ct findings of esophageal perforation: i...
23450817 - Acceleration apportionment: a method of improved 2d sense acceleration applied to 3d co...
25459197 - Diagnostic confidence and image quality of ct pulmonary angiography at 100 kvp in overw...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Medical physics     Volume:  41     ISSN:  0094-2405     ISO Abbreviation:  Med Phys     Publication Date:  2014 Mar 
Date Detail:
Created Date:  2014-03-05     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0425746     Medline TA:  Med Phys     Country:  United States    
Other Details:
Languages:  eng     Pagination:  031703     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Suitability of markerless EPID tracking for tumor position verification in gated radiotherapy.
Next Document:  Three-dimensional analysis of the respiratory interplay effect in helical tomotherapy: Baseline vari...