Automatic Sensitivity Control, unique to St. Jude defibrillators, allows for programmable sensitivity threshold and refractory periods. However, inappropriate programming can lead to marked undersensing and potentially lethal complications. We report a case of complete failure to sense ventricular fibrillation with inappropriate sensitivity adjustments. Detailed review of the parameters with induction of ventricular fibrillation must be performed to assure adequate safety.
Authors:
David Lin; Sanjay Dixit; Andrea M Russo; Henry H Hsia
Related Documents :
7777417 - Optimizing defibrillation through improved waveforms.
1382277 - Failure of a second and third generation implantable cardioverter defibrillator to sens...
17184687 - Bi-ventricular circulatory support with the abiomed ab5000 system in a patient with idi...
2703617 - A prospective, randomized evaluation of effect of ventricular fibrillation duration on ...
8193737 - Changes in autonomic tone following thrombolytic therapy for acute myocardial infarctio...
16352177 - Retroaortic course of the anomalous left main coronary artery: is it a benign anomaly? ...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  27     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  2004 Sep 
Date Detail:
Created Date:  2004-10-05     Completed Date:  2005-02-15     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1321-3     Citation Subset:  IM    
Affiliation:
Electrophysiology Service, Cardiovascular Division, University of Pennsylvania Health System, Philadelphia, Pennsylvania 19104, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Defibrillators, Implantable*
Equipment Design
Equipment Failure
Female
Humans
Ventricular Fibrillation / diagnosis*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Inappropriate implantable cardioverter defibrillator discharge following consumption of a dietary we...
Next Document:  Recurrent syncope triggered by inappropriate sinus tachycardia.