Time lag between oscillatory pressure and flow affecting accuracy of forced oscillation technique.
Jump to Full Text
MedLine Citation:
PMID:  21798069     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The forced oscillation technique (FOT) is a simple method for assessing the oscillatory mechanics of the respiratory system. The oscillatory properties, respiratory system resistance (Rrs) and reactance (Xrs), are calculated from the oscillatory pressure/flow relationship. Although the FOT has been a well-established technique, some detailed experimental conditions would be different among institutions.
METHODS: We evaluated whether time lags produced by the experimental conditions such as different positions of the sensors can affect the accuracy of the FOT. If the position of the pressure sensor is different from the flow sensor, a time lag may occur in the measurements. The effect of the time lag was studied by numerical analysis.
RESULTS: Rrs was estimated to be increased and Xrs decreased with an increase in the time lag, especially at a high oscillatory frequency of the medium-frequency range (5-35 Hz). At the high-frequency range (10-500 Hz), Rrs and Xrs were strikingly different in the values of the time lag.
CONCLUSION: A time lag between the oscillatory pressure and flow may be involved in the accuracy of the FOT, suggesting that it needs to be minimized or compensated for with signal processing. Researchers should pay attention to such detailed experimental conditions of the FOT apparatus.
Authors:
Junichi Ohishi; Hajime Kurosawa
Related Documents :
18448389 - Electroencephalography burst suppression in a patient with dengue encephalopathy: a cas...
21974179 - Statistical analysis of a characteristic shock formation distance for high-amplitude no...
7664559 - Dopamine-1-receptor stimulation and mucosal tissue oxygenation in the porcine jejunum.
17591059 - Comparison of the effects on cardiopulmonary function of ultrathin transnasal versus no...
15164369 - Morphometric analysis of the larval branchial chamber in the dragonfly aeshna cyanea m...
11243309 - Lead-induced hypertension. iii. increased hydroxyl radical production.
20404729 - Positive end-expiratory pressure improves end-expiratory lung volume but not oxygenatio...
3281649 - Assisted positive pressure breathing for augmentation of acceleration tolerance time.
24401029 - Intraocular pressure homeostasis: maintaining balance in a high-pressure environment.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2011-07-29
Journal Detail:
Title:  Biomedical engineering online     Volume:  10     ISSN:  1475-925X     ISO Abbreviation:  Biomed Eng Online     Publication Date:  2011  
Date Detail:
Created Date:  2011-09-06     Completed Date:  2011-11-21     Revised Date:  2013-06-28    
Medline Journal Info:
Nlm Unique ID:  101147518     Medline TA:  Biomed Eng Online     Country:  England    
Other Details:
Languages:  eng     Pagination:  65     Citation Subset:  IM    
Affiliation:
Department of Occupational Health, Tohoku University Graduate School of Medicine, Sendai 980, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Air Pressure
Airway Resistance
Diagnostic Techniques, Respiratory System
Linear Models
Oscillometry / methods*
Respiratory Function Tests / instrumentation*,  methods*
Respiratory Physiological Phenomena
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Biomed Eng Online
ISSN: 1475-925X
Publisher: BioMed Central
Article Information
Download PDF
Copyright ©2011 Ohishi and Kurosawa; licensee BioMed Central Ltd.
open-access:
Received Day: 13 Month: 5 Year: 2011
Accepted Day: 29 Month: 7 Year: 2011
collection publication date: Year: 2011
Electronic publication date: Day: 29 Month: 7 Year: 2011
Volume: 10First Page: 65 Last Page: 65
ID: 3166920
Publisher Id: 1475-925X-10-65
PubMed Id: 21798069
DOI: 10.1186/1475-925X-10-65

Time lag between oscillatory pressure and flow affecting accuracy of forced oscillation technique
Junichi Ohishi1 Email: ohishi-thk@umin.ac.jp
Hajime Kurosawa1 Email: kurosawa-thk@m.tohoku.ac.jp
1Department of Occupational Health, Tohoku University Graduate School of Medicine, Sendai, 980, Japan

The forced oscillation technique (FOT) is a simple method for assessing the oscillatory mechanics of the respiratory system, which has provided important findings in respiratory physiology [1,2]. Currently, the FOT using multi-frequency signals such as pseudo random noise or impulse signals is widely used in clinical research [3-6]. Although the FOT has been a well-established technique, some detailed experimental conditions would be different among institutions. To elucidate the effects of the experimental conditions may be important for more accurate FOT measurements. The purpose of this study was to evaluate whether time lags produced by the experimental conditions such as different positions of the sensors can affect the accuracy of the FOT.

The oscillatory properties, Rrs as the resistive component and Xrs as the reactive one, are calculated by a spectral relationship between the oscillatory pressure and flow. Generally, the oscillatory pressure and flow are measured simultaneously by sensors such as a pressure transducer or pneumotachograph near the subject's mouth, which are analyzed in the signal processing by Fourier transform [7]. In some previous papers, the sensors in the diagram were at different distances from the subject's mouth [2-4]. This difference would cause a time lag in the oscillatory pressure and flow in the signal processing. To evaluate whether the time lag could affect the measurement of the FOT, this effect was examined by numerical analysis as follows.

The respiratory impedance of a subject (Zsub) is often represented by a simple linear model of the resistance (R), inertance (L), and compliance (C). That is,

[Formula ID: bmcM1]
(1) 

where ω = 2πf (f = oscillatory frequency), . If the flow signals (f) are recorded τ seconds earlier than the pressure signals (p) due to differences in the positions of the sensors, the respiratory impedance (Zrs) is calculated as follows,

[Formula ID: bmcM2]
(2) 

The resonant and cutoff frequencies (f0 and fc respectively) are the properties of Xrs, the former of which is used to evaluate the respiratory system. Fc is only used to determine the Xrs-curve versus oscillatory frequency, which is not actually used in the FOT. Those parameters are defined by

[Formula ID: bmcM3]
(3) 

Then, Rrs and Xrs are given by

[Formula ID: bmcM4]
(4) 
[Formula ID: bmcM5]
(5) 
[Formula ID: bmcM6]
(6) 

Here, if R = 1 (standard value without units), f0 = 8 Hz, fc = 3 Hz, and τ varies from 0 to 600 μ seconds (every 200 μ seconds), Rrs and Xrs versus the medium oscillatory frequency (from 5 to 35 Hz) are as shown in figure 1. Rrs was estimated to be increased and Xrs decreased with an increase in τ, especially at the high oscillatory frequency of this range. Figure 2 shows those properties versus a high oscillatory frequency (from 10 to 500 Hz). Rrs and Xrs were strikingly different in the τ values in this frequency range. These results indicate that the time lag between the oscillatory pressure and flow can affect the oscillatory properties, suggesting that errors in the FOT measurements would occur. If the oscillatory pressure is regarded as the transmission of air, a time lag of 200 μ seconds would correspond to a distance of approximately 6.86 cm between the two points (under the condition in which the speed of sound is 343 m/s). Such a distance between the sensors may cause a time lag of several hundred microseconds, which can affect the accuracy of the FOT.

In conclusion, the time lag between the oscillatory pressure and flow may affect the accuracy of the FOT. This result suggests that the time lag needs to be minimized or compensated for in the signal processing for the optimal FOT measurement. Researchers should pay attention to such detailed experimental conditions of the FOT apparatus.


Competing interests

The authors declare that they have no competing interests.


Authors' contributions

OJ performed the analysis and drafted the manuscript. KH conceived of the study and participated in its design. All authors read and approved the final manuscript.


Acknowledgements and Funding

This study was supported by Grant-in-Aid for Japanese Society for the Promotion of Science Fellows.


References
DuBois AB,Brody AW,Lewis DH,Burgess BF Jr,Oscillation mechanics of lung and chest in manJ Appl PhysiolYear: 1956858759413331841
Oostveen E,MacLeod D,Lorino H,Farre R,Hantos Z,Desager K,Marchal F,ERS Task Force on Respiratory Impedance MeasurementsThe forced oscillation technique in clinical practice: methodology, recommendations and future developmentsEur Respir JYear: 2003221026104110.1183/09031936.03.0008940314680096
Frey U,Forced oscillation technique in infants and young childrenPaediatr Respir RevYear: 2005624625410.1016/j.prrv.2005.09.01016298307
Beydon N,Davis SD,Lombardi E,Allen JL,Arets HG,Aurora P,Bisgaard H,Davis GM,Ducharme FM,Eigen H,Gappa M,Gaultier C,Gustafsson PM,Hall GL,Hantos Z,Healy MJ,Jones MH,Klug B,Lødrup Carlsen KC,McKenzie SA,Marchal F,Mayer OH,Merkus PJ,Morris MG,Oostveen E,Pillow JJ,Seddon PC,Silverman M,Sly PD,Stocks J,Tepper RS,Vilozni D,Wilson NM,American Thoracic Society/European Respiratory Society Working Group on Infant and Young Children Pulmonary Function TestingAn official American Thoracic Society/European Respiratory Society statement: pulmonary function testing in preschool childrenAm J Respir Crit Care MedYear: 20071751304134510.1164/rccm.200605-642ST17545458
Faria AC,Lopes AJ,Jansen JM,Melo PL,Evaluating the forced oscillation technique in the detection of early smoking-induced respiratory changesBiomed Eng OnlineYear: 200982210.1186/1475-925X-8-22
Marchal F,Hall GL,Forced oscillation techniqueEur Respir MonYear: 201047121136
Michaelson ED,Grassman ED,Peters WR,Pulmonary mechanics by spectral analysis of forced random noiseJ Clin InvestYear: 1975561210123010.1172/JCI1081981184746

Figures

[Figure ID: F1]
Figure 1 

Effect of the time lag at a medium-frequency range. Rrs and Xrs at a frequency range from 5 to 35 Hz were calculated from equations 5 and 6, respectively. The time lag (τ) varies from 0 to 600 μ seconds (every 200 μ seconds).



[Figure ID: F2]
Figure 2 

Effect of the time lag at a high-frequency range. Rrs and Xrs at a frequency range from 10 to 500 Hz were calculated from equations 5 and 6, respectively. The time lag (τ) varies from 0 to 600 μ seconds (every 200 μ seconds).



Article Categories:
  • Letter to the Editor