Document Detail


Thrombogenic performance of a st. Jude bileaflet mechanical heart valve in a sheep model.
MedLine Citation:
PMID:  16436887     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The sheep model is preferred for chronic evaluation of prosthetic heart valves, surgical techniques, and endocardiographic studies. A bileaflet mechanical heart valve (MHV) was implanted into a sheep model to study its in vivo performance and to evaluate the thrombogenic potential of the valve. Transesophageal echocardiography and transcranial Doppler ultrasonography measurements were conducted before and after the valve implantation. Platelet activity state (PAS) assay measurements were also conducted before and after the implantation surgery. After sheep euthanasia, the MHV was explanted and scanning electron microscopy (SEM) was performed on the explanted valve to examine changes to the MHV surface. Tissue blocks were taken from the sheep brain, left ventricle, aorta, spleen, and lung lobes for histological examination. Our results indicated that after the MHV implantation, more embolic signals were detected in the sheep carotid artery, increasing monotonously as a function of implantation time. Echocardiographic parameters including blood aortic velocity, transvalvular pressure gradient, and velocity time integral increased. PAS increased significantly after valve implantation. SEM pictures demonstrated calcium and phosphate deposition on the valve surfaces. Histological examination demonstrated hemorrhage in the lung tissue, pulmonary thrombosis, and osteogenesis in heart tissue.
Authors:
Wei Yin; Irvin B Krukenkamp; Adam E Saltman; Glenn Gaudette; Krishnamurthy Suresh; Oscar Bernal; Jolyon Jesty; Danny Bluestein
Related Documents :
22033147 - The australian history of cardiac pacing: memories from a bygone era.
18095517 - Does valve design impact the maximum implantable bioprosthetic diameter? a prospective,...
7140157 - On-line classification of arterial stenosis severity using principal component analysis...
7878507 - Development of a new insertion technique and a new device for the placement of bifurcat...
1314677 - Interface between hydroxyapatite and mandibular human bone tissue.
19779727 - Neural response telemetry in patients with the double-array cochlear implant.
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  ASAIO journal (American Society for Artificial Internal Organs : 1992)     Volume:  52     ISSN:  1058-2916     ISO Abbreviation:  ASAIO J.     Publication Date:    2006 Jan-Feb
Date Detail:
Created Date:  2006-01-26     Completed Date:  2006-04-12     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9204109     Medline TA:  ASAIO J     Country:  United States    
Other Details:
Languages:  eng     Pagination:  28-33     Citation Subset:  IM    
Affiliation:
Department of Biomedical Engineering, Division of Hematology, Stony Brook University, Stony Brook, NY, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Aorta / physiopathology
Aortic Valve / surgery,  ultrasonography
Blood Flow Velocity
Blood Platelets / metabolism
Calcium / metabolism
Carotid Arteries / physiopathology
Echocardiography, Transesophageal
Female
Follow-Up Studies
Heart Valve Prosthesis / adverse effects*
Heart Valve Prosthesis Implantation / adverse effects*
Microscopy, Electron, Scanning
Phosphates / metabolism
Platelet Activation
Prosthesis Design
Sheep
Thrombosis / etiology*,  pathology
Time Factors
Treatment Outcome
Ultrasonography, Doppler, Transcranial
Chemical
Reg. No./Substance:
0/Phosphates; 7440-70-2/Calcium

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine