Document Detail


Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients.
MedLine Citation:
PMID:  18496416     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVES: No simple method exists for repeatedly measuring cardiac output in intensive care pediatric and neonatal patients. The purpose of this study is to present the theory and examine the in vitro accuracy of a new ultrasound dilution cardiac output measurement technology in which an extracorporeal arteriovenous tubing loop is inserted between existing arterial and venous catheters.
DESIGN: Laboratory experiments.
SETTING: Research laboratory.
SUBJECTS: None.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: In vitro validations of cardiac output, central blood volume, total end-diastolic volume, and active circulation volume were performed in a model mimicking pediatric (children 2-10 kg) and neonatal (0.5-3 kg) flows and volumes against flows and volumes measured volumetrically. Reusable sensors were clamped onto the arterial and venous limbs of the arteriovenous loop. A peristaltic pump was used to circulate liquid at 6-12 mL/min from the artery to the vein through the arteriovenous loop. Body temperature injections of isotonic saline (0.3-10 mL) were performed. In the pediatric setting, the absolute difference between cardiac output measured by dilution and cardiac output measured volumetrically was 3.97% +/- 2.97% (range 212-1200 mL/min); for central blood volume the difference was 4.59% +/- 3.14% (range 59-315 mL); for total end-diastolic volume the difference was 4.10% +/- 3.08% (range 24-211 mL); and for active circulation volume the difference was 3.30% +/- 3.07% (range 247-645 mL). In the neonatal setting the difference for cardiac output was 4.40% +/- 4.09% (range 106-370 mL/min); for central blood volume the difference was 4.90% +/- 3.69% (range 50-62 mL); and for active circulation volume the difference was 5.39% +/- 4.42% (range 104-247 mL).
CONCLUSIONS: In vitro validation confirmed the ability of the ultrasound dilution technology to accurately measure small flows and volumes required for hemodynamic assessments in small pediatric and neonatal patients. Clinical studies are in progress to assess the reliability of this technology under different clinical situations.
Authors:
Nikolai M Krivitski; Victor V Kislukhin; Naveen V Thuramalla
Related Documents :
568246 - Radioactive microsphere partitioning of blood flow between capillaries and arteriovenou...
2155626 - Effective pulmonary flow, aortic flow and cardiac output: in vitro and in vivo comparis...
8434826 - Mechanisms of blood flow during cpr.
6466526 - A sheep preparation for studying interactions between blood flow and drug disposition. ...
10749636 - Predictors of blood transfusions in spinal instrumentation and fusion surgery.
7047006 - In-vivo activation of circulating inactive renin by the ischemic kidney in man.
Publication Detail:
Type:  Journal Article; Research Support, N.I.H., Extramural    
Journal Detail:
Title:  Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies     Volume:  9     ISSN:  1529-7535     ISO Abbreviation:  Pediatr Crit Care Med     Publication Date:  2008 Jul 
Date Detail:
Created Date:  2008-10-09     Completed Date:  2009-03-09     Revised Date:  2011-09-26    
Medline Journal Info:
Nlm Unique ID:  100954653     Medline TA:  Pediatr Crit Care Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  423-8     Citation Subset:  IM    
Affiliation:
Transonic Systems Inc., Ithaca, NY, USA. nikolai.krivitski@transonic.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Cardiac Output
Catheterization, Central Venous
Catheterization, Peripheral
Child
Extracorporeal Circulation*
Hemodynamics*
Humans
Infant, Newborn
Intensive Care Units, Pediatric
Models, Cardiovascular*
Ultrasonics
Grant Support
ID/Acronym/Agency:
R44 HL061994/HL/NHLBI NIH HHS; R44 HL061994-03/HL/NHLBI NIH HHS
Comments/Corrections
Comment In:
Pediatr Crit Care Med. 2008 Jul;9(4):449-50   [PMID:  18496401 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A multidisciplinary survey on capillary refill time: Inconsistent performance and interpretation of ...
Next Document:  Risk factors for pneumothorax in very low birth weight infants.