Document Detail


Technical and Patient-related Characteristics Associated with Challenging Retrieval of Inferior Vena Cava Filters.
MedLine Citation:
PMID:  23831140     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE: To identify patient-related and device-specific predictors of challenging and failed inferior vena cava (IVC) filter retrievals.
METHODS: Retrospective single center review of consecutive retrievable IVC filters placed between 2004 and 2009. Retrieval was defined as challenging when it was unsuccessful owing to reported technical failure or when adjunctive endovascular maneuvers or access sites were recruited. Data regarding patient- and filter-specific information were collected. Logistic regression models were used to identify predictors of the reported outcomes. Statistical significance was set at p < .05.
RESULTS: Four hundred and one patients underwent retrievable IVC filter placement-the majority indicated for prophylaxis (67%). Two hundred and fifty-nine retrievals were attempted and 237 filters were successfully retrieved (overall retrieval rate: 59.1%). Eleven out of 259 (4.2%) attempts were aborted owing to significant thrombus within the filter and 11 (4.2%) were technically unsuccessful. In 142 patients no attempt for filter retrieval was made-the major reason being physician oversight (44.3%). Thirty-eight out of 248 (15.3%) non-aborted filter retrievals were recorded as challenging. Failed retrievals were predicted by prolonged dwell time (96.9 ± 111.9 vs. 29.5 ± 25.1 days, odds ratio [OR] 1.034, 95% confidence interval [CI] 1.016-1.053, p < .001), therapeutic indication (OR 5.197, 95% CI 1.200-22.511, p = .028), and filter wall apposition (OR 11.857, 95% CI 2.069-67.968, p = .006). Challenging retrievals were predicted by dwell time (51.1 ± 69.8 vs. 29.1 ± 24.5 days, OR 1.017, 95% CI 1.005-1.029, p = .007), filter tilt (OR 2.607, 95% CI 1.045-6.508, p = .040) and filter wall apposition (OR 6.149, 95% CI 2.398-15.763, p = <.001).
CONCLUSIONS: Physician oversight leads to poor IVC filter retrieval rates. Retrievals can be challenging or fail when the dwell time is >50 days and >90 days, respectively, and when the filter hook apposes the caval wall. Filter tilt increases retrieval difficulty but not failure rates.
Authors:
E D Avgerinos; J Bath; J Stevens; B McDaniel; L Marone; E Dillavou; J S Cho; M S Makaroun; R A Chaer
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2013-7-4
Journal Detail:
Title:  European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery     Volume:  -     ISSN:  1532-2165     ISO Abbreviation:  Eur J Vasc Endovasc Surg     Publication Date:  2013 Jul 
Date Detail:
Created Date:  2013-7-8     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9512728     Medline TA:  Eur J Vasc Endovasc Surg     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2013 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.
Affiliation:
Division of Vascular Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Televised relational and physical aggression and children's hostile intent attributions.
Next Document:  Interhuman transmissibility of Middle East respiratory syndrome coronavirus: estimation of pandemic ...