Technical Note on VDS. A System for the Endovascular Electrolytical Detachment of Platinum Coils at Variable Length.
MedLine Citation:
PMID:  20594529     Owner:  NLM     Status:  PubMed-not-MEDLINE    
A new system was deviced that allows the electrolytical detachment of platinum microcoils at variable lengths for the endovascular treatment of intracranial aneurysms. The detachment element consists of two short platinum coil segments, which are connected by a threat of stainless steel. The steel threat is interrupted by electrolysis, using a continuous current with 1-2 mA at a voltage of 4-8 Volt. The average detachment time in heparinized blood is about 30-40 sec. The detachment elements can be used to connect either an insertion wire with a platinum coil or to connect several segments of platinum coils with variable helices and lengths. If several detachment elements are integrated in a coil, electrolysis interrupts only the element, which is next beyond the distal tip of the catheter. The electrolytical process does not affect the detachment elements proximal and distal to the element adjacent to the tip of the catheter. Single or several coil segments can be pulled back into the microcatheter if necessary. The system is expected to allow a faster and more complete endovascular coil occlusion of intracranial aneurysms. The principles described in this paper are claimed by the German pending patent DE 100 10840 A1.
H Henkes; P Drepper; J Speder; S Felber; H Monstadt; D Kühne
Related Documents :
8733959 - Use of color power transcranial doppler sonography to monitor aneurysmal coiling.
15133679 - Purtscher-like retinopathy and horner's syndrome following coil embolization of an intr...
15018329 - Ischemic basilar artery dissecting aneurysm treated by stenting only--case report.
14739309 - Value of transcatheter arterial embolization with coils and n-butyl cyanoacrylate for l...
18948679 - Basilar apex aneurysm manifesting as third ventricular mass and obstructive hydrocephal...
16823299 - Feasibility and limitations of endovascular coil embolization of anterior communicating...
6741649 - Early effects of experimental arterial subarachnoid haemorrhage on the cerebral circula...
1961409 - Cervical vertebral artery aneurysms and arteriovenous fistulae in neurofibromatosis typ...
2534059 - Pulmonary thromboembolism following angioplasty for membranous occlusion of the vena ca...
Publication Detail:
Type:  Journal Article     Date:  2004-10-20
Journal Detail:
Title:  Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences     Volume:  8     ISSN:  1591-0199     ISO Abbreviation:  Interv Neuroradiol     Publication Date:  2002 Jun 
Date Detail:
Created Date:  2010-07-02     Completed Date:  2012-10-02     Revised Date:  2014-07-30    
Medline Journal Info:
Nlm Unique ID:  9602695     Medline TA:  Interv Neuroradiol     Country:  Italy    
Other Details:
Languages:  eng     Pagination:  197-200     Citation Subset:  -    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Previous Document:  Multiple dural arteriovenous fistulas. Radiologic progression and endovascular cure. Case report.
Next Document:  The usefulness of external marking in stenting for m1 segment of middle cerebral artery stenosis. Te...