Target Repositional Accuracy and PTV Margin Verification using Three-dimensional Cone-Beam Computed Tomography (CBCT) in Stereotactic Body Radiotherapy (SBRT) of Lung Cancers.
MedLine Citation:
PMID:  22402387     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
The purpose of this study was to assess target repositional accuracy with respect to the bony structures using daily CBCT, and to validate the planning target volume (PTV) margin used in the lung SBRT. All patients underwent 4D CT scanning in preparation for lung SBRT. The internal target volume (ITV) was outlined from the reconstructed 4D data using the maximum-intensity projection (MIP) algorithm. A 6 mm margin was added to the ITV to create the PTV. Conformal treatment planning was performed on the helical images, to which the MIP images were fused. Prior to each treatment, CBCT was taken after a patient was set up in the treatment position. The CBCT images were fused with the simulation CT based on the bony anatomy, in order to derive setup errors and separate them from the tumor repositional errors. The treating physician then checked and modified the alignment based on target relocalization within the PTV. The shifts determined in such a method were recorded and the subtractions of these shifts with respect to the corresponding setup errors were defined as the target relocalization accuracy. Our study of 36 consecutive patients, treating 38 targets for a total of 153 fractions shows that, after setup error correction, the target repositional accuracy followed a normal distribution with the mean values close to 0 in all directions, and standard deviations of 0.25 cm in A-P, 0.24 cm in Lat, and 0.28 cm in S-I directions, respectively. The probability of having the shifts ? 0.6 cm is less than 0.8% in A-P, 0.6% in Lat, and 1.7 % in S-I directions. For the patient population studied, the target centroid position relative to the bony structures changed minimally from day to day. This demonstrated that the PTV margin that is designed on the MIP image-based ITV was adequate for lung SBRT.
Authors:
Lu Wang; Steven Feigenberg; Jiajian Fan; Lihui Jin; Aruna Turaka; Lili Chen; C-M Charlie Ma
Related Documents :
22333427 - Pre-operative imaging for cochlear implantation: magnetic resonance imaging, computed t...
22451537 - Jpeg2000 compression of ct images used for measuring coronary artery calcification scor...
22528897 - Ct enterography at 80 kvp with adaptive statistical iterative reconstruction versus at ...
24049757 - A typical manifestation of birthmark-becker's nevus syndrome.
16103857 - The modic classification: inter- and intraobserver error in clinical practice.
14716237 - Evaluation of extraocular muscles in the edematous phase of graves ophthalmopathy on co...
Publication Detail:
Type:  Journal Article     Date:  2012-03-08
Journal Detail:
Title:  Journal of applied clinical medical physics / American College of Medical Physics     Volume:  13     ISSN:  1526-9914     ISO Abbreviation:  J Appl Clin Med Phys     Publication Date:  2012  
Date Detail:
Created Date:  2012-03-09     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101089176     Medline TA:  J Appl Clin Med Phys     Country:  United States    
Other Details:
Languages:  eng     Pagination:  3708     Citation Subset:  IM    
Affiliation:
Fox Chase Cancer Center. l_wang@fccc.edu.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Dosimetric effects of jaw tracking in step-and-shoot intensity-modulated radiation therapy.
Next Document:  Analysis of the optimum internal margin for respiratory-gated radiotherapy using end-expiratory phas...