Systematic Evaluation of a Novel Foot-Pump Ureteroscopic Irrigation System.
MedLine Citation:
PMID:  22149147     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Abstract Introduction: To evaluate forces exerted on a stone with different ureteroscopic irrigation systems. Methods: A 3 mm steel simulated stone was welded to a 3F stone basket and inserted into the working channel of a 4.5F Wolf semirigid ureteroscope. The basket shaft was attached to a 50 g load cell. The urterescope was placed in a ureteral model (14F silicon tubing). Simulated blood (McCormick™ Red Dye, 1 dye drop/15 mL H(2)O) was dripped adjacent to the stone at 12 drops/min. Endoirrigation devices were attached to the ureterscope and irrigation was applied at a rate sufficient to maintain visualization of the stone. Force on the stone was measured with the following endoirrigation systems: Boston Scientific™ Single-Action-Pump System (SAP) hand-pump and NuVista Medical™ Flo-Assist(®) foot-pump. Results: No significant difference (p=0.19) in the number of pumps required to maintain a clear endoscopic field was found between the SAP (0.20±0.075/second) and Flo-Assist device (0.25±0.056/second). The pump duration of the Flo-Assist was found to be less (1.12±0.40 seconds) than the SAP (1.35±0.31 seconds), but not significantly different (p=0.24). The average maximum impulse was significantly lower (p=0.0002) for the SAP (8.34×10(-5) Ns) than the Flo-Assist (1.96×10(-3) Ns). Total maximum impulse (2.02×10(-5) Ns) for the SAP and total average impulse (5.51×10(-6) Ns) were found to be lower than the Flo-Assist device. The Flo-Assist had a total maximum impulse of 4.49×10(-4) Ns and total average impulse of 8.85×10(-6) Ns, however, these differences were not statistically significant. Conclusion: The hand-pump (SAP) and foot-pump (Flo-Assist) irrigation devices tested require comparable amounts of pumps for similar durations of time to maintain endoscopic visualization. Overall, the SAP device exerts less average maximum force on the stone than the Flo-Assist device.
Authors:
Kari Hendlin; Carl Sarkissian; Branden Duffey; Manoj Monga
Related Documents :
17552007 - Long-term results of subtotal colectomy with cecorectal anastomosis for isolated coloni...
16470577 - Prenatal counselling of small bowel atresia: watch the fluid!
7209787 - Small bowel obstruction: the role of nonoperative treatment in simple intestinal obstru...
21435917 - Histologic and biomechanical evaluation of crosslinked and non-crosslinked biologic mes...
11379997 - Steroid injection and splinting in the treatment of carpal tunnel syndrome.
2617357 - Lateral electrical surface stimulation in idiopathic scoliosis. experience in two priva...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-12-13
Journal Detail:
Title:  Journal of endourology / Endourological Society     Volume:  -     ISSN:  1557-900X     ISO Abbreviation:  -     Publication Date:  2011 Dec 
Date Detail:
Created Date:  2011-12-13     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8807503     Medline TA:  J Endourol     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
1 Department of Urologic Surgery, University of Minnesota , Minneapolis, Minnesota.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Alternative Technique for Laparoscopic Port-Site Closure.
Next Document:  Prevalence of Legionella spp. in water systems of hospitals and hotels in South Western Greece.