Document Detail


System survival of nontransvenous implantable cardioverter-defibrillators compared to transvenous implantable cardioverter-defibrillators in pediatric and congenital heart disease patients.
MedLine Citation:
PMID:  20022820     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Nontransvenous (NTV) implantable cardioverter-defibrillator (ICD) systems with pericardial and/or subcutaneous coils are used in select pediatric and congenital heart disease patients who are not candidates for transvenous ICD leads. Outcomes with these hybrid configurations are not well understood. OBJECTIVE: The purpose of this study was to compare survival of NTV ICD systems to standard transvenous (TV) ICD systems. METHODS: We conducted a retrospective single-center study in which the TV group was matched to the NTV group 2:1 by type of cardiac disease and implant date. RESULTS: There were 39 patients in the NTV group and 78 matched in the TV group. Compared to the TV group, the NTV group was younger (median 7 vs 20 years) with a smaller body surface area at implant (0.9 vs 1.8 m(2); both P <.001). Median follow-up in the NTV group was 17 months. System survival at 12, 24, and 36 months was 73%, 55%, and 49% in the NTV group versus 91%, 83%, and 76% in the TV group (P = .003). A multivariable Cox proportional hazards model including group, body surface area, and age at implant revealed the NTV group to be an independent predictor of system failure (hazard ratio 2.9, P = .04). Rate of total unanticipated interventions in the NTV group was 18 versus 6 per 1,000 person-months in the TV group. In patients with NTV systems, 23% received appropriate shocks and 18% received inappropriate shocks. CONCLUSION: Survival of ICD systems using NTV defibrillation coils is significantly shorter than with TV ICD systems. Although NTV systems provide protection for this unique subset of patients, more durable options are needed.
Authors:
Andrew E Radbill; John K Triedman; Charles I Berul; Francis Fynn-Thompson; Joseph Atallah; Mark E Alexander; Edward P Walsh; Frank Cecchin
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't     Date:  2009-10-14
Journal Detail:
Title:  Heart rhythm : the official journal of the Heart Rhythm Society     Volume:  7     ISSN:  1556-3871     ISO Abbreviation:  Heart Rhythm     Publication Date:  2010  
Date Detail:
Created Date:  2010-02-04     Completed Date:  2010-05-18     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101200317     Medline TA:  Heart Rhythm     Country:  United States    
Other Details:
Languages:  eng     Pagination:  193-8     Citation Subset:  IM    
Affiliation:
Department of Cardiology, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Case-Control Studies
Child
Child, Preschool
Defibrillators, Implantable*
Equipment Failure Analysis*
Female
Follow-Up Studies
Heart Defects, Congenital / radiography,  therapy*
Heart Diseases / radiography,  therapy*
Humans
Male
Proportional Hazards Models
Retrospective Studies
Statistics, Nonparametric
Young Adult