Abstract/OtherAbstract:
A young woman presented with severe heart failure symptoms 4 years after percutaneous device closure of an atrial septal defect (ASD). There was residual left-to-right shunting, and the device was obstructing the inferior caval vein and tricuspid valve flow. Intraoperatively, the ASD was shown to be an inferior sinus venosus defect, and reconstruction of the mitral valve and the posterior wall of the left atrium was required. This case emphasizes the importance of appropriate patient selection for interventional ASD closure and demonstrates that patients with less-than-optimal results should be referred for surgical correction early, before the device produces secondary damage to atrial structures and atrioventricular valves.
Authors:
Kadir Yilmaz; Peter Ewert; Roland Hetzer; Christof Stamm
Related Documents :
6846953 - Oculoplethysmography. evaluation of patients with transient ischemic attacks from inter...
2597623 - Intimal fibroplasia: an unusual etiology of stroke.
21927863 - The short- and medium-term results of transcatheter closure of atrial septal defect wit...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-1-11
Journal Detail:
Title:  Interactive cardiovascular and thoracic surgery     Volume:  -     ISSN:  1569-9285     ISO Abbreviation:  -     Publication Date:  2012 Jan 
Date Detail:
Created Date:  2012-1-12     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101158399     Medline TA:  Interact Cardiovasc Thorac Surg     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Department of Thoracic and Cardiovascular Surgery, German Heart Institute Berlin, Berlin, Germany.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Estimation of Fan Beam and Parallel Beam Parameters in a Wire Mesh Design.
Next Document:  Assessment of the Prognostic Value of the World Health Organization Clinical Staging System for HIV/...