Suitable location and optimum bladder width for the occluding cuff used for indirect arterial blood pressure measurements at the wrist.
MedLine Citation:
PMID:  10354907     Owner:  NLM     Status:  MEDLINE    
Using a finite element method (FEM) and a volume oscillometric method with a photo-electric plethysmograph, we tried to determine which location(s) and which bladder width(s) of the occluding cuff would allow the most accurate indirect blood pressure measurements at the human wrist. Nuclear magnetic resonance imaging of the wrist allowed us to determine the geometry of two-dimensional (2D) finite element models. FEM analysis showed that the pressure transmission ratio (calculated tissue pressure: externally applied pressure) from the cuff to the radial artery (RA) was 100% when the cuff was placed over the site (L1) at which the RA crosses the most protuberant spot on the volar aspect of the distal end of the radius. In fact, the mean blood pressure (BPM) measured at this site in human subjects coincided with that measured at the upper arm (BPMA). In addition, a three-dimensional (3D) FEM was constructed by extending the 2D FEM at site L1 in the longitudinal direction to investigate the relation between pressure transmission to the RA and cuff bladder width. FEM analysis using the 3D finite element model showed that for all cuff widths greater than 0.25 times the diameter of the wrist, the external pressure was transmitted almost completely to the RA beneath the cuff center. Moreover, when the cuff width was as large as 0.45 times the diameter of the wrist, the BPM measured at site L1 in human subjects was similar to the BPMA. The theoretical and experimental results in the present study both suggest that for accurate indirect blood pressure measurements at the wrist, the L1 site is a suitable location for the occluding cuff and a cuff whose bladder width is more than one-third but less than one-half the wrist diameter can be used.
W Lu; A Tsukada; A Kawarada; K Sasaki
Related Documents :
12120967 - A prospective comparison of pedal ergometry with conventional treadmill testing in the ...
2310047 - Skin perfusion pressure assessed by measuring the external pressure required to stop bl...
12664047 - Hemodynamics of infrapopliteal ptfe bypasses and adjunctive arteriovenous fistulas.
1124477 - Systolic slope and other pressure measurements in patients with peripheral vascular dis...
21571067 - Diuretic bioactivity optimization of furosemide in rats.
25117597 - Use of a coverlet system for the management of skin microclimate.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Frontiers of medical and biological engineering : the international journal of the Japan Society of Medical Electronics and Biological Engineering     Volume:  9     ISSN:  0921-3775     ISO Abbreviation:  Front Med Biol Eng     Publication Date:  1999  
Date Detail:
Created Date:  1999-07-20     Completed Date:  1999-07-20     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  9011464     Medline TA:  Front Med Biol Eng     Country:  NETHERLANDS    
Other Details:
Languages:  eng     Pagination:  9-29     Citation Subset:  IM    
Division of Bio-Information Engineering, Faculty of Engineering, Toyama University, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Blood Pressure Determination / instrumentation*,  methods*
Computer Simulation
Finite Element Analysis
Magnetic Resonance Imaging
Models, Biological
Oscillometry / instrumentation,  methods*
Wrist / anatomy & histology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Previous Document:  The penetration characteristic of radio wave frequencies from oral cavities.
Next Document:  Reliability of measurement of muscle fiber conduction velocity using surface EMG.