| Suitability of helical multislice acquisition technique for routine unenhanced brain CT: an image quality study using a 16-row detector configuration. | |
| | |
MedLine Citation:
|
PMID: 16912851 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
Subjective and objective image quality (IQ) criteria, radiation doses, and acquisition times were compared using incremental monoslice, incremental multislice, and helical multislice acquisition techniques for routine unenhanced brain computed tomography (CT). Twenty-four patients were examined by two techniques in the same imaging session using a 16-row CT system equipped with 0.75-width detectors. Contiguous "native" 3-mm-thick slices were reconstructed for all acquisitions from four detectors for each slice (4x0.75 mm), with one channel available per detector. Two protocols were tailored to compare: (1) one-slice vs four-slice incremental images; (2) incremental vs helical four-slice images. Two trained observers independently scored 12 subjective items of IQ. Preference for the technique was assessed by one-tailed t test and the interobserver variation by two-tailed t test. The two observers gave very close IQ scores for the three techniques without significant interobserver variations. Measured IQ parameters failed to reveal any difference between techniques, and an approximate half radiation dose reduction was obtained by using the full 16-row configuration. Acquisition times were cumulatively shortened by using the multislice and the helical modality. |
| | |
Authors:
|
Danielle Hernalsteen; Guy Cosnard; Annie Robert; Cécile Grandin; Alain Vlassenbroek; Thierry Duprez |
Publication Detail:
|
Type: Journal Article; Validation Studies Date: 2006-08-16 |
Journal Detail:
|
Title: European radiology Volume: 17 ISSN: 0938-7994 ISO Abbreviation: Eur Radiol Publication Date: 2007 Apr |
Date Detail:
|
Created Date: 2007-03-07 Completed Date: 2007-07-31 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 9114774 Medline TA: Eur Radiol Country: Germany |
Other Details:
|
Languages: eng Pagination: 975-82 Citation Subset: IM |
Affiliation:
|
Department of Radiology and Medical Imaging, Université Catholique de Louvain, Cliniques Universitaires Saint-Luc, Avenue Hippocrate, 10, 1200, Brussels, Belgium. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Aged Aged, 80 and over Brain / radiography* Diagnostic Tests, Routine Female Humans Male Middle Aged Observer Variation Radiation Dosage Radiographic Image Enhancement* Tomography, Spiral Computed / methods* |