Subject-ventilator synchrony during neural versus pneumatically triggered non-invasive helmet ventilation.
Jump to Full Text
MedLine Citation:
PMID:  18512045     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: Patient-ventilator synchrony during non-invasive pressure support ventilation with the helmet device is often compromised when conventional pneumatic triggering and cycling-off were used. A possible solution to this shortcoming is to replace the pneumatic triggering with neural triggering and cycling-off-using the diaphragm electrical activity (EA(di)). This signal is insensitive to leaks and to the compliance of the ventilator circuit.
DESIGN: Randomized, single-blinded, experimental study.
SETTING: University Hospital. PARTICIPANTS AND SUBJECTS: Seven healthy human volunteers.
INTERVENTIONS: Pneumatic triggering and cycling-off were compared to neural triggering and cycling-off during NIV delivered with the helmet.
MEASUREMENTS AND RESULTS: Triggering and cycling-off delays, wasted efforts, and breathing comfort were determined during restricted breathing efforts (<20% of voluntary maximum EA(di)) with various combinations of pressure support (PSV) (5, 10, 20 cm H(2)O) and respiratory rates (10, 20, 30 breath/min). During pneumatic triggering and cycling-off, the subject-ventilator synchrony was progressively more impaired with increasing respiratory rate and levels of PSV (p < 0.001). During neural triggering and cycling-off, effect of increasing respiratory rate and levels of PSV on subject-ventilator synchrony was minimal. Breathing comfort was higher during neural triggering than during pneumatic triggering (p < 0.001).
CONCLUSIONS: The present study demonstrates in healthy subjects that subject-ventilator synchrony, trigger effort, and breathing comfort with a helmet interface are considerably less impaired during increasing levels of PSV and respiratory rates with neural triggering and cycling-off, compared to conventional pneumatic triggering and cycling-off.
Authors:
Onnen Moerer; Jennifer Beck; Lukas Brander; Roberta Costa; Michael Quintel; Arthur S Slutsky; Fabrice Brunet; Christer Sinderby
Related Documents :
16682245 - Advanced lung ventilation system (alvs) with linear respiratory mechanics assumption fo...
7836215 - Effect of arterial blood pressure and ventilation gases on cardiac depression induced b...
6580475 - Pao2 and intratracheal pressure in oscillatory ventilation in experimental respiratory ...
11457775 - Effects of hypergravity and anti-g suit pressure on intraregional ventilation distribut...
6127955 - Electrically coupled intrinsic responses of feline lower esophageal sphincter.
1905595 - Concurrent hydralazine administration prevents nitroglycerin-induced hemodynamic tolera...
Publication Detail:
Type:  Journal Article; Randomized Controlled Trial; Research Support, Non-U.S. Gov't     Date:  2008-05-30
Journal Detail:
Title:  Intensive care medicine     Volume:  34     ISSN:  0342-4642     ISO Abbreviation:  Intensive Care Med     Publication Date:  2008 Sep 
Date Detail:
Created Date:  2008-08-18     Completed Date:  2009-01-08     Revised Date:  2013-06-05    
Medline Journal Info:
Nlm Unique ID:  7704851     Medline TA:  Intensive Care Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1615-23     Citation Subset:  IM    
Affiliation:
Interdepartmental Division of Critical Care, St. Michael's Hospital, University of Toronto, Toronto, ON, Canada.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Equipment Design
Female
Head Protective Devices*
Humans
Male
Patient Satisfaction*
Positive-Pressure Respiration / instrumentation*,  methods
Respiration
Single-Blind Method
Comments/Corrections
Comment In:
Intensive Care Med. 2008 Sep;34(9):1562-4   [PMID:  18512044 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Intensive Care Med
ISSN: 0342-4642
ISSN: 1432-1238
Publisher: Springer-Verlag, Berlin/Heidelberg
Article Information
Download PDF
© The Author(s) 2008
Received Day: 21 Month: 12 Year: 2007
Accepted Day: 16 Month: 4 Year: 2008
Electronic publication date: Day: 30 Month: 5 Year: 2008
Print publication date: Month: 9 Year: 2008
Volume: 34 Issue: 9
First Page: 1615 Last Page: 1623
ID: 2517084
PubMed Id: 18512045
Publisher Id: 1163
DOI: 10.1007/s00134-008-1163-z
issue-copyright-statement: © Springer-Verlag 2008

Subject–ventilator synchrony during neural versus pneumatically triggered non-invasive helmet ventilation
Onnen Moerer12
Jennifer Beck3
Lukas Brander14
Roberta Costa5
Michael Quintel2
Arthur S. Slutsky168
Fabrice Brunet67
Christer Sinderby678 Address: +1-416-8646060 sinderbyc@smh.toronto.on.ca
1Interdepartmental Division of Critical Care, St. Michael’s Hospital, University of Toronto, Toronto, ON Canada
2Department of Anaesthesiology, Emergency and Critical Care Medicine, University of Göttingen, Göttingen, Germany
3Neonatal Intensive Care Unit, Department of Pediatrics, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, ON Canada
4Department of Intensive Care Medicine, University Hospital, Inselspital, Bern, Switzerland
5Department of Anesthesiology and Intensive Care Medicine, Catholic University of Rome, Rome, Italy
6Department of Critical Care Medicine, St. Michael’s Hospital, University of Toronto, 30 Bond Street, Room 4-072, Queen wing, Toronto, ON M5B 1W8 Canada
7Department of Medicine, University of Toronto, Toronto, ON Canada
8Keenan Research Centre in the Li Ka Shing Knowledge Institute of St. Michael’s Hospital, Toronto, ON Canada

Introduction

Non-invasive ventilation (NIV) is an important complement to invasive mechanical ventilation [14] in patients with acute exacerbations of chronic obstructive pulmonary disease (COPD) [5], and with severe cardiogenic pulmonary edema [6].

Problems with the commonly used interfaces for NIV application include air leakage [7, 8], patient discomfort [9] and pressure-related ulcerations of the nose [10], which can limit the duration of use and account for a large proportion of NIV failures [11]. Navalesi and co-workers [8] demonstrated that the type of interface used for NIV is important with respect to patient tolerance and the time of NIV application. A new NIV interface, the helmet, has been introduced recently and was tested in different clinical situations [1215]. The results are promising in terms of user acceptance [13]. However, due to the large collapsible and compliant chamber encompassing the patient’s head it significantly impairs patient–ventilator synchrony with conventional pneumatic systems [16, 17] and is less effective in reducing the work of breathing than the face mask [16, 18].

New methods for neural triggering and cycling-off, using the diaphragm electrical activity (EAdi), can be used to initiate and terminate ventilatory assist in synchrony with inspiratory efforts and, hence, may overcome some of the shortcomings of conventional pneumatic triggering and cycling-off with the helmet [19]. The aim of this study was to compare synchrony between inspiratory effort and ventilator assist, as well as breathing comfort during neurally or pneumatically triggered and cycled-off non-invasive pressure support ventilation delivered with the helmet interface in healthy volunteers.

Some of the data have been presented at the Congress of the European Society of Intensive Care Medicine in 2006 [20].


Materials and methods
Subjects

Seven healthy subjects (four female) were studied. Their mean (±SD) age, height and weight were 37 ± 5 years, 173 ± 10 cm and 67 ± 14 kg, respectively. Six subjects had prior knowledge about mechanical ventilation. The study was approved by the Research Ethics Committee of St. Michael’s Hospital in Toronto. Subjects gave their informed consent in writing.

Instrumentation

An 8-Fr catheter equipped with sensors for measurement of EAdi [21], and balloons for measurement of esophageal (Pes) and gastric (Pga) pressure [22] were inserted transnasally into the stomach [21, 23]. EAdi sensors were positioned at the level of the diaphragm by online supervision of recorded electrical and pressure signals [23]. Airway pressure was measured inside the helmet close to the subject’s mouth. All signals were digitized at 2 kHz and stored on a personal computer for off-line evaluation. Expiratory muscle activity was monitored by the differential EMG signal from two surface electrodes placed 30-cm apart on either side of the mid-line of the upper abdominal wall.

All subjects received NIV with the helmet (Starmed Castar “R”, Mirandola, Italy) via a modified conventional ICU ventilator (Servo 300, Maquet Critical Care, Solna, Sweden).

NIV was delivered either with pneumatic trigger (Ptr) and cycling-off (Poff) or with neural trigger (Ntr) and cycling-off (Noff). Ptr and Ntr were adjusted to avoid auto-cycling, verified by performing a 10-s period of apnea (group mean value for Ptr was −1.3 ±SD 0.6 cm H2O). Conventional flow cycling-off (Poff) was 5% of peak inspiratory flow (default on Servo300). Ntr was set to trigger when the EAdi exceeded the random noise-variability. Noff occurred when EAdi fell to 60% of peak EAdi. During all runs, a PEEP of 5 cm H2O was applied.

Protocol

All subjects were studied while seated. The maximum EAdi[\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{(EA}}_{{\text{di}}_{{ \max }} } ) $$\end{document}] was obtained by a maximal inspiratory maneuver [23]. Subsequently, during a practice period, subjects were instructed to follow a specific respiratory rate with a predetermined rate displayed as a time-line on a computer screen until they were able to maintain the predetermined respiratory rate (RR). Afterward the subjects repeated periods of breathing on the ventilator with RR of 10, 20, or 30 breaths per min (bpm), at pressure support (PSV) levels of 5, 10, or 20 cm H2O above PEEP (5 cm H2O). NIV was either neurally (Ntr and Noff) or pneumatically (Ptr and Poff) triggered and cycled-off. A total of 18 combinations of 2-min duration, were randomly performed using ballots. Before the 2-min period of data recording, each subject breathed until the new target respiratory rate and assist level was reached. Each measurement period was followed by 2 min of rest with CPAP of 5 cm H2O. Subjects were not informed about the applied PSV level, or the trigger type.

To limit the subjects’ breathing efforts, the inspiratory effort was monitored by the investigators throughout the protocol period, and subjects were instructed to lower their inspiratory effort if EAdi exceeded 20% of the voluntary maximum EAdi. Subjects were also instructed to not use expiratory muscles (supervised by monitoring the abdominal EMG).

Subjects comfort of breathing was assessed by a visual analogue scale (VAS) (0 mm = maximal comfort to 100 mm = unbearable) and marked by the subjects themselves at the end of each study period.

Off-line data analysis

Ntr and Ptr delays were determined by measuring the time between the onset of EAdi and the onset of ventilatory assist using the internal signal of the ventilator. Trigger delays during Ntr NIPSV could vary (see ESM for details). Noff and Poff were determined by measuring the time between the point where EAdi was reduced to 60% of its peak value and the onset of expiratory flow. Mean EAdi, [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{EA}}_{{\text{di}}_{{ \min }} } , $$\end{document}][\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{EA}}_{{\text{di}}_{{ \max }} } $$\end{document}] and the corresponding pressure time products were calculated for the unassisted period of inspiration [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{(EA}}_{{\text{di}}_{{\text{TR}}} } ,{\kern 1pt} \;{\text{PTPEA}}_{{\text{DI}}_{{\text{TR}}} } ) $$\end{document}] and for the entire neural inspiration [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{(EA}}_{{\text{di}}_{{\text{insp}}} } ,{\kern 1pt} \;{\text{PTPEA}}_{{\text{di}}_{{\text{Insp}}} } ). $$\end{document}] The peak EAdi was calculated for each breath and was expressed as percentage of the maximum EAdi obtained during the maximum inspiratory maneuvers. Mean tidal excursion of PesPes) and the pressure time product for the Pes (PTPes) were calculated for the unassisted pre-trigger inspiratory phase [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ (P_{{\text{es}}_{{\text{tr}}} } {\text{ and PTP}}_{{\text{es}}_{{\text{tr}}} } ) $$\end{document}] and for the entire neural inspiratory period [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ (P_{{\text{es}}_{{\text{tot}}} } {\text{ and PTP}}_{{\text{es}}_{{\text{tot}}} } ). $$\end{document}]

The number of wasted inspiratory efforts, i.e., failure to initiate PSV in the presence of a neural inspiratory effort is expressed as percentage of all neural efforts within the same time period.

For successfully triggered breaths, asynchrony was calculated over the whole breathing cycle on a breath by breath basis [24, 25] and was presented as a percentage of the duration of the neural breath (neural Ttot). Wasted efforts were counted as 100% asynchrony.

Statistical analysis

Data are presented as median and 25th and 75th percentiles if not stated otherwise. Analysis was performed using non-parametric repeated measures analysis [26], corrected for multiple testing by Bonferroni correction and pair-wise comparisons (Wilcoxon test). A p value <0.05 was considered to be significant. Correlation between comfort of breathing and the percentage of asynchrony was calculated using the Spearman’s R statistic (see ESM for details).


Results

All subjects successfully followed the breathing instructions and reached the targeted breathing frequency, i.e., there was neither significant difference between the neural breathing frequency during Ptr and Ntr at the same PSV level nor was there a difference in the neural inspiratory times (Table 1). [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{EA}}_{{\text{di}}_{{ \min }} } $$\end{document}] (Ntr: 26.0, 24.1–28.3; Ptr: 26.6, 24.8–28.8, p = 0.08) and [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{EA}}_{{\text{di}}_{{ \max }} } $$\end{document}] (Ntr: 64.0, 35.5–92.4; Ptr: 69.7, 47.1–96.5, p = 0.43) were comparable during Ptr and Ntr. The EAdi levels during inspiration were maintained at 9.5% (5.2–12.6) and 10.1% (6.1–16.0) of the [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{EA}}_{{\text{di}}_{{ \max }} } $$\end{document}] during Ntr and Ptr, respectively. There was no difference in [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{EA}}_{{\text{di}}_{{\text{insp}}} } $$\end{document}] between Ntr and Ptr at PSV levels of 5 cm H2O (p = 0.74) and 10 cm H2O (p = 0.33), whereas at PSV levels of 20 cm H2O the height of [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{EA}}_{{\text{di}}_{{\text{insp}}} } $$\end{document}] differed significantly (p = 0.0098) Thus, overall [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTPEA}}_{{\text{di}}_{{\text{insp}}} } $$\end{document}] was slightly higher during Ptr (Ntr: 38.6, 22.5–52.9; Ptr: 43.1, 28.5–59.9, p = 0.045) (Table 1).

Subject–ventilator asynchrony

Figure 1 depicts tracings of EAdi and ventilator assist during neural and pneumatic triggering and cycling-off of NIV with the helmet in one healthy subject. Figure 2 shows differences between the delays during pneumatic and neural triggering. Ptr was delayed relative to Ntr (p < 0.001, Table 1), and the difference between the neural and pneumatic trigger delays increased (p < 0.001) as the level of PSV and RR increased (Fig. 2). During pneumatic triggering, the number of wasted inspiratory efforts increased at higher respiratory rates and with higher pressure support levels (Table 1). At a respiratory rate of 10 bpm, no wasted inspiratory efforts occurred regardless of the level of PSV, whereas at a PSV of 20 cm H2O, at a respiratory rate of 30 bpm, there was a significant increase of wasted efforts to 52.0% (29.3–58.6; p = 0.001) of all neural inspiratory efforts. No wasted inspiratory efforts occurred during neural control of NIV.

The expiratory delay for all combinations of PSV levels and RR during Poff was longer than during Noff (p < 0.001) (Table 1), and the expiratory delays during Poff became longer with increasing PSV levels and target RR compared to Noff (p < 0.001) as depicted in Fig. 3.

Overall, the subject–ventilator synchrony was progressively impaired with increasing respiratory rate and levels of PSV during pneumatic triggering and cycling-off as depicted in Fig. 4, (p < 0.001). In contrast, increasing respiratory rate and levels of assists only had negligible influence on the asynchrony during neural triggering and cycling-off (Fig. 4).

Unassisted “pre-trigger” inspiratory effort during PSV

Depending on the combination between PSV and target RR, the efforts to trigger the ventilator were 5–60 times higher during Ptr than during Ntr. Median [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTPEA}}_{{\text{di}}_{{\text{Tr}}} } $$\end{document}] during Ntr was 3.4 (2.7–4.2) compared to 16.9 (10.2–26.6) during Ptr (p < 0.001). For all combinations, the [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTP}}_{{\text{es}}_{{\text{tr}}} } $$\end{document}] during Ptr was −264 (−470 to −130) cm H20 ms per breath compared to −11.7 (−30.6 to −6.2; p < 0.001) during Ntr (Table 1). The median [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTP}}_{{\text{es}}_{{\text{tr}}} } $$\end{document}] expressed as percent of [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTP}}_{{\text{es}}_{{\text{tot}}} } $$\end{document}] was 21.5% (11.1–23.5) during Ptr and significantly lower during Ntr, 1.4% (0.9–1.6) (p < 0.001).

Comfort of breathing

Breathing comfort was lower during Ptr than during Ntr in 84.4% of all runs (p < 0.001) (Fig. 5). Overall, comfort of breathing was more than two times worse during pneumatic triggering compared to neural triggering [Ptr median 43.0 mm (20.2–75.6) vs. Ntr 19.2 mm (10.1–28.5)]. Increasing the respiratory rate resulted in minor changes of breathing comfort during neural triggering (p = 0.003), while there was a reduction in breathing comfort during pneumatic triggering (p = 0.953). Increasing PSV at a given respiratory rate tended to decrease breathing comfort during both neural triggering and pneumatic triggering, although the reduction in comfort tended to be larger with pneumatic triggering (Fig. 5). During pneumatic triggering, comfort of breathing was significantly correlated with the amount of asynchrony (r = 0.59, p < 0.001), whereas no such correlation was observed during neural triggering (r = −0.146, p = 0.253).


Discussion

The present study demonstrates that the airway pressure generated by healthy subjects using ~10% of their maximal EAdi was less efficient to pneumatically control PSV with the helmet interface at increasing breathing frequencies and pressure support levels compared to a neural trigger and cycling-off algorithm using EAdi. Due to the trigger asynchrony during Ptr, also the neural and mechanical efforts during Ptr became several times higher than during Ntr. This suggests that during NIV with pneumatic triggering, it is necessary to force the ventilator into synchrony by increasing inspiratory effort, which would defeat the purpose of providing ventilatory assist.

Previous studies comparing the helmet to the face mask interface during pneumatically triggered and cycled-off PSV show that the helmet is less effective in unloading the respiratory muscles, which was partially explained by inspiratory trigger delays and the impaired pressurization rate [1618].

The difference in trigger delays between Ntr and Ptr could not be due to mechanical response of the ventilator since NIV was delivered with the same ventilator throughout the study. Hence, apart from changes in trigger delays, no changes in raised time or pressure delivery could occur.

A critique on the present study was that a pressure trigger of −1.3 ± 0.6 cm H2O was required to avoid auto-triggering. This likely increased trigger delays and the fraction of the pressure time product necessary to pneumatically trigger the ventilator [27]. In the absence of leaks, studies suggest that flow triggering is more efficient than pressure triggering [28, 29]. However, there are also results in favor of pressure triggering [30]. The reported improvements of delays between flow and pressure triggering (40–43) are not of a magnitude that can match the improvement of implementing neural trigger relative to pressure trigger using a helmet interface in the present study. In fact, the trigger delays during Ptr were similar to those observed with helmet interface in previous studies, whereas the trigger delays during Ntr in the present study are within the range of those reported with face mask in previous studies [1618]. Since the helmet interface is presumed to be frequently associated with leaks, which makes triggering and cycling very complex [31], it is questionable whether flow triggering is recommendable. It should be noted that in the present study, increasing breathing frequency and assist levels only increased trigger delays and [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTP}}_{{\text{es}}_{{\text{tr}}} } $$\end{document}] during Ptr and not during Ntr.

With regard to wasted inspiratory efforts, which are the worst type of trigger asynchrony, a previous study did not report wasted efforts during NIV (5 cm H2O) with helmet and face mask in a study with healthy volunteers who kept their breathing frequency around 15 bpm [12]. In a lung model study using the helmet interface, wasted inspiratory efforts occurred at a respiratory rate above 20 bpm (PSV level of 18 cm H2O) and were negatively influenced by increasing PEEP and PSV levels [17]. In the present study, where both inspiratory and expiratory efforts were monitored and restricted, wasted inspiratory efforts prevailed at highest PSV (20 cm H2O) or RR (30 bpm) levels.

Looking at the baseline characteristics of patients requiring non-invasive ventilatory support due to ARF respiratory rates above 30 would be expected [3234]. Thus, compared to our results where RR did not exceed 30 bpm, the amount of wasted efforts might be even higher. With regard to Ntr, increasing frequencies above 30 bpm should not increase the risk of wasted inspiratory efforts.

In terms of cycling-off the PSV, the ventilator used in the present study included a fixed cycling-off algorithm that terminates assist when flow has dropped to 5% of peak inspiratory flow. Given that the addition of the compliant helmet into the respiratory circuit increases the time constant of the total respiratory system (helmet, respiratory circuit and respiratory system) the late pneumatic cycling-off at 5% of peak inspiratory flow is likely not ideal. To match flow cycling-off with neural inspiratory termination, Du et al. [35] demonstrated that prolonging the time constant of the total respiratory system requires that flow cycles-off at a higher percentage of peak inspiratory flow. The need for varying flow cycling-off criteria suggested to match neural breath termination has been demonstrated in obstructive patients where the flow cycling-off had to take place at about 50% of peak inspiratory flow [36], whereas 5% appears sufficient in patients with restrictive lung disease [37]. However, since the time constant of the respiratory system (and the helmet) changes as the pressure assist level and/or breathing frequency change—and that there is currently no standard to what percentages of peak flow should be used to cycle-off at various levels of assist—the authors chose to let the cycling-off criteria be dictated by the ventilator used.

Since the present study compared different levels of PSV and breathing frequency and that the results are presented as the difference in cycling-off delays between pneumatic and neural cycling-off at the same level of PSV and RR, our findings should be adequate in terms of comparing changes in time delays for cycling-off during Poff compared to Noff with increasing respiratory rates and PSV levels. It should be noted that the findings of increasing cycling-off delays during increasing levels of pneumatically controlled PSV observed in the present study agree with previous studies on PSV in intubated patients [27, 38]. The findings of the present study substantiate that neural off-cycling with EAdi can reduce the problem of excessive prolongation of assist into neural exhalation and its associated influence on breathing pattern [24, 38].

As indicated by the asynchrony percentage as high as 30% during pneumatically triggered and cycled NIV, which was reduced to 5% during neurally triggered and cycled-off NIV, the advantage of the latter appears undisputable with regard to helmet ventilation. Our data confirm those of Racca et al. [39] that wasted efforts and impaired trigger synchrony are likely due to the properties of the helmet itself, being a collapsible device that will dampen the transmission of pressures delivered to the patient as well as reduce the ability to sense the pressure generated by the patient. Thus, the present study suggests that neural triggering can achieve an important reduction in the trigger effort with a helmet device.

In the present study, breathing comfort was rated closer to maximal comfort when PSV and RR were low, which was in contrast to observations in patients with ARF. Vitacca et al. [40] reported that the highest level of comfort in patients on NIV with acute exacerbation of COPD occurred at a PSV level of 17 + 6 cm H2O and a breathing frequency of 18 + 6 bpm.

During neurally triggered and cycled-off PSV, our findings showed that comfort decreased with increasing PSV levels at similar breathing frequency despite no changes in subject–ventilator asynchrony was observed. Thus, it is reasonable to assume that in healthy subjects “no assist” equals maximal comfort and increased PSV is associated with a decrease in comfort. Hence, one should be careful in interpreting the findings of the present study in relation to a patient population with respiratory failure.

However, as evidenced by the strong correlations found between asynchrony and comfort during pneumatically triggered and cycled-off NIV, which was abolished during neurally controlled NIV, subject–ventilator asynchrony plays an important role in the perception of breathing comfort. It should be noted that adding a nasogastric tube is uncomfortable. However, given the complications of endotracheal intubation, it appears reasonable to examine the possible advantage of trading tracheal for esophageal invasiveness, using non-invasive interfaces.


Conclusion

The present study demonstrates in healthy subjects that subject–ventilator synchrony, trigger effort, and breathing comfort with a helmet interface is considerably less impaired during increasing levels of PSV and respiratory rates with neural triggering and cycling-off, compared to conventional pneumatic triggering and cycling-off.


Electronic supplementary material

Below is the link to the electronic supplementary material.

Click here for additional data file (134_2008_1163_MOESM1_ESM.doc)


Notes

This article is discussed in the editorial available at: doi:10.1007/s00134-008-1164-y.

Electronic supplementary material

The online version of this article (doi:10.1007/s00134-008-1163-z) contains supplementary material, which is available to authorized users.

We are indebted to the volunteers for their participation, to Norman Comtois for his invaluable assistance, and to Orla Smith, RN and her team for the coordination given during the study. Statistical analysis of the data was performed after advisory service of the Department of Medical Statistics, University of Göttingen, Germany, with special gratitude to Dr. E. Kahler. This study was financed by departmental funds from the Department of Critical Care Medicine, St. Michael’s Hospital, University of Toronto, Canada and the Department of Anesthesiology, Emergency and Critical Care Medicine, University of Göttingen, Germany. Onnen Moerer holds a research grant from a research program, Faculty of Medicine, Georg-August-University Göttingen. Lukas Brander hold postdoctoral fellowships from the Swiss Foundation for Fellowships in Medicine and Biology (SSMBS) provided by Novartis AG and from the Division of Respirology at the University of Toronto provided by Merck-Frosst. Dr. Sinderby was financially supported by “R. Samuel McLaughlin Foundation”.

Conflict of interest Dr. Beck and Dr. Sinderby have made inventions related to neural control of mechanical ventilation that are patented. The license for these patents belongs to Maquet Critical Care. Future commercial uses of this technology may provide financial benefit to Dr. Beck and Dr. Sinderby through royalties. Dr. Beck and Dr. Sinderby each own 50% of Neurovent Research Inc (NVR). NVR is a research and development company that builds the equipment and catheters for research studies. NVR has a consulting agreement with Maquet Critical Care. Dr. Slutsky consults for companies that make ventilators, specifically, Maquet Critical Care and Hamilton Medical, and is compensated for these consultations St. Michael's Hospital will get a small royalty if Maquet includes SMH-patented discoveries in the ventilator.

Open Access This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.


References
1.. Ferrer M,Valencia M,Nicolas JM,Bernadich O,Badia JR,Torres A. Early noninvasive ventilation averts extubation failure in patients at risk: a randomized trialAm J Respir Crit Care Med 2006;173:164–170. [doi: 10.1164/rccm.200505-718OC]
1.. Ferrer M, Valencia M, Nicolas JM, Bernadich O, Badia JR, Torres A (2006) Early noninvasive ventilation averts extubation failure in patients at risk: a randomized trial. Am J Respir Crit Care Med 173:164–170 [pmid: 16224108]
2.. Ferrer M,Esquinas A,Arancibia F,Bauer TT,Gonzalez G,Carrillo A,Rodriguez-Roisin R,Torres A. Noninvasive ventilation during persistent weaning failure: a randomized controlled trialAm J Respir Crit Care Med 2003;168:70–76. [doi: 10.1164/rccm.200209-1074OC]
2.. Ferrer M, Esquinas A, Arancibia F, Bauer TT, Gonzalez G, Carrillo A, Rodriguez-Roisin R, Torres A (2003) Noninvasive ventilation during persistent weaning failure: a randomized controlled trial. Am J Respir Crit Care Med 168:70–76 [pmid: 12689847]
3.. Carlucci A,Richard JC,Wysocki M,Lepage E,Brochard L. Noninvasive versus conventional mechanical ventilation. An epidemiologic surveyAm J Respir Crit Care Med 2001;163:874–880.
3.. Carlucci A, Richard JC, Wysocki M, Lepage E, Brochard L (2001) Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. Am J Respir Crit Care Med 163:874–880 [pmid: 11282759]
4.. Brochard L,Mancebo J,Wysocki M,Lofaso F,Conti G,Rauss A,Simonneau G,Benito S,Gasparetto A,Lemaire F. Noninvasive ventilation for acute exacerbations of chronic obstructive pulmonary diseaseN Engl J Med 1995;333:817–822. [doi: 10.1056/NEJM199509283331301]
4.. Brochard L, Mancebo J, Wysocki M, Lofaso F, Conti G, Rauss A, Simonneau G, Benito S, Gasparetto A, Lemaire F (1995) Noninvasive ventilation for acute exacerbations of chronic obstructive pulmonary disease. N Engl J Med 333:817–822 [pmid: 7651472]
5.. Ram FS,Picot J,Lightowler J,Wedzicha JA. Non-invasive positive pressure ventilation for treatment of respiratory failure due to exacerbations of chronic obstructive pulmonary diseaseCochrane Database Syst Rev 2004;3:CD004104.
5.. Ram FS, Picot J, Lightowler J, Wedzicha JA (2004) Non-invasive positive pressure ventilation for treatment of respiratory failure due to exacerbations of chronic obstructive pulmonary disease. Cochrane Database Syst Rev 3:CD004104 [pmid: 15266518]
6.. Winck JC,Azevedo LF,Costa-Pereira A,Antonelli M,Wyatt JC. Efficacy and safety of non-invasive ventilation in the treatment of acute cardiogenic pulmonary edema—a systematic review and meta-analysisCrit Care 2006;10:R69. [doi: 10.1186/cc4905]
6.. Winck JC, Azevedo LF, Costa-Pereira A, Antonelli M, Wyatt JC (2006) Efficacy and safety of non-invasive ventilation in the treatment of acute cardiogenic pulmonary edema—a systematic review and meta-analysis. Crit Care 10:R69 [pmid: 16646987]
7.. Conti G,Antonelli M,Navalesi P,Rocco M,Bufi M,Spadetta G,Meduri GU. Noninvasive vs. conventional mechanical ventilation in patients with chronic obstructive pulmonary disease after failure of medical treatment in the ward: a randomized trialIntensive Care Med 2002;28:1701–1707. [doi: 10.1007/s00134-002-1478-0]
7.. Conti G, Antonelli M, Navalesi P, Rocco M, Bufi M, Spadetta G, Meduri GU (2002) Noninvasive vs. conventional mechanical ventilation in patients with chronic obstructive pulmonary disease after failure of medical treatment in the ward: a randomized trial. Intensive Care Med 28:1701–1707 [pmid: 12447511]
8.. Navalesi P,Fanfulla F,Frigerio P,Gregoretti C,Nava S. Physiologic evaluation of noninvasive mechanical ventilation delivered with three types of masks in patients with chronic hypercapnic respiratory failureCrit Care Med 2000;28:1785–1790. [doi: 10.1097/00003246-200006000-00015]
8.. Navalesi P, Fanfulla F, Frigerio P, Gregoretti C, Nava S (2000) Physiologic evaluation of noninvasive mechanical ventilation delivered with three types of masks in patients with chronic hypercapnic respiratory failure. Crit Care Med 28:1785–1790 [pmid: 10890620]
9.. Kramer N,Meyer TJ,Meharg J,Cece RD,Hill NS. Randomized, prospective trial of noninvasive positive pressure ventilation in acute respiratory failureAm J Respir Crit Care Med 1995;151:1799–1806.
9.. Kramer N, Meyer TJ, Meharg J, Cece RD, Hill NS (1995) Randomized, prospective trial of noninvasive positive pressure ventilation in acute respiratory failure. Am J Respir Crit Care Med 151:1799–1806 [pmid: 7767523]
10.. Gregoretti C,Confalonieri M,Navalesi P,Squadrone V,Frigerio P,Beltrame F,Carbone G,Conti G,Gamna F,Nava S,Calderini E,Skrobik Y,Antonelli M. Evaluation of patient skin breakdown and comfort with a new face mask for non-invasive ventilation: a multi-center studyIntensive Care Med 2002;28:278–284. [doi: 10.1007/s00134-002-1208-7]
10.. Gregoretti C, Confalonieri M, Navalesi P, Squadrone V, Frigerio P, Beltrame F, Carbone G, Conti G, Gamna F, Nava S, Calderini E, Skrobik Y, Antonelli M (2002) Evaluation of patient skin breakdown and comfort with a new face mask for non-invasive ventilation: a multi-center study. Intensive Care Med 28:278–284 [pmid: 11904656]
11.. Antonelli M,Pennisi MA,Conti G. New advances in the use of noninvasive ventilation for acute hypoxaemic respiratory failureEur Respir J Suppl 2003;42:65s–71s. [doi: 10.1183/09031936.03.00421003]
11.. Antonelli M, Pennisi MA, Conti G (2003) New advances in the use of noninvasive ventilation for acute hypoxaemic respiratory failure. Eur Respir J Suppl 42:65s–71s [pmid: 12946003]
12.. Chiumello D,Pelosi P,Carlesso E,Severgnini P,Aspesi M,Gamberoni C,Antonelli M,Conti G,Chiaranda M,Gattinoni L. Noninvasive positive pressure ventilation delivered by helmet vs. standard face maskIntensive Care Med 2003;29:1671–1679. [doi: 10.1007/s00134-003-1825-9]
12.. Chiumello D, Pelosi P, Carlesso E, Severgnini P, Aspesi M, Gamberoni C, Antonelli M, Conti G, Chiaranda M, Gattinoni L (2003) Noninvasive positive pressure ventilation delivered by helmet vs. standard face mask. Intensive Care Med 29:1671–1679 [pmid: 12802491]
13.. Antonelli M,Conti G,Pelosi P,Gregoretti C,Pennisi MA,Costa R,Severgnini P,Chiaranda M,Proietti R. New treatment of acute hypoxemic respiratory failure: noninvasive pressure support ventilation delivered by helmet—a pilot controlled trialCrit Care Med 2002;30:602–608. [doi: 10.1097/00003246-200203000-00019]
13.. Antonelli M, Conti G, Pelosi P, Gregoretti C, Pennisi MA, Costa R, Severgnini P, Chiaranda M, Proietti R (2002) New treatment of acute hypoxemic respiratory failure: noninvasive pressure support ventilation delivered by helmet—a pilot controlled trial. Crit Care Med 30:602–608 [pmid: 11990923]
14.. Pelosi P,Severgnini P,Aspesi M,Gamberoni C,Chiumello D,Fachinetti C,Introzzi L,Antonelli M,Chiaranda M. Non-invasive ventilation delivered by conventional interfaces and helmet in the emergency departmentEur J Emerg Med 2003;10:79–86. [doi: 10.1097/00063110-200306000-00002]
14.. Pelosi P, Severgnini P, Aspesi M, Gamberoni C, Chiumello D, Fachinetti C, Introzzi L, Antonelli M, Chiaranda M (2003) Non-invasive ventilation delivered by conventional interfaces and helmet in the emergency department. Eur J Emerg Med 10:79–86 [pmid: 12789060]
15.. Patroniti N,Foti G,Manfio A,Coppo A,Bellani G,Pesenti A. Head helmet versus face mask for non-invasive continuous positive airway pressure: a physiological studyIntensive Care Med 2003;29:1680–1687. [doi: 10.1007/s00134-003-1931-8]
15.. Patroniti N, Foti G, Manfio A, Coppo A, Bellani G, Pesenti A (2003) Head helmet versus face mask for non-invasive continuous positive airway pressure: a physiological study. Intensive Care Med 29:1680–1687 [pmid: 14564379]
16.. Navalesi P,Costa R,Ceriana P,Carlucci A,Prinianakis G,Antonelli M,Conti G,Nava S. Non-invasive ventilation in chronic obstructive pulmonary disease patients: helmet versus facial maskIntensive Care Med 2006;33:74–81. [doi: 10.1007/s00134-006-0391-3]
16.. Navalesi P, Costa R, Ceriana P, Carlucci A, Prinianakis G, Antonelli M, Conti G, Nava S (2006) Non-invasive ventilation in chronic obstructive pulmonary disease patients: helmet versus facial mask. Intensive Care Med 33:74–81 [pmid: 17039354]
17.. Moerer O,Fischer S,Hartelt M,Kuvaki B,Quintel M,Neumann P. Influence of two different interfaces for noninvasive ventilation compared to invasive ventilation on the mechanical properties and performance of a respiratory system: a lung model studyChest 2006;129:1424–1431. [doi: 10.1378/chest.129.6.1424]
17.. Moerer O, Fischer S, Hartelt M, Kuvaki B, Quintel M, Neumann P (2006) Influence of two different interfaces for noninvasive ventilation compared to invasive ventilation on the mechanical properties and performance of a respiratory system: a lung model study. Chest 129:1424–1431 [pmid: 16778258]
18.. Racca F,Appendini L,Gregoretti C,Stra E,Patessio A,Donner CF,Ranieri VM. Effectiveness of mask and helmet interfaces to deliver noninvasive ventilation in a human model of resistive breathingJ Appl Physiol 2005;99:1262–1271. [doi: 10.1152/japplphysiol.01363.2004]
18.. Racca F, Appendini L, Gregoretti C, Stra E, Patessio A, Donner CF, Ranieri VM (2005) Effectiveness of mask and helmet interfaces to deliver noninvasive ventilation in a human model of resistive breathing. J Appl Physiol 99:1262–1271 [pmid: 15961605]
19.. Sinderby C,Navalesi P,Beck J,Skrobik Y,Comtois N,Friberg S,Gottfried SB,Lindstrom L. Neural control of mechanical ventilation in respiratory failureNat Med 1999;5:1433–1436. [doi: 10.1038/71012]
19.. Sinderby C, Navalesi P, Beck J, Skrobik Y, Comtois N, Friberg S, Gottfried SB, Lindstrom L (1999) Neural control of mechanical ventilation in respiratory failure. Nat Med 5:1433–1436 [pmid: 10581089]
20.. Moerer O,Beck J,Brander L,Quintel M,Brunet F,Sinderby C. Subject–ventilator synchrony during neural vs. pressure triggered non-invasive helmet ventilationIntensive Care Med 2006;32:148.
20.. Moerer O, Beck J, Brander L, Quintel M, Brunet F, Sinderby C (2006) Subject–ventilator synchrony during neural vs. pressure triggered non-invasive helmet ventilation. Intensive Care Med 32:148
21.. Sinderby CA,Beck JC,Lindstrom LH,Grassino AE. Enhancement of signal quality in esophageal recordings of diaphragm EMGJ Appl Physiol 1997;82:1370–1377.
21.. Sinderby CA, Beck JC, Lindstrom LH, Grassino AE (1997) Enhancement of signal quality in esophageal recordings of diaphragm EMG. J Appl Physiol 82:1370–1377 [pmid: 9104877]
22.. Benditt JO. Esophageal and gastric pressure measurementsRespir Care 2005;50:68–75.
22.. Benditt JO (2005) Esophageal and gastric pressure measurements. Respir Care 50:68–75 [pmid: 15636646]
23.. Sinderby C,Beck J,Spahija J,Weinberg J,Grassino A. Voluntary activation of the human diaphragm in health and diseaseJ Appl Physiol 1998;85:2146–2158.
23.. Sinderby C, Beck J, Spahija J, Weinberg J, Grassino A (1998) Voluntary activation of the human diaphragm in health and disease. J Appl Physiol 85:2146–2158 [pmid: 9843538]
24.. Beck J,Tucci M,Emeriaud G,Lacroix J,Sinderby C. Prolonged neural expiratory time induced by mechanical ventilation in infantsPediatr Res 2004;55:747–754. [doi: 10.1203/01.PDR.0000119368.21770.33]
24.. Beck J, Tucci M, Emeriaud G, Lacroix J, Sinderby C (2004) Prolonged neural expiratory time induced by mechanical ventilation in infants. Pediatr Res 55:747–754 [pmid: 14739354]
25.. Beck J,Campoccia F,Allo JC,Brander L,Brunet F,Slutsky AS,Sinderby C. Improved synchrony and respiratory unloading by neurally adjusted ventilatory assist (NAVA) in lung-injured rabbitsPediatr Res 2007;61:289–294. [doi: 10.1203/01.pdr.0000257324.22406.93]
25.. Beck J, Campoccia F, Allo JC, Brander L, Brunet F, Slutsky AS, Sinderby C (2007) Improved synchrony and respiratory unloading by neurally adjusted ventilatory assist (NAVA) in lung-injured rabbits. Pediatr Res 61:289–294 [pmid: 17314685]
26.. Brunner E,Domhof S,Langer F. Nonparametric analysis of longitudinal data in factorial experiments. 2002New York: Wiley;
26.. Brunner E, Domhof S, Langer F (2002) Nonparametric analysis of longitudinal data in factorial experiments. Wiley, New York
27.. Aslanian P,el Atrous S,Isabey D,Valente E,Corsi D,Harf A,Lemaire F,Brochard L. Effects of flow triggering on breathing effort during partial ventilatory supportAm J Respir Crit Care Med 1998;157:135–143.
27.. Aslanian P, el Atrous S, Isabey D, Valente E, Corsi D, Harf A, Lemaire F, Brochard L (1998) Effects of flow triggering on breathing effort during partial ventilatory support. Am J Respir Crit Care Med 157:135–143 [pmid: 9445291]
28.. Giuliani R,Mascia L,Recchia F,Caracciolo A,Fiore T,Ranieri VM. Patient–ventilator interaction during synchronized intermittent mandatory ventilation. Effects of flow triggeringAm J Respir Crit Care Med 1995;151:1–9.
28.. Giuliani R, Mascia L, Recchia F, Caracciolo A, Fiore T, Ranieri VM (1995) Patient–ventilator interaction during synchronized intermittent mandatory ventilation. Effects of flow triggering. Am J Respir Crit Care Med 151:1–9 [pmid: 7812537]
29.. Nava S,Ambrosino N,Bruschi C,Confalonieri M,Rampulla C. Physiological effects of flow and pressure triggering during non-invasive mechanical ventilation in patients with chronic obstructive pulmonary diseaseThorax 1997;52:249–254.
29.. Nava S, Ambrosino N, Bruschi C, Confalonieri M, Rampulla C (1997) Physiological effects of flow and pressure triggering during non-invasive mechanical ventilation in patients with chronic obstructive pulmonary disease. Thorax 52:249–254 [pmid: 9093341]
30.. Goulet R,Hess D,Kacmarek RM. Pressure vs flow triggering during pressure support ventilationChest 1997;111:1649–1653. [doi: 10.1378/chest.111.6.1649]
30.. Goulet R, Hess D, Kacmarek RM (1997) Pressure vs flow triggering during pressure support ventilation. Chest 111:1649–1653 [pmid: 9187188]
31.. Calderini E,Confalonieri M,Puccio PG,Francavilla N,Stella L,Gregoretti C. Patient–ventilator asynchrony during noninvasive ventilation: the role of expiratory triggerIntensive Care Med 1999;25:662–667. [doi: 10.1007/s001340050927]
31.. Calderini E, Confalonieri M, Puccio PG, Francavilla N, Stella L, Gregoretti C (1999) Patient–ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger. Intensive Care Med 25:662–667 [pmid: 10470568]
32.. Ray P,Birolleau S,Lefort Y,Becquemin MH,Beigelman C,Isnard R,Teixeira A,Arthaud M,Riou B,Boddaert J. Acute respiratory failure in the elderly: etiology, emergency diagnosis and prognosisCrit Care 2006;10:R82. [doi: 10.1186/cc4926]
32.. Ray P, Birolleau S, Lefort Y, Becquemin MH, Beigelman C, Isnard R, Teixeira A, Arthaud M, Riou B, Boddaert J (2006) Acute respiratory failure in the elderly: etiology, emergency diagnosis and prognosis. Crit Care 10:R82 [pmid: 16723034]
33.. Honrubia T,Garcia Lopez FJ,Franco N,Mas M,Guevara M,Daguerre M,Alia I,Algora A,Galdos P. Noninvasive vs conventional mechanical ventilation in acute respiratory failure: a multicenter, randomized controlled trialChest 2005;128:3916–3924. [doi: 10.1378/chest.128.6.3916]
33.. Honrubia T, Garcia Lopez FJ, Franco N, Mas M, Guevara M, Daguerre M, Alia I, Algora A, Galdos P (2005) Noninvasive vs conventional mechanical ventilation in acute respiratory failure: a multicenter, randomized controlled trial. Chest 128:3916–3924 [pmid: 16354864]
34.. Phua J,Kong K,Lee KH,Shen L,Lim TK. Noninvasive ventilation in hypercapnic acute respiratory failure due to chronic obstructive pulmonary disease vs. other conditions: effectiveness and predictors of failureIntensive Care Med 2005;31:533–539. [doi: 10.1007/s00134-005-2582-8]
34.. Phua J, Kong K, Lee KH, Shen L, Lim TK (2005) Noninvasive ventilation in hypercapnic acute respiratory failure due to chronic obstructive pulmonary disease vs. other conditions: effectiveness and predictors of failure. Intensive Care Med 31:533–539 [pmid: 15742175]
35.. Du HL,Amato MB,Yamada Y. Automation of expiratory trigger sensitivity in pressure support ventilationRespir Care Clin N Am 2001;7:503–517. [doi: 10.1016/S1078-5337(05)70046-0]
35.. Du HL, Amato MB, Yamada Y (2001) Automation of expiratory trigger sensitivity in pressure support ventilation. Respir Care Clin N Am 7:503–517 [pmid: 11517036]
36.. Tassaux D,Gainnier M,Battisti A,Jolliet P. Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workloadAm J Respir Crit Care Med 2005;172:1283–1289. [doi: 10.1164/rccm.200407-880OC]
36.. Tassaux D, Gainnier M, Battisti A, Jolliet P (2005) Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med 172:1283–1289 [pmid: 16109983]
37.. Tokioka H,Tanaka T,Ishizu T,Fukushima T,Iwaki T,Nakamura Y,Kosogabe Y. The effect of breath termination criterion on breathing patterns and the work of breathing during pressure support ventilationAnesth Analg 2001;92:161–165. [doi: 10.1097/00000539-200101000-00031]
37.. Tokioka H, Tanaka T, Ishizu T, Fukushima T, Iwaki T, Nakamura Y, Kosogabe Y (2001) The effect of breath termination criterion on breathing patterns and the work of breathing during pressure support ventilation. Anesth Analg 92:161–165 [pmid: 11133620]
38.. Younes M,Kun J,Webster K,Roberts D. Response of ventilator-dependent patients to delayed opening of exhalation valveAm J Respir Crit Care Med 2002;166:21–30. [doi: 10.1164/rccm.2107143]
38.. Younes M, Kun J, Webster K, Roberts D (2002) Response of ventilator-dependent patients to delayed opening of exhalation valve. Am J Respir Crit Care Med 166:21–30 [pmid: 12091166]
39.. Racca F,Appendini L,Gregoretti C,Stra E,Patessio A,Donner CF,Ranieri VM. Effectiveness of mask and helmet interfaces to deliver noninvasive ventilation in a human model of resistive breathingJ Appl Physiol 2005;99(4):1262–1271. [doi: 10.1152/japplphysiol.01363.2004]
39.. Racca F, Appendini L, Gregoretti C, Stra E, Patessio A, Donner CF, Ranieri VM (2005) Effectiveness of mask and helmet interfaces to deliver noninvasive ventilation in a human model of resistive breathing. J Appl Physiol 99(4):1262–1271 [pmid: 15961605]
40.. Vitacca M,Bianchi L,Zanotti E,Vianello A,Barbano L,Porta R,Clini E. Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilationChest 2004;126:851–859. [doi: 10.1378/chest.126.3.851]
40.. Vitacca M, Bianchi L, Zanotti E, Vianello A, Barbano L, Porta R, Clini E (2004) Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation. Chest 126:851–859 [pmid: 15364766]

Figures

[Figure ID: Fig1]
Fig. 1 

Example of diaphragm electrical activity (EAdi), and ventilatory assist during pneumatically (left panel) and neurally (right panel) triggered and cycled-off NIV in one healthy subject breathing with the helmet interface. In this example, respiratory rate was 30 bpm, PSV level was 20 cm H2O. The dashed line shows the start of neural inspiration and the continuous line shows the end of neural inspiration. Note excessive asynchrony during pneumatically triggered and cycled-off NIV



[Figure ID: Fig2]
Fig. 2 

Differences in inspiratory delays (delay time Ptr − delay time Ntr) during NIV with a pressure support of 5, 10 and 20 cm H2O and respiratory rates of 10, 20 and 30 bpm were shown. During all combinations of respiratory rates and pressure support levels, there was a significant difference in inspiratory delays being highly increased during pneumatically triggered NIV. Symbols represent group median values and the bars indicate 25th and 75th percentiles. Delay-on inspiratory delay between the onset of the volunteers’ inspiratory effort and the start of the ventilatory support, Ptr pneumatic trigger, Ntr neural trigger, NIV non-invasive ventilation



[Figure ID: Fig3]
Fig. 3 

Differences in expiratory delays (delay time Poff − delay time Noff) during NIV with PSV of 5, 10 and 20 cm H2O and respiratory rates of 10, 20 and 30 bpm were shown. During all combinations of respiratory rates and pressure support levels, there was a significant difference in expiratory delays that were highly increased during pneumatically triggered NIV. Symbols represent group median values and the bars indicate 25th and 75th percentiles. Delay-off delay between the onset neural end of inspiration and the end of the ventilatory support, Poff pneumatic cycling-off, Noff neural cycling-off, NIV non-invasive ventilation



[Figure ID: Fig4]
Fig. 4 

Percentage of asynchrony over the whole breath duration as calculated during Ntr and Ptr NIV: (delay-on + delay-off/neural Ttot × 100). Wasted inspiratory efforts were counted as 100% asynchrony. Percentage of asynchrony during Ptr was markedly higher compared to Ntr. Symbols represent group median values and the bars indicate 25th and 75th percentiles. Ntr neurally triggered, Ptr pneumatically triggered, NIV non-invasive ventilation, PSV pressure support ventilation



[Figure ID: Fig5]
Fig. 5 

Breathing comfort in mm as assessed by a visual analogue scale during neurally and pneumatically triggered and cycled-off NIV at different levels of assist and respiratory rates (0 very comfortable, 100 unbearable). During all combinations of respiratory rates and PSV levels, comfort of breathing was better during neurally triggered and cycled-off NIV. Symbols represent group median values and the bars indicate 25th and 75th percentiles. NIV non-invasive ventilation, PSV pressure support ventilation, VAS visual analogue scale



Tables
[TableWrap ID: Tab1] Table 1 

Pneumatic-triggered versus neural-triggered NIV at different levels of PSV and respiratory rates


There was no difference between mean measured respiratory rate (RR), neural inspiratory time, total inspiratory diaphragmatic activity [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{(EA}}_{{\text{di}}_{{\text{insp}}} } ),{\text{ EA}}_{{\text{di}}_{{ \min }} } {\text{ and EA}}_{{\text{di}}_{{ \max }} } . $$\end{document}] Inspiratory delay time (delay-on) (p < 0.001), expiratory cycling-off (delay-off) (p < 0.001) [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTPEA}}_{{\text{di}}_{{\text{Tr}}} } $$\end{document}] (p < 0.001), [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{PTPEA}}_{{\text{di}}_{{\text{insp}}} } $$\end{document}] (p = 0.045) and the unassisted inspiratory effort [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{(PTP}}_{{\text{es}}_{{\text{Tr}}} } ) $$\end{document}] (p < 0.001) differed significantly. Wasted inspiratory efforts (WE) (expressed in percent of neural efforts) occurred at higher respiratory rates and levels of pressure support during Ptr while there were none during Ntr. Data are presented as median values (25th and 75th percentiles)

NIV non-invasive ventilation, PSV pressure support ventilation, RR respiratory rate, bpm breaths per minute, Ti neural inspiratory time, Tot total neural inspiratory and expiratory time, EAdi electrical diaphragmatic activity, [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ EA_{di_{\min } } {\text{ and }}EA_{di_{\max } } $$\end{document}] minimal and maximal EAdi during inspiration, [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ EA_{di_{insp} } $$\end{document}] mean EAdi during inspiration, [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ EA_{di_{Tr} } $$\end{document}] mean EAdi during unassisted phase of inspiration, [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ PTPEA_{di_{Tr} } {\text{ and }}PTPEA_{di_{insp} } $$\end{document}] pressure time product of unassisted inspiratory phase and inspiration, delay-on time between onset of EAdi and the onset of ventilatory assist, delay-off time between end of neural Ti defined as reduction to 60% of peak activity and end of ventilator assist, [\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ PTP_{es_{Tr} } $$\end{document}] pressure time product of esophageal pressure during unassisted phase of inspiration, WE wasted inspiratory efforts



Article Categories:
  • Original

Keywords: Keywords Non-invasive ventilation (NIV), Helmet, Neurally triggered ventilatory assist, Neural control, Pressure support ventilation, Trigger.