Document Detail


Statistical design of unicompartmental tibial implants and comparison with current devices.
MedLine Citation:
PMID:  17188876     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
This study defines, in the context of unicompartmental tibial replacement, the medial and lateral resection surfaces of 34 tibiae at a depth of 5 mm below the articular surface. Using statistical techniques, three optimal theoretical size and shape unicompartmental tibial designs, (i) implants of consistent shape in varying size (ii) implants symmetric about their ML axis; (iii) implants of varying size and shape, were constructed to best fit the population. Two currently available commercial implants, the Preservation Uni System (DePuy Orthopaedics Inc., Warsaw, IN) and the LCS Uni System (DePuy Orthopaedics Inc., Warsaw, IN), which were similar to types (i) and (ii) respectively, of the theoretical designs, were also included in the analysis. All implants, commercial and theoretical, were compared with one another to determine which implant designs gave the best cortical bone coverage on both the medial and lateral compartments. Of the commercial implants, the type (i) design fitted best, with an average of 67% of the implant edge lying on cortical bone, compared with 57% for the type (ii) implants. Of the theoretical implants, 72%, 67% and 76% of the implant edge lay on cortical bone for types (i), (ii) and (iii) designs, respectively, indicating that there is room for improvement in current implant designs to achieve better coverage in both the medial and lateral compartments combined.
Authors:
Clare Fitzpatrick; David FitzPatrick; Jordan Lee; Daniel Auger
Related Documents :
21458326 - Super-eba and irm as root-end fillings in periapical surgery with ultrasonic preparatio...
17070056 - Stereolithographic models for biomechanical testing.
23057046 - Stress distribution and micromotion analyses of immediately loaded implants of varying ...
23893176 - A new neosuchian with asian affinities from the jurassic of northeastern brazil.
16283906 - Long-term repetitive sodium lauryl sulfate-induced irritation of the skin: an in vivo s...
24431786 - Comparative evaluation of dimensional accuracy and tensile strength of a type iv gypsum...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2006-12-22
Journal Detail:
Title:  The Knee     Volume:  14     ISSN:  0968-0160     ISO Abbreviation:  Knee     Publication Date:  2007 Mar 
Date Detail:
Created Date:  2007-02-26     Completed Date:  2007-05-11     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9430798     Medline TA:  Knee     Country:  England    
Other Details:
Languages:  eng     Pagination:  138-44     Citation Subset:  IM    
Affiliation:
School of Electrical, Electronic and Mechanical Engineering, University College Dublin, Belfield, Dublin 4, Ireland. clare.fitzpatrick@ucd.ie <clare.fitzpatrick@ucd.ie>
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Arthroplasty, Replacement, Knee
Female
Humans
Image Processing, Computer-Assisted
Knee Joint / radiography*
Knee Prosthesis*
Male
Middle Aged
Models, Biological
Prosthesis Design*
Prosthesis Fitting
Tomography, X-Ray Computed

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  ortho-Substituted azoles as selective and dual inhibitors of VEGF receptors 1 and 2.
Next Document:  Neoligamentization process of BTPB used for ACL graft: histological evaluation from 6 months to 10 y...