Document Detail


Spatially varying longitudinal aliasing and resolution in spiral computed tomography.
MedLine Citation:
PMID:  10619247     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Spiral computed tomography (CT) has revolutionized conventional CT as a truly three-dimensional imaging modality. A number of studies aimed at evaluating the longitudinal resolution in spiral CT have been presented, but the spatially varying nature of the longitudinal resolution in spiral CT has been largely left undiscussed. In this paper, we investigate the longitudinal resolution in spiral CT as affected by the spatially varying longitudinal aliasing. We propose the treatment of aliasing as a signal dependent, additive noise, and define a new image quality parameter, the contrast-to-aliased-noise ratio (CNaR), that relates to possible image degradation or loss of resolution caused by aliasing. We performed CT simulations and actual phantom scans using a resolution phantom consisting of sequences of spherical beads of different diameters, extending along the longitudinal axis. Our results show that the off-isocenter longitudinal resolution differs significantly from the longitudinal resolution at the isocenter and that the CNaR decreases with distance from the isocenter, and is a function of pitch and the helical interpolation algorithm used. The longitudinal resolution was observed to worsen with decreasing CNaR. We conclude that the longitudinal resolution in spiral CT is spatially varying, and can be characterized by the CNaR measured at the transaxial location of interest.
Authors:
S Y Yen; G D Rubin; S Napel
Related Documents :
15278717 - Alternative diagnoses to stone disease on unenhanced ct to investigate acute flank pain.
18430827 - Dual-energy ct iodine-subtraction virtual unenhanced technique to detect urinary stones...
21984417 - A method for the automatic quantification of the completeness of pulmonary fissures: ev...
2003437 - Imaging the gallbladder: a historical perspective.
22048317 - Diagnostic imaging of an achilles tendon rupture.
11960237 - Cranial imaging in child abuse.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't; Research Support, U.S. Gov't, P.H.S.    
Journal Detail:
Title:  Medical physics     Volume:  26     ISSN:  0094-2405     ISO Abbreviation:  Med Phys     Publication Date:  1999 Dec 
Date Detail:
Created Date:  2000-03-13     Completed Date:  2000-03-13     Revised Date:  2007-11-14    
Medline Journal Info:
Nlm Unique ID:  0425746     Medline TA:  Med Phys     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  2617-25     Citation Subset:  IM    
Affiliation:
Department of Electrical Engineering, Information Systems Laboratory, Stanford University, California 94305, USA. shin@s-word.stanford.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Computer Simulation
Models, Statistical
Phantoms, Imaging
Tomography, X-Ray Computed / methods*
Grant Support
ID/Acronym/Agency:
1R01-HL50305-01/HL/NHLBI NIH HHS

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Correction for geometric image distortion in the x-ray imaging chain: local technique versus global ...
Next Document:  Experimental verification of a CT-based Monte Carlo dose-calculation method in heterogeneous phantom...