A sensitive branched DNA HIV-1 signal amplification viral load assay with single day turnaround.
Jump to Full Text
MedLine Citation:
PMID:  22479381     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Branched DNA (bDNA) is a signal amplification technology used in clinical and research laboratories to quantitatively detect nucleic acids. An overnight incubation is a significant drawback of highly sensitive bDNA assays. The VERSANT® HIV-1 RNA 3.0 Assay (bDNA) ("Versant Assay") currently used in clinical laboratories was modified to allow shorter target incubation, enabling the viral load assay to be run in a single day. To dramatically reduce the target incubation from 16-18 h to 2.5 h, composition of only the "Lysis Diluent" solution was modified. Nucleic acid probes in the assay were unchanged. Performance of the modified assay (assay in development; not commercially available) was evaluated and compared to the Versant Assay. Dilution series replicates (>950 results) were used to demonstrate that analytical sensitivity, linearity, accuracy, and precision for the shorter modified assay are comparable to the Versant Assay. HIV RNA-positive clinical specimens (n = 135) showed no significant difference in quantification between the modified assay and the Versant Assay. Equivalent relative quantification of samples of eight genotypes was demonstrated for the two assays. Elevated levels of several potentially interfering endogenous substances had no effect on quantification or specificity of the modified assay. The modified assay with drastically improved turnaround time demonstrates the viability of signal-amplifying technology, such as bDNA, as an alternative to the PCR-based assays dominating viral load monitoring in clinical laboratories. Highly sensitive bDNA assays with a single day turnaround may be ideal for laboratories with especially stringent cost, contamination, or reliability requirements.
Authors:
Mark A Baumeister; Nan Zhang; Hilda Beas; Jesse R Brooks; Jesse A Canchola; Carlo Cosenza; Felix Kleshik; Vinod Rampersad; Johan Surtihadi; Thomas R Battersby
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2012-03-27
Journal Detail:
Title:  PloS one     Volume:  7     ISSN:  1932-6203     ISO Abbreviation:  PLoS ONE     Publication Date:  2012  
Date Detail:
Created Date:  2012-04-05     Completed Date:  2012-08-28     Revised Date:  2013-06-26    
Medline Journal Info:
Nlm Unique ID:  101285081     Medline TA:  PLoS One     Country:  United States    
Other Details:
Languages:  eng     Pagination:  e33295     Citation Subset:  IM    
Affiliation:
Siemens Healthcare Diagnostics Inc., Berkeley, California, United States of America.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Branched DNA Signal Amplification Assay / methods*
HIV Infections / diagnosis*,  virology*
HIV-1 / genetics*
Humans
Nucleic Acid Amplification Techniques / methods
Polymerase Chain Reaction
Reagent Kits, Diagnostic / standards
Reproducibility of Results
Sensitivity and Specificity
Time Factors
Viral Load
Chemical
Reg. No./Substance:
0/Reagent Kits, Diagnostic
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): PLoS One
Journal ID (iso-abbrev): PLoS ONE
Journal ID (publisher-id): plos
Journal ID (pmc): plosone
ISSN: 1932-6203
Publisher: Public Library of Science, San Francisco, USA
Article Information
Download PDF
Baumeister et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Received Day: 17 Month: 10 Year: 2011
Accepted Day: 11 Month: 2 Year: 2012
collection publication date: Year: 2012
Electronic publication date: Day: 27 Month: 3 Year: 2012
Volume: 7 Issue: 3
E-location ID: e33295
ID: 3314011
PubMed Id: 22479381
Publisher Id: PONE-D-11-20620
DOI: 10.1371/journal.pone.0033295

A Sensitive Branched DNA HIV-1 Signal Amplification Viral Load Assay with Single Day Turnaround Alternate Title:Branched DNA Assay with Single Day Turnaround
Mark A. Baumeister
Nan Zhang
Hilda Beas
Jesse R. Brooks
Jesse A. Canchola
Carlo Cosenza
Felix Kleshik
Vinod Rampersad
Johan Surtihadi
Thomas R. Battersbyaff1*
Richard C. Willsonedit1 Role: Editor
Siemens Healthcare Diagnostics Inc., Berkeley, California, United States of America
University of Houston, United States of America
Correspondence: * E-mail: thomas.battersby@siemens.com
Contributed by footnote: Conceived and designed the experiments: MAB JRB FK TRB. Performed the experiments: MAB HB JRB FK. Analyzed the data: NZ JAC CC JS. Contributed reagents/materials/analysis tools: VR. Wrote the paper: MAB TRB.

Introduction

Branched DNA (bDNA) technology, first developed over twenty years ago [1], is today widely used in clinical [2] and research [3] laboratories to quantitatively detect specific nucleic acid sequences. bDNA quantitative hybridization technology has a wide dynamic range and is sensitive enough for applications intended to reliably detect very few target molecules [4]. The technology has been described in detail elsewhere [4]. bDNA technology has some advantages over PCR technology, commonly used for applications requiring high sensitivity. Unlike PCR, where a region of the intended target is exponentially amplified in order to generate detectable signal, in bDNA assays only signal is amplified. bDNA assays are consequently not susceptible to contamination risks associated with PCR-based assays [5]. Minor sequence variation in probe-binding regions of target will not compromise assay performance, in contrast to PCR [6], leading to more robust quantitation across genotypes [7]. Furthermore, reproducibility, and therefore reliability, of bDNA assays is superior to PCR assays, especially at the low end of an assay's dynamic range [8].

However, a reduction in the processing time allowing the single day turnaround now standard in clinical PCR assays, would improve the productivity of laboratories utilizing bDNA. bDNA is a hybridization technology in which a series of probes is introduced to a sample well. Target is immobilized at the well surface and an extended branching structure with many signal-generating moieties is constructed to produce an amplified signal [4]. The primary component of the run time is an overnight (16–18 h) target incubation. Target incubation involves annealing of unpurified lysed target nucleic acids to a set of probes used in immobilizing the target and another set of probes to initiate construction of the extended signaling structure. We have modified the Versant HIV-1 RNA 3.0 Assay (bDNA) (“Versant Assay”) currently in use in clinical laboratories to allow the viral load assay to be run in a single day (assay in development; not commercially available). An evaluation of the modified assay consisting of analytical performance testing, a method comparison with clinical specimens, a genotype equivalence study, and testing of potentially interfering endogenous substances is presented.


Results

In an initial experiment we simply shortened the standard 16–18 h target incubation to a target period of 3 h, forcing the otherwise unmodified Versant Assay to provide results in little over a standard 8 h work shift. We examined analytical performance of the unmodified shortened assay with a target dilution series of non-infectious HIV-18E5/LAV virus (8E5) [9]. Performance suffered when comparing analogous parameters previously reported [2] for the Versant Assay. Assay sensitivity and precision were significantly worse with a simple truncation of the target incubation (Table S1).

Therefore, we focused on accelerating signal generation in the assay by speeding the diffusion of the ∼10 Kb target HIV RNA to the well surface and accelerating hybridization of the assay probes and target. Dilution series of target were analyzed to assess the impact of assay modification in these evaluations. Diffusion is most simply addressed by reducing the volume in the assay wells during target incubation. However, there are at least two potential problems with this approach. Surfaces of the wells have attached capture probe oligonucleotides. When the assay solution volume is reduced, some capacity for signal generation is inevitably lost as the solution level falls below some of the capture oligonucleotides. Additionally, sufficient solution must be present in the wells to ensure that the small volume lost to evaporation during assay manipulations does not significantly affect assay results.

Reducing the target incubation solution volume as low as 80 µL still left sufficient solution in the wells such that evaporation was not a problem under the assay conditions. Signal generation at 80 µL target incubation solution volume was much more rapid. Salt and surfactant concentrations in the solution at the lower volume were also adjusted to give rapid signal generation without proportionally increasing background signal in negative samples. Doubling these concentrations in the “Lysis Diluent” component gave optimal results for the lower volume target incubation (Table S2). Signal at 3 h with this modified, “low volume” condition was virtually identical to signal at 15 h with the standard Lysis Diluent volume across the quantitative range of the assay (Fig. 1).

Various polyanions were tested for ability to accelerate nucleic acid hybridization [10] in the assay (Fig. 2). Several polyanions that were examined significantly accelerated signal generation in the assay. Among the polyanionic species in our experiments, polyvinylsulfonic acid (PVSA) was consistently superior in rapidly boosting assay signal across the quantitative range of the assay. Optimization of PVSA concentration in the assay demonstrated that 0.3% (w/v) PVSA in the Lysis Diluent significantly increased signal generation, yet did not cause problems with precipitation upon storage (data not shown). Finally, the lower volume Lysis Diluent was prepared with PVSA (designated “M1 Lysis Diluent”) and used in assays with low volume target incubation, demonstrating that the two modifications together had some cumulative benefit in rapid signal generation (Fig. 3).

Preliminary experiments indicated that the levels of signal generated in the Versant Assay could be reached in as little as 2.5 h target incubation time simply by using M1 Lysis Diluent in the assay. Probe-containing components and associated volumes of addition were not changed. Assay evaluations with the M1 Lysis Diluent and assay software modified to accommodate 2.5 h target incubation (the “modified assay”) were performed, including analytical performance testing, a method comparison with clinical specimens, a genotype equivalence study, and testing of potentially interfering endogenous substances.

Using data gathered from 450 total individually collected HIV RNA-negative plasma specimens across eighteen runs with three kit/M1 Lysis Diluent lot combinations, a detection cutoff of 29 copies/mL was defined to deliver specificity with a lower level of the 95% confidence interval of at least 95% (Table 1, Negative cutoff determination). Specificity with the defined detection cutoff was verified as >95% by 144 additional individual HIV-negative plasma specimens across twelve more runs with the remaining three kit/M1 Lysis Diluent lot combinations (Table 1, Negative cutoff verification). The verified detection cutoff was then used in analyzing all other experiments.

Wells across all thirty runs containing dilution series replicates were used to determine analytical sensitivity, linearity, accuracy, and precision for the modified assay. The assay detection cutoff ascertained from the individual negative specimens was used in determining the assay limit of detection (LoD, 95% detection rate) as 70 copies/mL (95% CI from 64 to 76 copies/mL) by logistic curve fit. Linearity of the modified assay was evaluated by determining the absolute deviation of geometric mean quantitations from the predicted values on the linear regression line for dilution series levels above the LoD. These mean differences for all levels of the dilution series above the LoD were ≤0.07 log10 (Table 2, Log Difference). Accuracy of the modified assay was determined from the difference in measured and expected mean quantitations for dilution series levels. These differences were ≤0.06 log10 for all levels above the assay LoD with the modified assay (Table 2, Log Recovery). The range of total percent coefficients of variation (%CVs) corresponding to dilution series levels above the LoD in the modified assay were 13.8–31.4% for the modified assay (Table 2). Analytical sensitivity, linearity, accuracy, and precision determined with the modified assay are comparable to the analogous parameters previously reported [2] for the Versant Assay.

Quantitative equivalence of the modified assay and the Versant Assay was established with a set of common clinical specimens quantified in both assays (Fig. 4). Results quantifying below the LoD of the modified assay (6 results) were excluded from analysis. Average log difference between the Versant and modified assays was −0.025 (95% CI from −0.067 to 0.018), indicating no statistically significant difference in quantification between the two assays. Deming regression showed linearity for a plot of log quantifications for the Versant Assay against the log quantifications of the modified assay across the range examined (slope = 0.974 with 95% CI from 0.913 to 1.04). A Bland-Altman plot (Fig. 5) demonstrated no evidence of proportional bias across the linear range of the assay (regression slope = −0.026; 95% CI from −0.086 to 0.035).

Genotype equivalence was demonstrated with a high concentration series (∼50,000 copies/mL) and a low concentration series (∼500 copies/mL) of HIV subtype samples. The modified assay and the Versant Assay quantified each sample very similarly (Figs. S1 and S2).

Like the Versant Assay [2], none of the endogenous substances tested for potential interference in the modified assay had a significant effect on assay specificity with HIV-negative samples or quantification of target at approximately 1,000 copies/mL in Seracon II (Table S3).


Discussion

The modified bDNA assay demonstrated performance characteristics comparable to the more time-consuming Versant HIV-1 RNA 3.0 Assay (bDNA). The dramatically shortened target incubation was enabled by modification of only the Lysis Diluent used in target incubation. The assay retains advantages of bDNA technology, including superior reproducibility, reliability, and insusceptibility to contamination. Throughput with the modified assay is comparable to analogous PCR-based assays used in clinical laboratories. A single Versant 440 Molecular System can run two 96 well plates simultaneously, allowing the modified HIV-1 assay to routinely process 192 tests per day. bDNA assays with a single day turnaround may be ideal for laboratories with especially stringent cost, contamination, or reliability requirements.

Similar shortened target incubation can be implemented in bDNA assays targeting a variety of clinically relevant nucleic acid analytes. The HIV analyte targeted in the viral load assay presented here has a relatively stringent sensitivity requirement. The target incubation time of a bDNA assay can most likely be further shortened when targeting analytes lacking acute sensitivity requirements.


Materials and Methods
Assay Kits and Systems

The modified assay used Versant Assay kits, removing the original Lysis Diluent component and replacing it with M1 Lysis Diluent. All other Versant Assay reagents were used unchanged. Assays were performed on the VERSANT® 440 Molecular System with software modified to allow a target incubation period of 2.5 h. Some preliminary experimentation used in developing the M1 Lysis Diluent formulation was performed on the older System 340 bDNA Analyzer. Unmodified kit calibrators and controls from the Versant Assay were used in the modified assay to generate a standard curve and verify plate validity.

Reagent Modification

In M1 Lysis Diluent, salt and surfactant concentrations were doubled relative to the Lysis Diluent in the Versant Assay. Concentrations of the remaining components of the Lysis Diluent were identical in M1 Lysis Diluent. Additionally, polyvinylsulfonic acid was added to M1 Lysis Diluent (0.3% (w/v)). The target incubation solution volume in the modified assay was reduced to 80 µL from 140 µL per well in the Versant Assay, by adding 60 µL, instead of 120 µL, of “Lysis Working Reagent” made from M1 Lysis Diluent (Table S2).

Analytical Performance Evaluation

Analytical sensitivity, specificity, linearity, accuracy and precision were evaluated by testing three M1 Lysis Diluent lots made from diverse raw materials in all six permutations available with two assay kit lots. Specificity testing was performed on individual HIV-negative plasma specimens (ProMedDx). A dilution series of non-infectious HIV-18E5/LAV virus (8E5) [9] in Seracon II (SeraCare) was gravimetrically prepared. The 8E5 stock was previously value assigned from a reference transcript quantified by phosphate analysis [11]. The dilutions were then value assigned in the Versant Assay. This is identical to the dilution series preparation in a published evaluation of the Versant Assay [2], except 8E5 was substituted for β-propiolactone-treated virus. The dilution series was used for all analytical performance testing. Dilution series observations in analytical performance studies with Studentized residuals >4 for relative light unit signal were considered outliers and excluded from analysis (5 of 970 results).

Method Comparison

In the method comparison experiment, aliquots from 141 HIV-positive clinical specimens (ProMedDx, Bio Collections, or Northwest Biomedical) with viral loads distributed across the assay range were quantified in the Versant Assay and the modified assay. These specimens were collected with written informed consent of the participants and following protocols approved by New England IRB (ProMedDx), Independent Investigational Review Board, Inc. (BioCollections), or Beijing Youan Hospital IRB (Northwest Biomedical). The specimens were purchased delinked and deidentified from the participants by the commercial sources. An unweighted Deming regression was used in the method comparison analysis [12]. No outlier analysis was performed on the method comparison data.

Genotype Equivalence and Potentially Interfering Endogenous Substances

A panel of HIV RNA-positive samples composed of eight HIV-1 subtypes (SeraCare) was used to examine genotype quantification equivalence. Two levels of each panel member (∼50,000 and ∼500 copies/mL) were prepared by gravimetric dilution in Seracon II. Elevated levels of several endogenous substances, including hemoglobin (200 mg/dL), triglycerides (2000 mg/dL), albumin (6 g/dL), γ-globulins (6 g/dL), and bilirubin (20 mg/dL), were tested for potential interference in the modified assay. These levels are derived from recommended test levels in CLSI document EP7-A2 [13]. The endogenous substance samples were prepared in Seracon II, with 8E5 gravimetrically spiked to create HIV-positive samples.


Supporting Information Figure S1

Eight HIV subtypes at ∼50,000 copies/mL quantified by the modified assay and the VERSANT HIV-1 bDNA 3.0 Assay. Quantification by the VERSANT HIV-1bDNA 3.0 Assay (32 replicates of each subtype across 4 plates) is very similar to quantification by the modified assay (5 replicates of each subtype on a single plate).

(TIF)


Click here for additional data file (pone.0033295.s001.png)

Figure S2

Eight HIV subtypes at ∼500 copies/mL quantified by the modified assay and the VERSANT HIV-1bDNA 3.0 Assay. Quantification by the VERSANT HIV-1 bDNA 3.0 Assay (32 replicates of each subtype across 4 plates) is very similar to quantification by the modified assay (5 replicates of each subtype on a single plate).

(TIF)


Click here for additional data file (pone.0033295.s002.png)

Table S1

Analytical performance summary for the unmodified HIV-1 bDNA assay with truncated target incubation.

(DOC)


Click here for additional data file (pone.0033295.s003.doc)

Table S2

HIV Target Incubation: VERSANT versus modified assay volumes. In the Versant Assay, all reagents (except M1 Lysis Diluent) in Table S3 are supplied in separate vials. The components are combined following the instructions for use provided with the Versant Assay to make a “Lysis Working Reagent”, which is added (120 µL) to each sample well (total incubation volume 140 µL). In the modified assay, M1 Lysis Diluent (45 µL) is substituted for Lysis Diluent and the resulting Lysis Working Reagent is added (60 µL) to each sample well (total incubation volume 80 µL).

(DOC)


Click here for additional data file (pone.0033295.s004.doc)

Table S3

Endogenous substances: specificity with HIV-negative samples and quantification with HIV-positive samples in the modified assay.

(DOC)


Click here for additional data file (pone.0033295.s005.doc)


Notes

Competing Interests: The authors have read the journal's policy and have the following conflicts. The authors were all employed by the funder of the work, Siemens Healthcare Diagnostics, at the time this work was performed, as is implied by the institutional affiliations disclosed. A patent application [US provisional 61/138,601] has been filed for methods of speeding bDNA hybridization based on this work. As disclosed in the manuscript, there are no current products using the work in the manuscript, but the techniques in the manuscript are under development and a related product or products may appear in the future. This does not alter the authors' adherence to all the PLoS ONE policies on sharing data and materials.

Funding: Siemens Healthcare Diagnostics was the sole source of funding for this work. As the authors are all employees of Siemens Healthcare Diagnostics, and the authors performed all of the work, Siemens Healthcare Diagnostics was involved in the study design, data collection and analysis, decision to publish, and preparation of the manuscript.

Phanitha Koganti, Lhia Dolores, and the Siemens Clinical Laboratory ran some of the assays in the experiments presented. Yuvraj Sujan and John Piovesan assisted with software modification. Jason Veilleux, Alan Alpert, and Diane Powers provided the 8E5 dilution series used in evaluating analytical performance. Rebecca Reynolds and Jonathan Arnold provided helpful suggestions.


References
1. Horn T,Urdea M. Year: 1989Forks and combs and DNA: the synthesis of branched oligonucleotides.Nucleic Acids Res17695969672780317
2. Gleaves CA,Welle J,Campbell M,et al. Year: 2002Multicenter evaluation of the Bayer VERSANT HIV-1 RNA 3.0 assay: analytical and clinical performance.J Clin Virol2520521612367656
3. Knudsen BS,Allen AN,McLerran DF,Vessella RL,Karademos J,et al. Year: 2008Evaluation of the branched-chain DNA assay for measurement of RNA in formalin-fixed tissues.J MolDiagn10169176
4. Collins ML,Irvine B,Tyner D,Fine E,Zayati C,et al. Year: 1997A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml.Nucleic Acids Res25297929849224596
5. Aslanzadeh J. Year: 2004Preventing PCR amplification carryover contamination in a clinical laboratory.Ann Clin Lab Sci3438939615648778
6. Korn K,Weissbrich B,Henke-Gendo C,Heim A,Jauer CM,et al. Year: 2009Single-point mutations causing more than 100-fold underestimation of human immunodeficiency virus type 1 (HIV-1) load with the Cobas TaqMan HIV-1 real-time PCR assay.J Clin Microbiol471238124019193839
7. Murphy DG,Côté L,Fauvel M,René P,Vincelette J. Year: 2000Multicenter comparison of Roche COBAS AMPLICOR MONITOR Version 1.5, Organon Teknika NucliSens QT with Extractor, and Bayer Quantiplex Version 3.0 for quantification of human immunodeficiency virus type 1 RNA in plasma.J Clin Microbiol384034404111060065
8. Lubelchek RJ,Max B,Sandusky CJ,Hota B,Barker DE. Year: 2009Reliability at the lower limits of HIV-1 RNA quantification in clinical samples: a comparison of RT-PCR versus bDNA assays.PLoS ONE4e600819547711
9. Folks TM,Powell D,Lightfoote M,Koenig S,Fauci AS,et al. Year: 1986Biological and biochemical characterization of a cloned Leu-3- cell surviving infection with the acquired immune deficiency syndrome retrovirus.J Exp Med1642802903014036
10. Petrie CR III,Rutledge LM,Morgan JR. Year: 1998Chemical Promotion of Nucleic Acid Hybridization.Patent US5,853,986
11. Collins M,Zayati C,Detmer J,Daly B,Kolberg J,et al. Year: 1995Preparation and characterization of RNA standards for use in quantitative branched DNA hybridization assays.J Anal Biochem226120129
12. Cornbleet PJ,Gochman N. Year: 1979Incorrect least-squares regression coefficients in method-comparison analysis.Clin Chem2543238262186
13. Clinical and Laboratory Standards InstituteYear: 2005Interference Testing in Clinical Chemistry; Approved Guideline—Second Edition, CLSI document EP7-A2Wayne, PennsylvaniaClinical and Laboratory Standards Institute

Figures

[Figure ID: pone-0033295-g001]
doi: 10.1371/journal.pone.0033295.g001.
Figure 1  Signal generation in standard and low volume assays.

The standard assay was run with 3 and 15 h target incubation (140 µL) and the low volume assay was run with 3 h target incubation (80 µL). Identical lots of all components except Lysis Diluent were used on a single instrument for the three conditions. A dilution series of 8E5 in Seracon II at three concentrations (n = 10 for each condition) generated nearly identical average Relative Light Units (RLUs) for standard volume target incubation at 15 h and low volume target incubation volume at 3 h. Signal in negative samples remained unchanged (n = 6 for each condition).



[Figure ID: pone-0033295-g002]
doi: 10.1371/journal.pone.0033295.g002.
Figure 2  Polyanions in target incubation solution.

In this example, addition of 1% (w/V) of polyacrylic acid (PAA), polyantholesulfuric acid (PASA), or polyvinylsulfonic acid (PVSA) to the Lysis Diluent (LD, control) caused different enhancement of signal generation relative to the control condition. An 8E5 sample at 80,000 copies/mL (n = 5 for each condition) was analyzed here on a single plate with 3 h target incubation. Superior signal enhancement by PVSA relative to alternative polyanions was observed throughout our experiments. No polyanion caused discernable increase in signal with negative samples (n = 1 for each condition).



[Figure ID: pone-0033295-g003]
doi: 10.1371/journal.pone.0033295.g003.
Figure 3  Cumulative effect of lower volume and PVSA addition.

Three conditions were compared in this experiment: Versant HIV-1 RNA 3.0 Assay with standard volume target incubation (140 µL), low volume target incubation (80 µL), or low volume target incubation plus PVSA (80 µL, PVSA). Two concentrations of 8E5 (n = 4 for each level of each condition) were analyzed on a single plate with 2 h target incubation. Enhanced signal generation (RLU average) was observed for low volume target incubation, and further improvement was observed upon addition of PVSA to the low volume target incubation. Although signal increased slightly for negative samples (n = 6 for each condition) in the low volume conditions, signal for the low concentration positive samples in the low volume conditions increased more than enough to compensate.



[Figure ID: pone-0033295-g004]
doi: 10.1371/journal.pone.0033295.g004.
Figure 4  Method comparison of the modified bDNA assay and the VERSANT HIV-1 bDNA 3.0 Assay.

Log quantitations of 135 HIV-positive clinical specimens with the modified assay were measured and plotted. The Deming regression line (slope = 0.974 with 95% CI from 0.913 to 1.04) is indicated in the plot.



[Figure ID: pone-0033295-g005]
doi: 10.1371/journal.pone.0033295.g005.
Figure 5  Bland-Altman plot of method comparison data between the modified bDNA assay and the VERSANT HIV-1 bDNA 3.0 Assay.

No evidence of proportional bias across the linear range of the assay is apparent (regression slope = −0.026; 95% CI from −0.086 to 0.035).



Tables
[TableWrap ID: pone-0033295-t001] doi: 10.1371/journal.pone.0033295.t001.
Table 1  Specificity with individual HIV-negative clinical specimens: determination of 29 copies/mL as the negative cutoff and subsequent verification in the modified assay.
Negative cutoff determination Negative cutoff verification
N N below negative cutoff Specificity (95% Lower Confidence Limit) N N below negative cutoff Specificity (95% Lower Confidence Limit)
450 436 96.9% (95.2%) 144 140 97.2% (93.8%)

[TableWrap ID: pone-0033295-t002] doi: 10.1371/journal.pone.0033295.t002.
Table 2  Analytical performance summary for the modified HIV-1 bDNA assay.
Level (copies/mL) Total N Positive N Log Recovery Log Difference Total %CV
584926 197 197 0.06 0.06 16.6
58493 197 197 0.01 0.01 20.3
5849 199 199 0.06 0.07 13.8
585 196 196 0.01 0.01 31.3
117 347 345 −0.05 −0.04 30.4
88 347 341 −0.06 −0.05 31.4
58 348 313 −0.04 −0.03 29.7


Article Categories:
  • Research Article
Article Categories:
  • Biology
    • Molecular Cell Biology
      • Nucleic Acids
Article Categories:
  • Medicine
    • Diagnostic Medicine


Previous Document:  Genetic and non-genetic influences during pregnancy on infant global and site specific DNA methylati...
Next Document:  RUNX3 regulates intercellular adhesion molecule 3 (ICAM-3) expression during macrophage differentiat...