Abstract/OtherAbstract:
PURPOSE: To prospectively assess the efficacy of self-expanding nitinol aortic stents in the treatment of coarctation of the aorta (CoA). METHODS: Between July 2005 and July 2008, 21 patients (14 men; mean age 19.2+/-5.5 years, range 11-34) with CoA were treated with self-expanding Sinus-Aorta stents. The predilation balloon was selected to be <5 times the stenosis diameter. The stent diameter was selected to be 20% to 30% greater than the diameter of the undiseased aorta at the level of the diaphragm. RESULTS: All procedures were successfully performed without any major complications. Predilation (mean balloon diameter 12.3 mm) and postdilation (mean balloon diameter 15.4 mm) were performed in 12 and 14 procedures, respectively. The mean diameter of the stents was 21.6+/-2.3 mm (range 18-26). The mean peak transcoarctation pressure gradient decreased from 57.4+/-19.5 mmHg (range 30-100) before the procedure to 1.2+/-2.2 mmHg (range 0-7; p<0.001). Cephalad stent dislodgement with the first-generation device occurred in 3 of 12 patients; 2 were treated with a second stent overlapping the first, and the third received 3 overlapped stents after the second stent migrated distally. No stent dislodgement occurred in the subsequent 9 patients treated with longer second-generation stents with anti-jump markers. None of the patients had dissection, arterial rupture, or other complications. On follow-up, 1 (5%) patient had recoarctation and minor stent migration after 18 months; another stent was deployed successfully. No evidence of aneurysm formation was seen in 7 patients undergoing arch imaging. CONCLUSION: CoA can be successfully and safely managed with self-expanding nitinol aortic stents without aortic wall complications. Stent malpositioning can be overcome using oversized stents with anti-jump markers.
Authors:
Ali-Mohammad Haji-Zeinali; Payam Ghazi; Mohammad Alidoosti
Related Documents :
9602654 - Self expandable stents for relief of venous baffle obstruction after the mustard operat...
19660344 - Factors favoring stent-graft collapse after thoracic endovascular aortic repair.
8933984 - The v-stent: a novel technique for coronary bifurcation stenting.
15284464 - Sonographic follow-up after visceral artery stenting.
10892954 - Pneumoperitoneum secondary to endoscopic harvest of saphenous vein graft.
9157954 - In vivo transfer of lipoprotein(a) into human atherosclerotic carotid arterial intima.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists     Volume:  16     ISSN:  1526-6028     ISO Abbreviation:  J. Endovasc. Ther.     Publication Date:  2009 Apr 
Date Detail:
Created Date:  2009-05-21     Completed Date:  2009-06-19     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  100896915     Medline TA:  J Endovasc Ther     Country:  United States    
Other Details:
Languages:  eng     Pagination:  224-32     Citation Subset:  IM    
Affiliation:
Department of Interventional Cardiology, Tehran Heart Center, Tehran University of Medical Sciences, Tehran, Iran. ali_zeinali_cardio@yahoo.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Alloys*
Angioplasty, Balloon / adverse effects,  instrumentation*
Aortic Coarctation / radiography,  therapy*
Aortography
Child
Female
Humans
Male
Prospective Studies
Prosthesis Design
Stents*
Treatment Outcome
Young Adult
Chemical
Reg. No./Substance:
0/Alloys; 52013-44-2/nitinol

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Clinical significance and technical assessment of stent cell geometry in carotid artery stenting.
Next Document:  An unusual case of Dysphagia after endovascular exclusion of thoracoabdominal aortic aneurysm.