Safe scanning, but frequent artifacts mimicking bradycardia and tachycardia during magnetic resonance imaging (MRI) in patients with an implantable loop recorder (ILR).
MedLine Citation:
PMID:  16255749     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Patients with implantable devices are generally not permitted to undergo magnetic resonance imaging (MRI) because of potentially deleterious interactions. Little has been reported regarding the safety and effects of MRI scanning of patients with implantable loop recorders (ILRs). We evaluated the safety of scanning patients with ILRs and the output of the ILR after undergoing MRI. METHODS: Ten patients underwent 11 MRI scanning events. All patients had Reveal Plus (Medtronic, Minneapolis, MN) ILRs. Seven cranial, two lumbar-spine, one shoulder, and one knee MRI were performed. All of the MRIs were performed with the understanding that the patient had an ILR. In each patient, the ILR was cleared moments before the scan and the integrity of the signal and time date stamp were verified. The devices were reinterrogated immediately after MRI in 10 patients and two days post MR scanning in one patient. Each patient was questioned post MRI regarding any symptoms experienced during the scan. RESULTS: Both tachy and bradyarrhythmias appeared as artifacts as a result of ILR exposure to MRI. Post MRI, none of the ILRs showed diminished signal integrity, altered programmed parameters, diminished battery status, inability to communicate or be reprogrammed. No sensations of tugging or warmth at the implant site were noted. CONCLUSION: MRI was performed in ILR patients without harm to the patient or permanent damage to the ILR. MRI scanning of the Reveal appears safe. Artifact mimicking an arrhythmia was common, however, and must be excluded in any ILR patient undergoing MRI to avoid mistakenly attributing a syncopal episode, or palpitations to the artifacts produced from MRI exposure.
Authors:
J Rod Gimbel; Jamal Zarghami; Christian Machado; Bruce L Wilkoff
Related Documents :
12699799 - Perinatal mortality: clinical value of postmortem magnetic resonance imaging compared w...
10275609 - Swiss hospital institute's approach to the problems of magnetic resonance imaging.
18307629 - Feasibility of magnetic resonance imaging in patients with an implantable loop recorder.
11545109 - Magnetic resonance imaging in cardiomyopathies.
8533569 - Magnetic resonance imaging (mri) in endometrial carcinoma; preoperative estimation of d...
10300489 - Is it now time to acquire magnetic resonance imaging? issues and concerns.
24575889 - Hit the mark with diffusion-weighted imaging: metastases of rhabdomyosarcoma to the ext...
7220889 - Computed tomography in the evaluation of sacroiliac arthritis.
21493029 - Evaluation of mr diffusion-weighted imaging in differentiating endometriosis infiltrati...
Publication Detail:
Type:  Clinical Trial; Journal Article    
Journal Detail:
Title:  Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc     Volume:  10     ISSN:  1082-720X     ISO Abbreviation:  Ann Noninvasive Electrocardiol     Publication Date:  2005 Oct 
Date Detail:
Created Date:  2005-10-31     Completed Date:  2006-01-20     Revised Date:  2008-03-10    
Medline Journal Info:
Nlm Unique ID:  9607443     Medline TA:  Ann Noninvasive Electrocardiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  404-8     Citation Subset:  IM    
Affiliation:
Parkwest Hospital, 9330 Parkwest Boulevard, Ste. 202, Knoxville, TN 37923, USA. gimbeljr@ix.netcom.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Artifacts*
Bradycardia / diagnosis
Burns / epidemiology
Causality
Comorbidity
Diagnostic Errors / prevention & control,  statistics & numerical data*
Electrocardiography, Ambulatory / statistics & numerical data*
Equipment Failure / statistics & numerical data*
Humans
Magnetic Resonance Imaging / adverse effects*,  statistics & numerical data*
Prevalence
Reproducibility of Results
Sensitivity and Specificity
Tachycardia / diagnosis
United States / epidemiology

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Noninvasive risk stratification of subjects with a Brugada-type electrocardiogram and no history of ...
Next Document:  Effects of mobile telephones on the function of implantable cardioverter defibrillators.