Document Detail


Robot-assisted stenting of a high-grade anastomotic pulmonary artery stenosis following single lung transplantation.
MedLine Citation:
PMID:  20939718     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: To report robot-assisted stenting of a stenosis at the pulmonary artery anastomosis following lung transplantation, a rare complication that conveys poor prognosis even after surgical correction.
TECHNIQUE: The technique is illustrated in a 72-year-old man with end-stage lung disease who received a left single lung transplant. On postoperative day 54, he was evaluated for recurrent dyspnea on exertion that was due to a severe stenosis at the site of the pulmonary artery anastomosis. Balloon angioplasty was performed, and a 10-mm stent was deployed, with marked clinical improvement. Fourteen months later, he presented with recurrent symptoms due to in-stent restenosis. Multiple attempts at catheterization and balloon angioplasty of the stent failed. Due to the technical difficulty involved in maneuvering the balloon while maintaining stability, it was decided to repeat the angioplasty with the assistance of a Hansen Sensei remote robotic navigation system. The robotic arm markedly enhanced stability and facilitated successful navigation of the stented site. A 16-mm-diameter Wallstent was placed through the previously placed balloon-expandable stent and postdilated.
CONCLUSION: A remote robotic catheter navigation system was able to assist stenting of an anastomotic pulmonary artery stenosis following failure of conventional interventional techniques.
Authors:
Alan B Lumsden; Javier E Anaya-Ayala; Itamar Birnbaum; Mark G Davies; Jean Bismuth; Zulfiqar F Cheema; Hossam F El Sayed; Harish Seethamraju; Matthias Loebe; Miguel Valderrabano
Related Documents :
18388338 - Recanalization of an acute middle cerebral artery occlusion using a self-expanding, rec...
20667238 - Indications and problems of intracranial stenting. report of two cases.
10762768 - Experimental evaluation of a new single wire tantalum coil coronary stent (wiktor-iª).
20635738 - Endovascular stenting for native coarctation in older children and adolescents using ad...
8431038 - Damus-kaye-stansel with cavopulmonary connection for single ventricle and subaortic obs...
24952798 - Simple formulas to determine optimal subclavian central venous catheter tip placement i...
Publication Detail:
Type:  Case Reports; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists     Volume:  17     ISSN:  1545-1550     ISO Abbreviation:  J. Endovasc. Ther.     Publication Date:  2010 Oct 
Date Detail:
Created Date:  2010-10-13     Completed Date:  2011-02-09     Revised Date:  2011-03-02    
Medline Journal Info:
Nlm Unique ID:  100896915     Medline TA:  J Endovasc Ther     Country:  United States    
Other Details:
Languages:  eng     Pagination:  612-6     Citation Subset:  IM    
Affiliation:
Methodist DeBakey Heart & Vascular Center, Department of Cardiovascular Surgery, The Methodist Hospital, Houston, Texas 77030, USA. ABLumsden@tmhs.org
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Anastomosis, Surgical
Angioplasty, Balloon / instrumentation*
Arterial Occlusive Diseases / etiology,  radiography,  therapy*
Constriction, Pathologic
Humans
Lung Transplantation / adverse effects*
Male
Pulmonary Artery* / radiography
Recurrence
Robotics*
Stents*
Therapy, Computer-Assisted*
Treatment Outcome
Comments/Corrections
Erratum In:
J Endovasc Ther. 2010 Dec;17(6):A6

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Current status of hybrid procedures for thoracoabdominal and pararenal aortic aneurysm repair: techn...
Next Document:  Acute type B aortic intramural hematoma: treatment strategy and the role of endovascular repair.