Document Detail


Risks in using siphon-reducing devices in adult patients with normal-pressure hydrocephalus: bench test investigations with Delta valves.
MedLine Citation:
PMID:  8613856     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
To elucidate the reason for malfunction of Delta valves in patients with normal-pressure hydrocephalus, the authors applied a new concept of the bench test and simulated intracranial pressure (ICP) to measure the resultant flow with a complete shunt system. Subcutaneous pressure on the valve, valve implantation site, and postural hydrostatic differential-pressure changes were simulated in this bench test designed for adult patients with normal-pressure hydrocephalus. Subcutaneous pressure ranged within 6.4 +/- 0.09 cm H2O (mean +/- standard deviation) on the 7th day after the implantation of the valve in rats. A linear correlation between valve closing pressures and the external pressure was observed: gradual increase of the latter resulted in an approximately equal rise in the closing pressure. Closing pressure ranged within the physiological variation of ICP (10 +/- 5 cm H2O) in the supine position. In the erect position it was higher than the physiological variation (+5 to -5 cm H2O) when the valve was placed at the level of the foramen of Monro and lower when placed at the level of the clavicle. This observation indicates that the subcutaneous pressure around the Delta valve significantly affects its closing pressure. An excessive reduction of flow or a functional obstruction occurs when the patient with a Delta valve implanted at the foramen of Monro assumes an erect posture. The authors conclude that Delta valves are not recommended for implantation at the level of the foramen of Monro in patients with normal-pressure hydrocephalus because of the risk of underdrainage. The mastoid process or clavicle may be the alternative levels for its implantation to prevent underdrainage; however, the possibility of shunt malfunction should be kept in mind because subcutaneous pressure is variable among patients and that might affect the performance characteristics of these valves.
Authors:
M Hassan; S Higashi; J Yamashita
Related Documents :
23732146 - Efficacy of baroreflex activation therapy for the treatment of resistant hypertension.
2339826 - The vascular resistance of arterial stenoses in series.
14514546 - Doppler echocardiographic evaluation of prosthetic valves in tricuspid position.
10392496 - Adjustable antisiphon shunt.
22379186 - Blood pressure target attainment in the background of guidelines: the very elderly in s...
23528996 - High pressure extraction of antioxidants from solanum stenotomun peel.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of neurosurgery     Volume:  84     ISSN:  0022-3085     ISO Abbreviation:  J. Neurosurg.     Publication Date:  1996 Apr 
Date Detail:
Created Date:  1996-06-04     Completed Date:  1996-06-04     Revised Date:  2003-11-14    
Medline Journal Info:
Nlm Unique ID:  0253357     Medline TA:  J Neurosurg     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  634-41     Citation Subset:  AIM; IM    
Affiliation:
Department of Neurosurgery, Kanazawa University School of Medicine, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Hydrocephalus, Normal Pressure / physiopathology,  surgery*
Intracranial Pressure
Rats
Rats, Inbred F344
Risk Factors
Time Factors
Ventriculoperitoneal Shunt*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effects of prednisone on ventriculoperitoneal shunt function in hydrocephalus secondary to cysticerc...
Next Document:  Brain tissue pressure gradients created by expanding frontal epidural mass lesion.