Retrospective analysis of 88,429 serum and urine glucose EQA results obtained from professional laboratories and medical offices participating in surveys organized by three European EQA centers between 1996 and 2007.
MedLine Citation:
PMID:  20678030     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The aim of this study was to provide inter-laboratory imprecision comparisons of different groups of diagnostic systems as well as a comparison of professional laboratories with medical offices performance on the basis of 88,429 glucose results obtained in external quality assessment (EQA) schemes organized by three European EQA centers between 1996 and 2007.
METHODS: A simple, non-parametrical statistical model suited to all EQA results, including outliers, was used to calculate yearly and global performance.
RESULTS: The best performance was obtained from professional laboratories with a group of three diagnostic systems--Hitachi, Integra, and Vitros, followed by Cobas Mira, and finally by Reflotron. For medical offices, the best performance was achieved with the Cobas Mira diagnostic systems, followed by the Reflotron, SpotChem, and Vitros DT60 diagnostic systems. A slight but overall improvement in performance over time was observed for most diagnostic devices.
CONCLUSIONS: The analysis of glucose EQA results collected over a 12-year period showed that professional laboratories obtained better performances than medical offices, and that a general improvement in yearly performance was observed for both types of laboratories.
Authors:
Pierre-Alain Morandi; André Deom; Dagmar Kesseler; Richard Cohen
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Clinical chemistry and laboratory medicine : CCLM / FESCC     Volume:  48     ISSN:  1434-6621     ISO Abbreviation:  Clin. Chem. Lab. Med.     Publication Date:  2010 Sep 
Date Detail:
Created Date:  2010-09-22     Completed Date:  2011-01-05     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9806306     Medline TA:  Clin Chem Lab Med     Country:  Germany    
Other Details:
Languages:  eng     Pagination:  1255-62     Citation Subset:  IM    
Affiliation:
CSCQ, Chêne-Bourg, Switzerland.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Chemical Analysis / standards*
Blood Glucose / analysis*
Data Collection*
Europe
Glycosuria / diagnosis*
Humans
Laboratories / standards*
Medical Office Buildings / standards*
Quality Control
Retrospective Studies
Urinalysis / standards*
Chemical
Reg. No./Substance:
0/Blood Glucose

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Transfer of myxovirus-protein-A mRNA assay for interferon-? bioactivity measurement in multiple scle...
Next Document:  Effect of storage temperature or collection tube type on blood methylmalonic acid concentrations mea...