Document Detail

Retrieval of a partially degloved stent strut during percutaneous coronary intervention.
Jump to Full Text
MedLine Citation:
PMID:  22977756     Owner:  NLM     Status:  PubMed-not-MEDLINE    
Abstract/OtherAbstract:
Although stent entrapment is a rare event during percutaneous coronary intervention, stent entrapment can cause stent breakage or loss, which results in fatal complications such as stent embolism or acute myocardial infarction. We report one case of stent entrapment that was successfully treated by a snare via a contralateral transfemoral approach.
Authors:
Woohyeun Kim; Yong Hyun Kim; Sang Yup Lim; Seong Hwan Kim; Jeong Chun Ahn; Woo Hyuk Song
Related Documents :
8605616 - Correlates and in-hospital outcome of painless presentation of acute myocardial infarct...
15194026 - Prognostic significance of serum creatinine concentration for in-hospital mortality in ...
19032996 - Trends in presenting characteristics and hospital mortality among patients with st elev...
19752326 - Perioperative risk predictors of cardiac outcomes in patients undergoing liver transpla...
23212806 - Urinary albumin concentration and long-term cardiovascular risk in acute coronary syndr...
2857086 - Somatostatin in the human heart and comparison with guinea pig and rat heart.
Publication Detail:
Type:  Journal Article     Date:  2012-08-24
Journal Detail:
Title:  Chonnam medical journal     Volume:  48     ISSN:  2233-7393     ISO Abbreviation:  Chonnam Med J     Publication Date:  2012 Aug 
Date Detail:
Created Date:  2012-09-14     Completed Date:  2012-10-02     Revised Date:  2013-05-30    
Medline Journal Info:
Nlm Unique ID:  101564659     Medline TA:  Chonnam Med J     Country:  Korea (South)    
Other Details:
Languages:  eng     Pagination:  130-2     Citation Subset:  -    
Affiliation:
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Chonnam Med J
Journal ID (iso-abbrev): Chonnam Med J
Journal ID (publisher-id): CMJ
ISSN: 2233-7385
ISSN: 2233-7393
Publisher: Chonnam National University Medical School
Article Information
Download PDF
© Chonnam Medical Journal, 2012
open-access:
Received Day: 22 Month: 5 Year: 2012
Revision Received Day: 18 Month: 7 Year: 2012
Accepted Day: 23 Month: 7 Year: 2012
Print publication date: Month: 8 Year: 2012
Electronic publication date: Day: 24 Month: 8 Year: 2012
Volume: 48 Issue: 2
First Page: 130 Last Page: 132
ID: 3434794
PubMed Id: 22977756
DOI: 10.4068/cmj.2012.48.2.130

Retrieval of a Partially Degloved Stent Strut During Percutaneous Coronary Intervention
Woohyeun KimA1
Yong Hyun KimA1
Sang Yup LimA1
Seong Hwan KimA1
Jeong Chun AhnA1
Woo Hyuk SongA1
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Correspondence: Corresponding Author: Sang Yup Lim. Department of Internal Medicine, Korea University Ansan Hospital, 516 Gojan-dong, Danwon-gu, Ansan 425-707, Korea. TEL: +82-31-412-6755, FAX: +82-31-412-5594, vnlover@hanmail.net

INTRODUCTION

Complications during percutaneous coronary intervention (PCI) include coronary artery complications (perforation, distal embolization, side branch occlusion, and stent thrombosis) and vascular complications (puncture site hematoma, retroperitoneal bleeding, and atheroembolism). Among them, stent entrapment with resulting stent deployment failure is rare but can cause fatal conditions such as stent thrombosis or myocardial infarction.

Although many methods have been used for stent retrieval, there are few reports of the use of a snare via a contralateral femoral approach. Here we report one case of half-degloved stent entrapment that occurred during PCI and that was successfully treated by a snare via a contralateral approach.


CASE REPORT

An 80-year-old male was admitted owing to recurrent chest pain for elective PCI of a chronic total occlusion (CTO) in a left circumflex (LCX) lesion. Six months previously, the patient had undergone emergent PCI in the left anterior descending artery (LAD) owing to unstable angina. Follow-up coronary angiography revealed no significant in-stent restenosis in the LAD and CTO in the LCX (Fig. 1). We decided to revascularize the LCX lesion. A 7-Fr JL4 guiding catheter was engaged and predilation with a 2.0×20-mm balloon was done. After ballooning, we attempted to insert the 2.5×30-mm Resolute-Integrity (Medtronic, Minneapolis, MN, USA) stent into the LCX, but the stent failed to cross the CTO lesion in the LCX. We used the deep intubation technique, but the stent could not cross the lesion entirely because the lesion was severely calcified and angulated. While attempting to push the stent, we noticed that the stent was stuck within the lesion. After a struggle, the guiding catheter was taken out but we observed that the stent was half-degloved off the balloon (Fig. 2). Because of the risk of stent loss and embolization, we decided to pull out the catheter and stent strut by use of an Amplatz GooseNeck snare (EV3, Plymouth, MN, USA). We pushed down the guiding catheter and stent strut together to the iliac bifurcation level. The left femoral artery was punctured and we used a 6Fr Balkin sheath to grab the stent with the snare and both the stent strut and the balloon catheter were removed successfully by the snare (Fig. 3). We completed the procedure and decided on medical therapy instead of PCI for the LCX lesion. Two days later, the patient was discharged and there have been no adverse events during 3 months of clinical follow-up.


DISCUSSION

Entrapment of coronary angioplasty devices, especially the stent strut, is rare but is a serious complication of PCI for which cardiac surgery is sometimes required. Calcified, long lesions and angulated lesions may predispose to entrapment of the stent strut. The incidence of stent dislodgement or migration ranges from 1.4% to 8%.1-4 To prevent stent dislodgement or migration, predilatation with a balloon catheter or use of stents with a smaller profile and good tractability are recommended in case of severe angulated or calcified lesions, as in this patient.5

A variety of stent retrieval methods have been described in the literature, such as using a snare device, a multipurpose basket, a variety of forceps, an Angioguard (a distal protection device), a simple balloon and just crushing of the stent into the vessel wall by a balloon.6-10

However, using a balloon or a snare to rescue an entrapped stent in a half-degloved state may contribute to stent loss; to prevent stent loss, it would be safer to approach both sides by inserting the snare via another vascular route.

In this case, we pushed down the stent strut and guiding catheter together to the iliac bifurcation level. After that, the contralateral femoral artery was punctured and both stent strut and balloon catheter were removed successfully by a snare with the aid of a 6Fr Balkin sheath. We think that this technique is safer than other methods in this situation to prevent stent loss and embolization.

However, to avoid serious complications, interventional cardiologists should keep in mind the various complications during PCI and should understand the lesion anatomy and characteristics in their daily practice.


References
1. Elsner M,Peifer A,Kasper W. Intracoronary loss of balloon-mounted stents: successful retrieval with a 2 mm-"Microsnare"-deviceCathet Cardiovasc DiagnYear: 1996392712768933971
2. Carrozza JP Jr,Kuntz RE,Levine MJ,Pomerantz RM,Fishman RF,Mansour M,et al. Angiographic and clinical outcome of intracoronary stenting: immediate and long-term results from a large single-center experienceJ Am Coll CardiolYear: 1992203283371634668
3. Schatz RA,Baim DS,Leon M,Ellis SG,Goldberg S,Hirshfeld JW,et al. Clinical experience with the Palmaz-Schatz coronary stent. Initial results of a multicenter studyCirculationYear: 1991831481611984878
4. Schömig A,Kastrati A,Mudra H,Blasini R,Schühlen H,Klauss V,et al. Four-year experience with Palmaz-Schatz stenting in coronary angioplasty complicated by dissection with threatened or present vessel closureCirculationYear: 199490271627247994813
5. Feldman T. Tricks for overcoming difficult stent deliveryCatheter Cardiovasc IntervYear: 19994828528610525229
6. Eggebrecht H,Haude M,von Birgelen C,Oldenburg O,Baumgart D,Herrmann J,et al. Nonsurgical retrieval of embolized coronary stentsCatheter Cardiovasc IntervYear: 20005143244011108675
7. Foster-Smith KW,Garratt KN,Higano ST,Holmes DR Jr. Retrieval techniques for managing flexible intracoronary stent misplacementCathet Cardiovasc DiagnYear: 19933063688402869
8. Eeckhout E,Stauffer JC,Goy JJ. Retrieval of a migrated coronary stent by means of an alligator forceps catheterCathet Cardiovasc DiagnYear: 1993301661688221873
9. Kim MH,Cha KS,Kim JS. Retrieval of dislodged and disfigured transradially delivered coronary stent: report on a case using forcep and antegrade brachial sheath insertionCatheter Cardiovasc IntervYear: 20015248949111285606
10. Berder V,Bedossa M,Gras D,Paillard F,Le Breton H,Pony JC. Retrieval of a lost coronary stent from the descending aorta using a PTCA balloon and biopsy forcepsCathet Cardiovasc DiagnYear: 1993283513538462088

Article Categories:
  • Case Report

Keywords: Stents, Angioplasty, Complications.