BACKGROUND: In the last fifty years since plate and screw osteosynthesis has been implemented in fracture treatment, osteosporotic bone fractures were observed as a special problem. Due to special histologic, anatomic, physical and biomehanic properties of osteoporotic changed bone the laws of biomechanics suggest that stable osteosynthesis for osteoporotic bone is necessary to increase the contact surface of metallic implants and bone and the stability of the screw-plate-bone compound. There are numerous surgical techniques and methods for treatment of osteoporotic proximal humeral fractures. Every surgical procedure has to establish anatomical reduction and stable fixation that will enable early mobilisation.
SUBJECTS AND METHODS: The aim of this study was to present results of internal fixation of proximal humeral osteoporotic fractures with PHILOS locking plate. Between 2007 and 2012, a total of 67 patients older than 65 years with closed proximal humerus fractures underwent surgical treatment with PHILOS plate system (Synthes, Switzerland). 42 patients were operated with deltopectoral approach and 25 with deltoid split approach. After a mean follow up period of 14.68 (6-28) months functional and radiologic results were assessed.
RESULTS: We noted 9 postoperative complications related to surgical technique (1 intraarticular screw placement, 1 displacement in major tuberculum fragment, 1 displacement in major tuberculum fragment along with oblique placement of the plate, 2 cases of inadequate reduction, 1 case of humeral head avascular necrosis, varus humeral head fixation in 3 cases). None of the patients developed superficial or deep surgical infection. There was no nonunions. In the final evaluation, the Constant shoulder score was 91.75 (72-100).
CONCLUSIONS: In this study PHILOS locking plate showed good applicability, respecting bone biologic properties because of negligible interference with blood supply of the humeral head. There was no requirement to shape the plate enabling stabilization at constant angles as clear benefit of this plate. All that enables early mobilisation, and no implant insufficiency resulting in satisfactory treatment results and high Constant shoulder scores.
Authors:
Ante Bandalović; Fabijan Cukelj; Josip Knežević; Marko Ostojić; Arsen Pavić; Zlatko Parać; Matko Rošin
Related Documents :
19576398 - The history of the anterior approach to the hip.
25284548 - Endodontic management of a young patient with avulsion and root fracture: a case report...
21644408 - Delayed posttraumatic unilateral occipital epidural hygroma in early childhood.
20658098 - Operative fixation of a subtrochanteric fracture in a patient with previous spontaneous...
20595128 - Bilateral insufficiency fracture of the femoral head and neck in a case of oncogenic os...
8486298 - Enhancing independence in the older hip fracture patient.
Publication Detail:
Type:  JOURNAL ARTICLE    
Journal Detail:
Title:  Psychiatria Danubina     Volume:  26     ISSN:  0353-5053     ISO Abbreviation:  Psychiatr Danub     Publication Date:  2014 Dec 
Date Detail:
Created Date:  2014-11-29     Completed Date:  -     Revised Date:  2014-11-30    
Medline Journal Info:
Nlm Unique ID:  9424753     Medline TA:  Psychiatr Danub     Country:  -    
Other Details:
Languages:  ENG     Pagination:  376-381     Citation Subset:  -    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Computer Representation of Osteosynthesis Stability in Locking Plates Used for the Treatment of Oste...
Next Document:  Postoperative infections after posterior spondylodesis of thoracic and lumbal spine. Surgical spine ...