Reproducibility of ultrasound scan in the assessment of volume flow in the veins of the lower extremities.
MedLine Citation:
PMID:  11877703     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: The purpose of this study was the identification of the optimal settings of ultrasound scan flow measurement in the veins and the determination of whether the standardization of these settings can provide acceptable reproducibility of the venous flow measurements in individual segments of the lower extremity veins.
METHODS: The venous cross-sectional area, the time average mean velocity, and the venous volume flow of 25 healthy volunteers were examined with duplex ultrasound scanning. Reproducibility was examined with different measurement settings. Doppler scan sample volume size, ultrasound scan beam incident angle, and time interval of measurement were varied across a spectrum for arrival at the setting for highest reproducibility of the flow volume measurements. Test-retest reproducibility of venous flow volume measurements then was investigated with optimized settings.
RESULTS: The highest repeatability of volume flow measurements was achieved when the full lumen of the vein was insonated (coefficient of repeatability [CR] = 1.88 cm/s), the ultrasound scan beam incident angle was equal to 60 degrees (CR = 1.56 cm/s), and the measurement time was more than 40 seconds (CR = 1.64 cm/s). The mean values of volume flow were 360 mL/min in the common femoral vein, 147 mL/min in the superficial femoral vein, 86 mL/min in the profunda femoral vein, and 38 mL/min in the greater saphenous vein. Test-retest repeatability coefficients were 96.9 mL/min for the common femoral vein, 70.2 mL/min for the superficial femoral vein, 40.8 mL/min for the profunda femoral vein, and 16.8 mL/min for the greater saphenous vein.
CONCLUSION: The reproducibility of ultrasound scan measurements of volume flow in veins is optimized with the use of sampling volumes that cover the entire venous lumen, with an incident angle of 60 degrees and measuring for 40-second intervals or longer. With these defined variables, volumetric measurements are sufficiently repeatable. the values of flow volume measured with duplex ultrasound scanning were comparable to those with thermodilution techniques that were reported previously.
Authors:
Tomohiro Ogawa; Fedor Lurie; Robert L Kistner; Bo Eklof; Frank L Tabrah
Related Documents :
2260663 - Regulation of canine jejunal transit.
8190523 - Cerebral responses to single and multiple cocaine injections in newborn sheep.
6818113 - The potentiating effect of cortisone (prednisolone) on the fibrinolytic and clotting fa...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of vascular surgery     Volume:  35     ISSN:  0741-5214     ISO Abbreviation:  J. Vasc. Surg.     Publication Date:  2002 Mar 
Date Detail:
Created Date:  2002-03-05     Completed Date:  2002-04-24     Revised Date:  2012-10-03    
Medline Journal Info:
Nlm Unique ID:  8407742     Medline TA:  J Vasc Surg     Country:  United States    
Other Details:
Languages:  eng     Pagination:  527-31     Citation Subset:  IM    
Affiliation:
Department of Surgery, University of Hawaii, John A. Burns School of Medicine, and the Straub Foundation and Straub Clinic and Hospital, Honolulu, HI 96814, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Flow Velocity / physiology*
Humans
Leg / blood supply*,  ultrasonography*
Models, Theoretical
Reference Values
Reproducibility of Results
Time Factors
Ultrasonography, Doppler, Duplex*
Veins / physiology*,  ultrasonography

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Changes in venous lumen size and shape do not affect the accuracy of volume flow measurements in hea...
Next Document:  Differential effects of vascular growth factors on arterial and venous angiogenesis.