Document Detail


Remote magnetic control of a wireless capsule endoscope in the esophagus is safe and feasible: results of a randomized, clinical trial in healthy volunteers.
MedLine Citation:
PMID:  20855064     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
BACKGROUND: Remote control of esophageal capsule endoscopes could enhance diagnostic accuracy.
OBJECTIVE: To assess the safety and efficacy of remote magnetic manipulation of a modified capsule endoscope (magnetic maneuverable capsule [MMC]; Given Imaging Ltd, Yoqneam, Israel) in the esophagus of healthy humans.
DESIGN: Randomized, controlled trial.
SETTING: Academic hospital.
PATIENTS: This study involved 10 healthy volunteers.
INTERVENTION: All participants swallowed a conventional capsule (ESO2; Given Imaging) and a capsule endoscope with magnetic material, the MMC, which is activated by a thermal switch, in random order (1 week apart). An external magnetic paddle (EMP; Given Imaging) was used to manipulate the MMC within the esophageal lumen. MMC responsiveness was evaluated on a screen showing the MMC film in real time.
MAIN OUTCOME MEASUREMENTS: Safety and tolerability of the procedure (questionnaire), responsiveness of the MMC to the EMP, esophageal transit time, and visualization of the Z-line.
RESULTS: No adverse events occurred apart from mild retrosternal pressure (n = 5). The ability to rotate the MMC around its longitudinal axis and to tilt it by defined movements of the EMP was clearly demonstrated in 9 volunteers. Esophageal transit time was highly variable for both capsules (MMC, 111-1514 seconds; ESO2, 47-1474 seconds), but the MMC stayed longer in the esophagus in 8 participants (P < .01). Visualization of the Z-line was more efficient with the ESO2 (inspection of 73% ± 18% of the circumference vs 33% ± 27%, P = .01).
LIMITATIONS: Magnetic forces were not strong enough to hold the MMC against peristalsis when the capsule approached the gastroesophageal junction.
CONCLUSION: Remote control of the MMC in the esophagus of healthy volunteers is safe and feasible, but higher magnetic forces may be needed.
Authors:
Jutta Keller; Christiane Fibbe; Frank Volke; Jeremy Gerber; Alexander C Mosse; Meike Reimann-Zawadzki; Elisha Rabinovitz; Peter Layer; Paul Swain
Related Documents :
16628904 - Monitoring presence and advance of the alien invasive western corn rootworm beetle in e...
20578014 - Vital signs and cognitive function are not affected by 23-sodium and 17-oxygen magnetic...
17148824 - Rapid calculations of susceptibility-induced magnetostatic field perturbations for in v...
20373424 - Induced magnetic force in human heads exposed to 4 t mri.
12876044 - Mr imaging of bone marrow edema and joint effusion in patients with osteonecrosis of th...
6311484 - The application of nuclear magnetic resonance data to radiation therapy.
Publication Detail:
Type:  Journal Article     Date:  2010-09-19
Journal Detail:
Title:  Gastrointestinal endoscopy     Volume:  72     ISSN:  1097-6779     ISO Abbreviation:  Gastrointest. Endosc.     Publication Date:  2010 Nov 
Date Detail:
Created Date:  2010-11-01     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0010505     Medline TA:  Gastrointest Endosc     Country:  United States    
Other Details:
Languages:  eng     Pagination:  941-6     Citation Subset:  IM    
Copyright Information:
Copyright © 2010 American Society for Gastrointestinal Endoscopy. Published by Mosby, Inc. All rights reserved.
Affiliation:
Department of Internal Medicine, Israelitic Hospital, Academic Hospital, University of Hamburg, Germany. keller@ik-h.de
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Esophageal squamous cell carcinoma with intramural metastasis presenting as a pendiculated polyp.
Next Document:  Caught on camera: an unusual type of bug in the gut.