Rapid assessment of cardiac anatomy and function with a new hand-carried ultrasound device (OptiGo): a comparison with standard echocardiography.
MedLine Citation:
PMID:  11888820     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
AIMS: The aim of this study was to evaluate image quality and accuracy of a new hand-carried ultrasound device, OptiGo (Agilent Technologies) when compared to standard echocardiography in the setting of a focused examination in the assessment of cardiac anatomy and function. METHODS AND RESULTS:One-hundred and twenty-one patients were prospectively enrolled. Image quality and accuracy in assessment of chamber sizes, left ventricular (LV) wall thickness and contractility, right ventricular (RV) function, mitral and aortic leaflet thickening, mitral annular calcification, pericardial effusion and valvular regurgitation were assessed. Two-dimensional (2D) findings were graded on a four-point scale, except for LV function (six-point) and valvular leaflet opening (two-point). Colour Doppler assessment of valvular regurgitation was graded on a seven-point scale. A one-point difference was considered minor; a two or more point difference was considered major. There was no statistically significant difference in image quality between the two devices. For 2D data, the number of total (minor and major) differences between the hand-carried and standard echocardiograph examinations was significantly greater than the inter-observer variability (14.3% vs 10.7%, P< 0.05), however, major differences alone were not statistically different. For the colour Doppler assessment of regurgitation there was a significant difference between the devices for total (minor and major) differences, (40.0% vs 31.8%,P < 0.007) however, the number of major differences is explained by inter-observer variability. CONCLUSIONS: Image quality and diagnostic accuracy of the hand-carried device, OptiGo, was adequate for the purpose of performing a focused assessment of a limited number of 2D and Doppler parameters for the evaluation of cardiac anatomy and function.
Authors:
M Rugolotto; B S Hu; D H Liang; I Schnittger
Publication Detail:
Type:  Comparative Study; Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  European journal of echocardiography : the journal of the Working Group on Echocardiography of the European Society of Cardiology     Volume:  2     ISSN:  1525-2167     ISO Abbreviation:  Eur J Echocardiogr     Publication Date:  2001 Dec 
Date Detail:
Created Date:  2002-03-12     Completed Date:  2002-04-29     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  100890618     Medline TA:  Eur J Echocardiogr     Country:  England    
Other Details:
Languages:  eng     Pagination:  262-9     Citation Subset:  IM    
Affiliation:
Division of Cardiovascular Medicine, Stanford University Medical Center, Stanford, California 94305-5233, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Aged, 80 and over
Echocardiography, Doppler / instrumentation*,  standards
Female
Humans
Male
Middle Aged
Point-of-Care Systems / standards*
Prospective Studies
Comments/Corrections
Comment In:
Eur J Echocardiogr. 2001 Dec;2(4):216-8   [PMID:  11888815 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Quantification of left ventricular function with contrast-enhanced harmonic colour Doppler and a sem...