Quantitation of the human basal ganglia with positron emission tomography: a phantom study of the effect of contrast and axial positioning.
MedLine Citation:
PMID:  18222819     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
The accurate measurement of the concentration of a radioisotope in small structures with positron emission tomography (PET) requires a correction for quantitation loss due to the partial volume effect and the effect of scattered radiation. To evaluate errors associated with measures in the human basal ganglia (BG), a unilateral model of the BG has been inserted in a 20-cm cylinder. The recovery coefficient (RC=measured activity/true activity) for the BG phantom has been measured on a CTI tomograph with different background concentrations (contrast) and at different axial locations in the gantry. The BG was visualized on four or five slices, depending on its position in the gantry and on the contrast used. The RC was 0.75 with no background (contrast equal to 1.0). Increasing the relative radioactivity concentration in the background increased the RC from 0.75 to 2.00 when the contrast was -0.7 (BG<background). The RC was also affected by the size and the shape of the region of interest (ROI) used (RC from 0.75 to 0.67 with ROI size from 0.12 to 1.41 cm(2)). These results show that accurate RC correction depends not only on the volume of the structure but also on its contrast with its surroundings, as well as on the selection of the ROI.
Authors:
B Bendriem; S L Dewey; D J Schlyer; A P Wolf; N D Volkow
Related Documents :
21716829 - Magnetic resonance "flip-flop" in idiopathic intracranial hypertension.
17025269 - Micropet investigation of chronic long-term neurotoxicity from heavy ion irradiation.
16684819 - The diagnosis of ectopic focal hyperinsulinism of infancy with [18f]-dopa positron emis...
15770969 - An intra-operative positron probe with background rejection capability for fdg-guided s...
1966959 - Potential of labeling monoclonal antibodies for positron emission tomography and magnet...
18261529 - Non-oncologic imaging of the larynx.
8357619 - Imaging the sphenoid bone and basiocciput: pathological considerations.
11545119 - An aortocaval fistula diagnosed with 1.5-t magnetic resonance angiography.
21701109 - Hydrocephalus due to diffuse villous hyperplasia of the choroid plexus.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  IEEE transactions on medical imaging     Volume:  10     ISSN:  0278-0062     ISO Abbreviation:  IEEE Trans Med Imaging     Publication Date:  1991  
Date Detail:
Created Date:  2008-01-28     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8310780     Medline TA:  IEEE Trans Med Imaging     Country:  United States    
Other Details:
Languages:  eng     Pagination:  216-22     Citation Subset:  -    
Affiliation:
Hopital d'Orsay.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Discrimination between healthy and diseased hearts by spectra decomposition of their left ventricula...
Next Document:  An investigation of methods of pileup rejection for 2-D array detectors employed in high resolution ...