Abstract/OtherAbstract:
Pulse oximetry is a widely used technique in biomedical optics, but currently available pulse oximeters rely on empirical calibration approaches, which perform poorly at low saturations. We present an exact solution for pulse oximetry and show how this can be used as the basis for the development of a semiempirical calibration approach that may be useful, especially at low saturations and variable probe geometries. This new approach was experimentally tested against traditional empirical calibration techniques on transmission pulse oximetry for monitoring of fetal sheep using a minimally invasive spiral probe. The results open the way for the development of more accurate pulse oximetry.
Authors:
George Zonios; Uday Shankar; Vijay K Iyer
Related Documents :
18853828 - North american neonatal extracorporeal membrane oxygenation (ecmo) devices and team rol...
7832338 - Cerebral oxygenation during pediatric cardiac surgery using deep hypothermic circulator...
2399508 - Neonatal extracorporeal membrane oxygenation in texas.
18725788 - Perforation of the superior vena cava during ecmo catheterization in two neonates with ...
567458 - Effect of heat treatment of human milk on absorption of nitrogen, fat, sodium, calcium,...
10608908 - Participants' perceptions of dimensions of the therapeutic alliance over the course of ...
Publication Detail:
Type:  Comparative Study; Evaluation Studies; Journal Article; Validation Studies    
Journal Detail:
Title:  IEEE transactions on bio-medical engineering     Volume:  51     ISSN:  0018-9294     ISO Abbreviation:  IEEE Trans Biomed Eng     Publication Date:  2004 May 
Date Detail:
Created Date:  2004-05-10     Completed Date:  2004-06-10     Revised Date:  2009-11-11    
Medline Journal Info:
Nlm Unique ID:  0012737     Medline TA:  IEEE Trans Biomed Eng     Country:  United States    
Other Details:
Languages:  eng     Pagination:  818-22     Citation Subset:  IM    
Affiliation:
Respironics, Inc., Kennesaw, GA 30144, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Calibration / standards
Computer Simulation
Diagnosis, Computer-Assisted / methods*
Equipment Failure Analysis / methods*
Models, Cardiovascular*
Oximetry / instrumentation*,  methods*,  standards
Oxygen / analysis
Quality Control
Reproducibility of Results
Sensitivity and Specificity
Sheep / embryology,  metabolism
Transducers*
Chemical
Reg. No./Substance:
7782-44-7/Oxygen

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Multichannel magnetic stimulation system design considering mutual couplings among the stimulation c...
Next Document:  A closed-loop mechanical ventilation controller with explicit objective functions.