Pulse conductance and flow-induced hemolysis during pulsatile cardiopulmonary bypass.
MedLine Citation:
PMID:  20420610     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
In this study, the hypothesis was tested that a low-resistant, high-compliant oxygenator provides better pulse conductance and less hemolysis than a high-resistant, low-compliant oxygenator during pulsatile cardiopulmonary bypass. Forty adults undergoing coronary artery bypass surgery were randomly divided into two groups using either an oxygenator with a relatively low hydraulic resistance (Quadrox BE-HMO 2000, Maquet Cardiopulmonary AG, Hirrlingen, Germany) or with a relatively high hydraulic resistance (Capiox SX18, Terumo Cardiovascular Systems, Tokyo, Japan). The phase shift between the flow signals measured at the inlet and outlet of the oxygenator was used to assess compliance. Pulse conductance in terms of pressure attenuation was calculated by dividing the outlet pulse pressure of the oxygenator by the inlet pulse pressure. A normalized index was used to assess hemolysis. The phase shifts in time of the flow pulses were 36 +/- 6 ms in the low-resistant (high-compliant) oxygenator, and 14 +/- 2 ms in the high-resistant (low-compliant) oxygenator group (P < 0.001). The low-resistant, high-compliant oxygenator provided 27% better pulse conductance compared with the high-resistant, low-compliant oxygenator (0.84 +/- 0.02 and 0.66 +/- 0.01, respectively, P < 0.001). Inlet pulse pressures were significantly higher (29%) in the high-resistant, low-compliant (Capiox) group than in the low-resistant, high-compliant (Quadrox) group (838 +/- 38 mm Hg and 648 +/- 25 mm Hg respectively, P < 0.001), but no significant difference in hemolysis was found. A low-resistant, high-compliant oxygenator provides better pulse conduction than a high-resistant, low-compliant oxygenator. However, the study data could not confirm the association of high pressures with increased hemolysis.
Authors:
Antoine P Simons; Pieter Wortel; Rob A T van Kan; Frederik H van der Veen; Patrick W Weerwind; Jos G Maessen
Publication Detail:
Type:  Journal Article; Randomized Controlled Trial    
Journal Detail:
Title:  Artificial organs     Volume:  34     ISSN:  1525-1594     ISO Abbreviation:  Artif Organs     Publication Date:  2010 Apr 
Date Detail:
Created Date:  2010-04-27     Completed Date:  2010-07-23     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7802778     Medline TA:  Artif Organs     Country:  United States    
Other Details:
Languages:  eng     Pagination:  289-94     Citation Subset:  IM    
Affiliation:
Department of Cardiothoracic Surgery and Cardiovascular Research Institute Maastricht, Maastricht University Medical Center, Maastricht, The Netherlands. antoinepsimons@ctc.unimaas.nl
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Blood Pressure
Cardiopulmonary Bypass / adverse effects*,  instrumentation*
Coronary Artery Bypass / adverse effects*,  instrumentation*
Equipment Design
Female
Heart Rate
Hemolysis*
Humans
Male
Middle Aged
Oxygenators
Pulsatile Flow