Abstract/OtherAbstract:
|
PURPOSE: The purpose of this work was to compare the extent of pseudoenhancement (artifactual increase in measured attenuation of a simple cyst after contrast medium administration) in a phantom model on single detector and multidetector helical CT scanners. METHOD: The phantom consisted of four water-filled spheres varying in size from 8 to 28 mm, suspended in an aqueous contrast medium bath. Iodine concentration in the bath was varied: 0, 6, 12, and 24 mg/ml corresponding to attenuation values of 0, +108, +180, and +300 HU. The phantom was scanned on single detector and multidetector helical CT scanners during the same session. Collimation (1, 3, and 5 mm) and pitch (1 and 1.5:1, single detector; 3:1 and 6:1, multidetector) were varied at each concentration. All scans were performed at 140 kVp and 170 mA. The region of interest was measured at the center of each sphere. The effects were analyzed using a linear regression model. RESULTS: The degree of pseudoenhancement was more pronounced with increasing iodine concentration, decreasing cyst size, and wider collimation (all p = 0.0001). Pseudoenhancement was also more marked on the multidetector than the single detector scanner (p = 0.0001). At physiological levels of renal enhancement, the average pseudoenhancement was +18 HU for the single detector versus +23 HU for the multidetector scanner. Variation in pitch had no effect. CONCLUSION: Pseudoenhancement is greater on a multidetector than a single detector helical CT scanner and may exceed 20 HU at physiological levels of renal enhancement. |
| | |
Authors:
|
Joan P Heneghan; Audrey L Spielmann; Douglas H Sheafor; Mark A Kliewer; David M DeLong; Rendon C Nelson |
Related Documents
:
|
11455749 - Visual survey of cancer dissemination: classic patterns on helical ct of the neck and t... 1566679 - Assessment of the thoracic aorta by spiral ct. 1457539 - Three-dimensional display of hepatic venous anatomy generated from spiral computed tomo... 7567259 - Reduced need for sedation in patients undergoing helical ct of the chest and abdomen. 17705009 - Improvement in image quality of noncontrast head images in multidetector-row ct by volu... 15789609 - Comparison of helical and cine acquisitions for 4d-ct imaging with multislice ct. 20164559 - Diagnostic utility of novel mri-based biomarkers for alzheimer's disease: diffusion ten... 2006279 - Senile dementia and healthy aging: a longitudinal ct study. 19617599 - Evaluation of radiation doses from mdct-imaging in otolaryngology. |
Publication Detail:
|
Type: Comparative Study; Journal Article |
Journal Detail:
|
Title: Journal of computer assisted tomography Volume: 26 ISSN: 0363-8715 ISO Abbreviation: J Comput Assist Tomogr Publication Date: 2002 Jan-Feb |
Date Detail:
|
Created Date: 2002-01-21 Completed Date: 2002-02-15 Revised Date: 2006-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 7703942 Medline TA: J Comput Assist Tomogr Country: United States |
Other Details:
|
Languages: eng Pagination: 90-4 Citation Subset: IM |
Affiliation:
|
Department of Radiology, Duke University Medical Center, Durham, NC 27710, USA. heneg002@mc.duke.edu |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Contrast Media Humans Image Enhancement Kidney Diseases, Cystic / radiography* Phantoms, Imaging Tomography, X-Ray Computed / methods* |
| Chemical | |
Reg. No./Substance:
|
0/Contrast Media |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Right-sided colonic diverticulitis: CT findings.
Next Document: Radiologic findings of gastrointestinal tract involvement in hepatocellular carcinoma.