Document Detail

Pseudo-no-reflow phenomenon in carotid artery stenting using FilterWire EX: successful recovery by aspiration thrombectomy.
Jump to Full Text
MedLine Citation:
PMID:  19795002     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Distal protection devices such as FilterWire EX have been widely used in carotid artery stenting, however, the large amount of atherothrombotic debris entrapped in the filter could reduce or stop antegrade flow. We present a case of pseudo-no-reflow phenomenon after postdilatation of the stent in a patient with asymptomatic carotid artery stenosis. After several passes using an Export Aspiration catheter, normal flow in the internal carotid artery was restored. Aspiration thrombectomy can successfully recover pseudo-no-reflow phenomenon.
Authors:
Seung Hwan Han; Woong Chol Kang; Tae Hoon Ahn; Eak Kyun Shin
Publication Detail:
Type:  Case Reports; Journal Article     Date:  2009-09-24
Journal Detail:
Title:  Journal of Korean medical science     Volume:  24     ISSN:  1598-6357     ISO Abbreviation:  J. Korean Med. Sci.     Publication Date:  2009 Oct 
Date Detail:
Created Date:  2009-10-01     Completed Date:  2010-01-27     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8703518     Medline TA:  J Korean Med Sci     Country:  Korea (South)    
Other Details:
Languages:  eng     Pagination:  967-9     Citation Subset:  IM    
Affiliation:
Division of Cardiology, Heart Center, Gil Hospital Gachon University of Medicine and Science Incheon, Korea.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Blood Vessel Prosthesis Implantation
Carotid Arteries / radiography,  surgery
Carotid Stenosis / diagnosis,  radiography,  surgery*
Humans
Male
No-Reflow Phenomenon*
Stents
Thrombectomy / instrumentation*
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): J Korean Med Sci
Journal ID (publisher-id): JKMS
ISSN: 1011-8934
ISSN: 1598-6357
Publisher: The Korean Academy of Medical Sciences
Article Information
Download PDF
Copyright ? 2009 The Korean Academy of Medical Sciences
open-access:
Received Day: 24 Month: 1 Year: 2008
Accepted Day: 06 Month: 5 Year: 2008
Print publication date: Month: 10 Year: 2009
Electronic publication date: Day: 24 Month: 9 Year: 2009
Volume: 24 Issue: 5
First Page: 967 Last Page: 969
ID: 2752787
PubMed Id: 19795002
DOI: 10.3346/jkms.2009.24.5.967

Pseudo-no-Reflow Phenomenon in Carotid Artery Stenting using FilterWire EX: Successful Recovery by Aspiration Thrombectomy
Seung Hwan HanA1
Woong Chol KangA1
Tae Hoon AhnA1
Eak Kyun ShinA1
Division of Cardiology, Heart Center, Gil Hospital Gachon University of Medicine and Science Incheon, Korea.
Correspondence: Address for correspondence: Eak Kyun Shin, M.D. Division of Cardiology, Gil Heart Center, Gachon Medical School, 1198 Guwol-dong, Namdong-gu, Incheon 405-760, Korea. Tel: +82.32-460-3665, Fax: +82.32-460-3117, ekshin@gilhospital.com

INTRODUCTION

The interventions for carotid artery stenosis, vein graft and acute myocardial infarction could embolize a large amount of plaque debris, especially unstable, ulcerative, and thrombus containing lesions (1). Distal protection devices such as FilterWire EX have been widely used in carotid artery stenting to prevent distal embolization of plaque debris (2). The large amount of atherothrombotic debris entrapped in the filter could reduce or stop antegrade flow and may be shown as no reflow phenomenon in case using filter type distal protection device.

We present a case of pseudo no-reflow phenomenon after postdilatation of the stent in severe carotid artery stenosis and successful recovery with aspiration thrombectomy using Export Aspiration catheter.


CASE REPORT

A 70-yr-old man was diagnosed with asymptomatic carotid artery stenosis during preoperative risk assessments. He was planned to surgery for rectosigmoid cancer. He was a current smoker and denied history of hypertension and diabetes mellitus. He had been treated in our hospital for congestive heart failure and stable angina for 2 yr. He did not complain of dyspnea or chest pain and had no neurological abnormality.

Carotid angiogram showed no significant stenosis in the left carotid artery and a 90% stenosis in the right internal carotid artery (Fig. 1). Shuttle sheath (7F, Cook Medical, Bloomington, IN, U.S.A.) was positioned to right common carotid artery via right femoral artery. To prevent distal embolization of plaque debris during angioplasty, FilterWire EX (Boston Scientific, Natick, MA, U.S.A.) was positioned in the right cervical internal carotid artery distal to stenotic lesion. After predilation with 4.0 mm balloon, 8 mm self-expanding Wall stent (Boston Scientific) was positioned across the lesion and was deployed in stenotic lesion. After postdilation of the stent with 6.0 mm balloon, carotid angiogram showed abrupt complete flow interruption (Fig. 2A). We suspected this phenomenon induced by obliteration of filter pore by massive atherosclerotic emboli. We decided to perform aspiration thrombectomy using Export Aspiration catheter (Medtronic AVE, Santa Rosa, CA, U.S.A.) (Fig. 2B). After several passes of Export Aspiration catheter into the basket (over the FilterWire EX), antegrade flow was fully restored (Fig. 3). After retrieval of FilterWire Ex, whitish debris was noted within the basket (Fig. 4). After finishing carotid artery stenting, the patient complained of slight speech disturbance and this neurological abnormality was completely recovered within 24 hr after stenting.


DISCUSSION

Obstructive carotid artery lesions are known to contain friable, ulcerated and thrombolic material that have the potential to embolize during intervention and may be responsible for the majority of the neurologic events during carotid artery stenting (1). A number of "distal protection" strategies, designed to capture embolic debris released during carotid intervention, have been evaluated for their efficacy in minimizing the risk of embolic neurologic events. The intravascular filter type distal protection device offers the advantage of a constant cerebral perfusion during carotid artery stenting and allows more time for careful and precise intervention (2). During carotid artery stenting using filter type distal protection device, microscopic particles larger than the size of the filter pores (80 to 130 ?m) could be captured. Reimers et al. (3) reported that carotid artery stenting with filter type distal protection device experienced flow impairment due to filter obstruction in 7.9% of procedures, which was resolved in all cases after filter removal. In our case, complete flow interruption was developed after postdilatation of the stent. We suspected this pseudo-no-reflow phenomenon was induced by obliteration of filter pore by massive atherosclerotic debris. We considered to remove filter by retrieval catheter, however, this process may cause entrapment of filter device to retrieval catheter and embolize massive amount of atherosclerotic debris. Therefore, we choose aspiration thrombectomy using Export Aspiration catheter. After aspiration thrombectomy, flow interruption in the right internal carotid artery was completely restored, therefore, we confirmed this phenomenon as pseudo-no-reflow phenomenon instead of true no-reflow phenomenon. Yadav (4) recommended practically in his review paper if there is no forward blood flow during carotid angioplasty with filter type distal protection device, this should be aspirated before the filter is collapsed. Our case is the first case report of pseudo-no-reflow phenomenon in carotid artery stenting using filter type distal protection device and successful recovery of this phenomenon by aspiration thrombectomy using Export Aspiration catheter. Lee et al. (5) recently reported that successful management of pseudo-no-reflow phenomenon by the Export Aspiration catheter in ostial saphenous vein graft intervention using FilterWire EX protection.

In conclusion, the large amount of atherothrombotic debris entrapped in the filter during carotid intervention could stop antegrade flow. This phenomenon is pseudo-no-reflow phenomenon instead of true no-reflow phenomenon. Aspiration thrombectomy can successfully recover pseudo-no-reflow phenomenon.


References
1. Imparato AM,Riles TS,Gorstein F. The carotid bifurcation plaque: pathologic findings associated with cerebral ischemiaStrokeYear: 197910238245462508
2. Al-Mubarak N,Colombo A,Gaines PA,Iyer SS,Corvaja N,Cleveland TJ,Macdonald S,Brennan C,Vitek JJ. Multicenter evaluation of carotid artery stenting with a filter protection systemJ Am Coll CardiolYear: 20023984184611869850
3. Reimers B,Schl?ter M,Castriota F,T?bler T,Corvaja N,Cernetti C,Manetti R,Picciolo A,Liistro F,Di Mario C,Cremonesi A,Schofer J,Colombo A. Routine use of cerebral protection during carotid artery stenting: results of a multicenter registry of 753 patientsAm J MedYear: 200411621722214969648
4. Yadav JS. Embolic protection devices: methods, techniques, and dataTech Vasc Interv RadiolYear: 2004719019316084443
5. Lee CH,Low A,Tan HC,Lim YT. Pseudo-no-reflow phenomenon in ostial saphenous vein graft intervention using FilterWire EX protectionInt J CardiolYear: 200510423323416168821

Article Categories:
  • Case Report

Keywords: Carotid Artery Stenting, Distal Protection Device, Pseudo-no Reflow Phenomenon, Aspiration Thrombectomy.

Previous Document:  A case of gastritis associated with gastric capillariasis.
Next Document:  Kaposi sarcoma herpes virus-associated hemophagocytic syndrome complicated by multicentric castleman...