Abstract/OtherAbstract:
AIM: To identify whether pulse oximetry technology is associated with decreased retinopathy of prematurity (ROP) and laser treatment.
METHODS: Inborn infants <1250 g who had eye exams were compared at two centres in three periods. In Period 1, SpO₂ target was ≥93% and pulse oximetry technology was the same in both Centres. In Period 2, guidelines for SpO₂ 88-93% were implemented at both centres and Centre B changed to oximeters with signal extraction technology (SET(®)) while Centre A did not, but did so in Period 3. One ophthalmology department performed eye exams using international criteria.
RESULTS: In 571 newborns <1250 g, birth weight and gestational age were similar in the different periods and centres. At Centre A, severe ROP and need for laser remained the same in Periods 1 and 2, decreasing in Period 3-6% and 3%, respectively. At Centre B, severe ROP decreased from 12% (Period 1) to 5% (Period 2) and need for laser decreased from 5% to 3%, remaining low in Period 3.
CONCLUSION: In a large group of inborn infants <1250 g, a change in clinical practice in combination with pulse oximetry with Masimo SET, but not without it, led to significant reduction in severe ROP and need for laser therapy. Pulse oximetry selection is important in managing critically ill infants.
Authors:
Armando Castillo; Richard Deulofeut; Ann Critz; Augusto Sola
Related Documents :
15827664 - Incidence, risk factors of retinopathy of prematurity among very low birth weight infan...
484624 - Cigarette smoking as an etiologic factor in cleft lip and palate.
4038594 - Histopathology of otitis media in infants with cleft and high-arched palates.
17440954 - Ultrastructure of the protonephridial system in neodasys chaetonotoideus (gastrotricha:...
23329914 - Maternal progesterone level in fetal growth restriction and its relationship with doppl...
3373424 - Effects of vehicles and prodrug properties and their interactions on the delivery of 6-...
Publication Detail:
Type:  Journal Article; Multicenter Study; Research Support, Non-U.S. Gov't     Date:  2010-10-15
Journal Detail:
Title:  Acta paediatrica (Oslo, Norway : 1992)     Volume:  100     ISSN:  1651-2227     ISO Abbreviation:  Acta Paediatr.     Publication Date:  2011 Feb 
Date Detail:
Created Date:  2011-01-11     Completed Date:  2011-05-03     Revised Date:  2013-05-28    
Medline Journal Info:
Nlm Unique ID:  9205968     Medline TA:  Acta Paediatr     Country:  Norway    
Other Details:
Languages:  eng     Pagination:  188-92     Citation Subset:  IM    
Copyright Information:
© 2010 The Author(s)/Acta Paediatrica © 2010 Foundation Acta Paediatrica.
Affiliation:
Northeast Georgia Medical Center, Gainesville, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Female
Humans
Infant, Newborn
Male
Oximetry*
Oxygen / analysis
Prospective Studies
Retinopathy of Prematurity / diagnosis,  prevention & control*
Severity of Illness Index
Chemical
Reg. No./Substance:
7782-44-7/Oxygen
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Acta Paediatr
Journal ID (publisher-id): apa
ISSN: 0803-5253
ISSN: 1651-2227
Publisher: Blackwell Publishing Ltd
Article Information
Download PDF
Acta Pædiatrica © 2011 Foundation Acta Pædiatrica
open-access:
Received Day: 15 Month: 7 Year: 2010
Revision Received Day: 19 Month: 8 Year: 2010
Accepted Day: 26 Month: 8 Year: 2010
Print publication date: Month: 2 Year: 2011
Volume: 100 Issue: 2
First Page: 188 Last Page: 192
ID: 3040295
PubMed Id: 20825604
DOI: 10.1111/j.1651-2227.2010.02001.x

Prevention of retinopathy of prematurity in preterm infants through changes in clinical practice and SpO2 technology
Armando Castillo1
Richard Deulofeut2
Ann Critz3
Augusto Sola4
1Northeast Georgia Medical CenterGainesville, GA, USA
2Neonatal Service, Pediatrix Medical GroupDallas, TX, USA
3Division of Neonatal- Perinatal Medicine, Emory UniversityAtlanta, GA, USA
4Iberoamerican Society of NeonatologyMorristown, NJ, USA
Correspondence: Augusto Sola MD, 20 Egbert Hill Road, Morristown, NJ 07960, USA.Tel.: +1 862 812 8846 | Fax: +1 949 297 7398 | Email: augustosolaneo@gmail.com
Re-use of this article is permitted in accordance with the Terms and Conditions set out at http://wileyonlinelibrary.com/onlineopen#OnlineOpen_Terms
(augustosolaneo@gmail.com)

Introduction

Retinopathy of prematurity (ROP) is a devastating and common blinding disease in developed and developing countries (1,2). However, most cases of ROP are mild and regress spontaneously. The incidence of this disease has been reported to be as high as 29% in very low birth weight infants (VLBW) and is strongly associated with smaller and sicker infants (35). There are many risk factors involved in the pathogenesis of ROP (58), and two phases of abnormal vessel development have been described in ROP. Phase I is characterized by delayed retinal vascular growth and partial regression of existing vessels. Phase II is characterized by pathological vessel growth associated with hypoxia (2). The use of oxygen and fluctuations of arterial oxygen are considered major risk factors for the development severe ROP in preterm infants. (2,6,8). We and others have shown that lower saturation targets are associated with improved rates of ROP (914) and that there is a treatment-by-gender effect when aiming to avoid hyperoxia in preterm infants in the Neonatal Intensive Care Unit (NICU) (15). Strategies to lower the incidence of ROP include the following: education and commitment of bedside care providers, guidelines to decrease hyperoxemic periods and wide changes in oxygenation and the advances in saturation monitoring technology. The impact of each of these interventions on the incidence of ROP is not easy to discriminate. In the last decade, it has become clear that performance of pulse oximeters differ by brand, with some brands having a much higher incidence of false alarms and holding periods and reduced reliability (1621). Of the next generation pulse oximetry technologies, only one has signal extraction technology (SET), which has been shown to have fewer false alarms and missed events in challenging clinical situations such as during motion and low perfusion. (2226). However, no studies have explored the role that SpO2 technology plays in the prevention or ROP. We designed this study to determine whether the incidence of severe ROP is associated with the type pulse oximetry technology utilized in the NICU, when oxygen saturation limits are aimed to be maintained between 88 and 93% after education and guidelines are implemented. Our hypothesis was that the relative risk reduction in severe retinopathy of prematurity is associated with the saturation technology utilized in neonatal intensive care units.


Patients and Methods

After Internal Review Board approval, this observational descriptive study was conducted in two centres at Emory University during two consecutive 3 -year periods (Periods 1 and 2) and an 18- month follow-up period (Period 3). Centre A was Grady Memorial Hospital, and Centre B was Emory Crawford Long Hospital. Table 1 shows a schematic representation of the study design. Inborn infants, with birth weight <1250 g that had detailed ROP examinations by an ophthalmologist from Emory University following AAP and AAO guidelines (27), were included in this study. The exclusion criteria were as follows: newborns with major congenital anomalies, outborn infants, infants with birth weight ≥1250 g and those who had no ophthalmologic examination.

During the first period, oxygen saturation targets were between 92% and 100%, and the pulse oximeters used at both centres were Nellcor N-300 and Nellcor N-395, (Nellcor®, Covidien, Boulder, CO, USA). As part of an intense quality improvement effort, an educational process was initiated for oxygen utilization and monitoring in the two centres. This led to guideline development and implementation, as described previously (9,10). In Period 2, the new practice guidelines for lower oxygen saturation targets of 88–93% for premature infants receiving supplemental oxygen were implemented at both centres (Table 1). During Period 2, the pulse oximeters were changed to Masimo SET® (Masimo, Irvine, CA, USA) in Centre B while Centre A continued to use the Nellcor® technology. During Period 3, both centres continued to practise the guidelines for oxygen saturation targets and Centre A changed to pulse oximetry with SET so that both centres were using SET technology. During all three periods, the medical leadership, attending neonatologists, neonatal fellows and neonatal nurse practitioners were the same in both centres and followed the same guidelines for the care of neonates (Table 1). Of course, other improvements other than oxygen management did occur over the study period, but they affected at all infants regardless of the type of monitoring used.

Data analyses and statistics

In both centres, the data were collected prospectively and entered into a single, combined database by the same personnel for subsequent analysis. The data for all infants in both centres were also reported to Vermont Oxford Network (VON). Statistics included chi square, Student’s t-test and ANOVA when appropriate. Only infants with detailed ROP examinations were used in the denominator to calculate the incidence of severe ROP and need for laser therapy in all periods and centres. All the statistical analyses were carried out using the same software (SPSS version 15.0 for Windows; SPSS, Chicago, IL, USA).


Results

During the three study periods, there were 774 newborns with BW <1250 g admitted to the study centres. Of them, 571 had eye exams and met all the inclusion criteria. From Centre A, there were 138 infants included during Period 1, 113 during Period 2 and 65 during Period 3. From Centre B, there were 83 infants included in Period 3, 115 infants in Period 2 and 57 in Period 3. There were no significant differences in patient characteristics between the two centres. Table 2 summarizes some of the characteristics of the patient population. During the three study periods, we did not identify any child who suffered from stage V ROP in either of the centres.

The rate of ROP III and IV and laser treatment in examined infants with birth weight <1250 g in Centre B were 12% and 5% in Period 1 and significantly decreased to 5% and 3% in Period 2 following the change to SET technology (p < 0.05) (Table 3). Based on this decrease in the incidence of severe ROP in Centre B, the relative risk reduction can be calculated at 58%, with a number needed to treat of 14. In Centre B, the incidence remained low in Period 3, 4% and 2%, respectively. In Centre A in Period 1, ROP III–IV and laser rates were 13% and 4.5%, respectively, and did not decrease in Period 2. In Period 3, following the change to SET technology, ROP III–IV decreased to 6% with a relative risk reduction of 54% (p < 0.05), and laser rates decreased to 3%.

We also examined the individual incidence of severe ROP for each of the years of Periods 1 and 2 at each centre. As expected, there was variability between each year. At Centre A, the incidence of ROP III–IV varied between 15% and 10% during the years of Period 1 and was relatively stable with an average of 13% per year during Period 2 while the technology was not changed. Similarly, at Centre B, the incidence of severe ROP varied during the years of Period 1 (10–13%). Unlike Centre A, severe ROP decreased each year during Period 2, starting at 10% in the first year and decreasing to 3% in the last year of the period, The rates of laser surgery treatment for severe ROP in Centre A decreased during Period 3 while at Centre B the rates decreased during Period 2 and remained as low during Period 3. With ‘lower’ oxygen saturation targets of 88–93%, the mortality was similar and there was no increase in periventricular leucomalacia; the rate of bronchopulmonary displasia was improved, and the incidence of necrotizing enterocolitis and patent ductus arteriosus did not increase. Additionally, the longer term outcome was better in the period with lower targets in the infants available for follow up (10).


Discussion

This study confirms that while ROP is a potentially serious condition, it can be reduced in at risk newborns. The findings of this study show a significant decrease in the incidence of severe ROP and need for laser surgery during the period after education and practice changes when SET pulse oximetry technology was used. Furthermore, the lower incidence of ROP following the practice changes and change in pulse oximetry was progressively more significant year after year. In this study of 571 infants, none was blinded by stage V ROP and only about ≤5% needed treatment to prevent blindness.

Among the most important ROP, prevention efforts are to decrease wide changes in oxygenation and hyperoxemic periods (916,28). At the two centres studied, the same intense educational process and clinical guidelines aimed to decrease hyperoxemia, and wide changes in oxygenation were implemented. In 1992, Flynn et al. were the first to describe an association between the incidence and severity of retinopathy of prematurity and the duration of exposure to arterial oxygen levels of 80 mmHg or higher, measured transcutaneously (29). As we have shown that saturations >94% are likely to be associated with PaO2 > 80 mmHg in infants breathing supplemental oxygen (30), the target values were to avoid those levels of saturation. The leadership, physicians and nurse practitioners were the same at both centres during the three study periods. The pulse oximetry technology changed to SET technology at one centre after Period 1 and stayed the same at the other centre but then was changed at the second centre during Period 3. The findings of this study show a strong positive association between the use of pulse oximetry with SET and a reduction in the incidence of ROP.

Many factors contribute to good medical care and outcome improvement. Among them are education, training, commitment, team work, eradicating bad practices, and the correct use of the best available technology. In general, it is well accepted in most areas of medical care that technology by itself, without education and appropriate use, serves very little value for care improvement. On the other hand, well-trained committed clinicians obtain worse outcomes when using technology that is obsolete, outdated or of lesser quality. The findings clearly suggests that, when all other interventions are the same, an educational process and improved care guidelines along with a change to improved pulse oximetry technology are useful for reducing the incidence of severe neonatal ROP and need for laser surgery.

In contrast to the interventional devices and therapies, there is a tendency by some clinicians to believe that non-therapeutic monitoring devices with similar functionality deliver similar performance. However, as clinicians, we all understand that the decisions we make, right or wrong, are often based on the numbers shown on patient monitors. Therefore, the accuracy and reliability of monitored physiologic information can be just as important, if not more important, than the therapeutic interventions that follow. In this study, we have shown that use of one type of pulse oximetry technology appears to strongly affect the incidence of ROP.

We acknowledge that there are some limitations in this study. This was not a randomized controlled trial, so the results cannot definitively isolate the pulse oximetry technology as the major only factor contributing to decreasing incidence of ROP. However, it is unlikely that such a study could ever be performed in a randomized way. In addition, while the physicians and nurse practitioners were the same at both centres, the bedside nurses and respiratory therapists were different. As the rates of severe ROP were significantly lower at Centre B compared to Centre A in period 2 (Table 3) in association with implementation of SET technology and a reduction in severe ROP and need for laser surgery occurred in Centre A in Period 3, when change of technology occurred, it is unlikely that bedside nurses and respiratory therapists played a major role in the different outcomes during the different periods. Nonetheless, the significance or ‘weight’ of different bedside health care providers cannot be separately discriminated in this study.

Finally, a very recent article reports markedly decrease ROP but worse mortality when saturation targets are 85–89% in very tiny infants. In the current and in all other studies published to date, no one has aimed for a target of 85–89%. Therefore, we cannot compare or comment on those findings. Further data are waited from other large ongoing randomized clinical trials.


Conclusions

In this study of inborn infants with birth weight <1250 g treated by the same physicians and nurse practitioners using the same clinical guidelines aimed at decreasing hyperoxemia and wide changes in oxygenation, a reduction in the incidence of severe ROP and need for laser therapy were associated with the use of signal extraction pulse oximetry. The findings lend further support to the significance of using improved saturation monitors in managing critically ill infants. Education, guidelines and practice changes for reducing ROP may be more effective when accurate and reliable pulse oximetry technology is used.


Acknowledgement and Disclosures

We thank the Goddard Scholarship at Emory University for the support during the period of this study. We thank the neonatologists, fellows, NNPs, RRTs and RNs for all their effort in improving quality of care and the professionals that put tireless effort to accurately collect neonatal data. No industry sponsored the study. The study design was developed by the authors. None of the authors owe any stock in the companies mentioned. As of July 2010, Dr Sola consults for Masimo in neonatal and paediatric research and development. Dr Sola lectures on oxygenation, ROP and monitoring around the USA, Latin America and Europe. If such activity is sponsored by Masimo, Dr Sola does not collect the assigned honorarium, which Masimo donates to an educational not for profit organization (SIBEN).


References
1. Tasman W,Patz A,McNamara JA,Kaiser RS,Trese MT,Smith BT. Retinopathy of prematurity: the life of a lifetime diseaseAm J OphthalmolYear: 20061411677416386993
2. Chen J,Smith LE. Retinopathy of prematurityAngiogenesisYear: 2007101334017332988
3. Shah VA,Yeo CL,Ling YL,Ho LY. Incidence, risk factors of retinopathy of prematurity among very low birth weight infants in SingaporeAnn Acad Med SingaporeYear: 2005341697815827664
4. Delport SD,Swanepoel JC,Odendaal PJ,Roux P. Incidence of retinopathy of prematurity in very-low-birth-weight infants born at Kalafong Hospital, PretoriaS Afr Med JYear: 2002929869012561416
5. Saunders RA,Donahue ML,Christmann LM,Pakalnis AV,Tung B,Hardy RJ,et al. Racial variation in retinopathy of prematurity. The Cryotherapy for Retinopathy of Prematurity Cooperative GroupArch OphthalmolYear: 199711560489152127
6. Smith LE. Pathogenesis of retinopathy of prematuritySemin NeonatolYear: 200384697315001119
7. Karna P,Muttineni J,Angell L,Karmaus W. Retinopathy of prematurity and risk factors: a prospective cohort studyBMC PediatrYear: 2005285
8. York JR,Landers S,Kirby RS,Arbogast PG,Penn JS. Arterial oxygen fluctuation and retinopathy of prematurity in very-low-birth-weight infantsJ PerinatolYear: 20042482714762452
9. Chow LC,Wright KW,Sola A,for the CSMC Oxygen Administration Study GroupCan changes in clinical practice decrease the incidence of severe retinopathy of prematurity in very low birth weight infants?PediatricsYear: 20031113394512563061
10. Deulofeut R,Critz A,Adams-Chapman I,Sola A. Avoiding hyperoxia in infants ≤ 1,250 g is associated with improved short- and long-term outcomesJ PerinatolYear: 200626700517036032
11. Wright KW,Sami D,Thompson L,Ramanathan R,Joseph R,Farzavandi S. A physiologic reduced oxygen protocol decreases the incidence of threshold retinopathy of prematurityTrans Am Ophthalmol SocYear: 2006104788417471328
12. Sears JE,Pietz J,Sonnie C,Dolcini D,Hoppe G. A change in oxygen supplementation can decrease the incidence of retinopathy of prematurityOphthalmolYear: 20091165138
13. Chen ML,Guo L,Smith LE,Dammann CE,Dammann O. High or low oxygen saturation and severe retinopathy of prematurity: a meta-analysisPediatricsYear: 2010125148392
14. SUPPORT Study GroupTarget ranges of oxygen saturation in extremely preterm infantsNEJMYear: 201036219596920472937
15. Deulofeut R,Golde D,Augusto S. Treatment-by-gender effect when aiming to avoid hyperoxia in preterm infants in the NICUActa PaediatrYear: 200796990417577339
16. Ahlborn V,Bohnhorst B. False alarms in very low birthweight infants: comparison between three intensive care monitoring systemsActa PaediatrYear: 200089571610852195
17. Bohnhorst B,Peter CS,Owitz CF. Pulse oximeters’ reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximetersCrit Care MedYear: 2000281565810834712
18. Malviya S,Reynolds PI,Voepel-Lewis T,Siewert M,Watson D,Tait AR,et al. False alarms and sensitivity of conventional pulse oximetry versus the Masimo SET technology in the pediatric postanesthesia care unitAnesth AnalgYear: 20009013364010825316
19. Poets CF,Urschitz MS,Bohnhorst B. Pulse oximetry in the neonatal care unit. Detection of hyperoxemia and false alarm ratesAnesth AnalgYear: 200294SupplS41311900037
20. Hay WW,Rodden DJ. Reliability of conventional and new oximetry in neonatal patientsJ PerinatolYear: 200222360612082469
21. Durbin CG,Rostow SK. Advantages of new technology pulse oximetry with adults in extremisAnesth AnalgYear: 20021Suppl813
22. Durbin CG,Rostow SK. More reliable oximetry reduces the frequency of arterial blood gas analysis and hastens oxygen weaning after cardiac surgery: a prospective, randomized trial of the clinical impact of the new technologyCrit Care MedYear: 20023017354012163785
23. Murthy LCT,Goyal RM. Masimo – a new reliable noninvasive method of detecting oxygen saturation in critically illIndian J AnaesthYear: 2005491336
24. Torres A,Skender KM. Pulse oximetry in children with congenital heart disease: effects of cardiopulmonary bypass and cyanosisJ Intensive Care MedYear: 2004192293415296623
25. Sahni R,Gupta A,Ohira-Kist K,Rosen TS. Motion resistant pulse oximetry in neonatesArch Dis Child Fetal Neonatal EdYear: 200388F505814602699
26. Kawagishi T,Kanaya N,Nakawama M,Kurosawa S,Namiki A. A comparison of the failure times of pulse oximeters during blood pressure cuff-induced hypo perfusion in volunteersAnesth AnalgYear: 200499793615333413
27. American Academy of PediatricsSection on Ophthalmology, American Academy of Ophthalmology, and American Association for Pediatric Ophthalmology and Strabismus. Screening examination of premature infants for retinopathy of prematurityPediatricsYear: 20011088091111533356
28. Sola A,Saldeño YP,Favareto V. Clinical practices in neonatal oxygenation: where have we failed? What can we do?J PerinatolYear: 200828Suppl 1S283418446174
29. Flynn JT,Bancalari E,Snyder ES. A cohort study of transcutaneous oxygen tension and the incidence and severity of retinopathy of prematurityNEJMYear: 1992326105041549150
30. Castillo A,Sola A,Baquero H,Neira F,Alvis R,Deulofeut R,et al. Pulse oxygen saturation levels and arterial oxygen tension values in newborns receiving oxygen therapy in the neonatal intensive care unit: is 85% to 93% an acceptable range?PediatricsYear: 2008121882918450890

Tables
[TableWrap ID: tbl1] Table 1 

Schematic description of the study design


Period I3 years2001–2003 Period II3 years2004–2006 Period III18 months2007–2008
SpO2 Range SpO2 ≥92% SpO2 88–93% SpO2 88–93%
Pulse Oximetry Type Centres A and B: Nellcor® Centre A: Nellcor®Centre B: Masimo SET Centres A and B: Masimo SET®
Practice Intervention None Educational process, guidelines development and implementation
Practitioners and data Leadership and health care team members: same (but for RN's and RRT's; see text). Data from the two centres collected prospectively into same data base

[TableWrap ID: tbl2] Table 2 

Demographics of the studied population


A B


Centres Period (n) Period 1 (138) Period 2 (113) Period 3 (65) Period 1 (83) Period 2 (115) Period 3 (57)
Gestational weeks 27 ± 2 27 ± 2.3 27 ± 1.6 26.8 ± 2.1 26.8 ± 2.2 27 ± 2.4
Birth Weight (g) 922.8 ± 190 914 ± 208 907 ± 183 889.6 ± 203 866 ± 198 897 ± 206
Antenatal Steroids 78% 79% 80% 64% 77% 81%
Female 56% 51% 53% 55% 53% 52%
Mechanical Ventilation 92% 83% 86% 85% 81% 83%
Small for gestation 21% 21% 20% 21% 23% 21%

No significant statistical significance was identified for any variable.


[TableWrap ID: tbl3] Table 3 

Rates of retinopathy of prematurity (ROP) and laser surgery for each centre in each period in examined infants <1250 g


Centre A Centre B


Period 1 Period 2 Period 3 Period 1 Period 2 Period 3
ROP III–IV 13% 13%**,† 6%† 12%¶ 5%**,¶ 4%
LASER 4.5% 5% 3% 5%* 3%* 2%

All comparisons (†;**; ¶; *): p < 0.05.



Article Categories:
  • Regular Articles
    • Oxygen

Keywords: Hyperoxemia, Laser surgery, Newborns, Oxygen therapy, Prevention, Pulse oximetry saturation, Retinopathy of prematurity (ROP), Signal extraction technology (SET).

Previous Document:  Motor development curve from 0 to 12?months in infants born preterm.