Pressure volume parameters for optimal function of an intraaortic balloon without a pump.
MedLine Citation:
PMID:  3870575     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
An air filled intraaortic balloon connected to an extravascular air chamber (compression chamber system-CCS) was used in 10 anesthetized dogs to increase its stroke volume (SV). Highest SV increases (up to over 200%) were observed at low SV states, with a high CCS volume and with high increases of the systolic pressure. The CCS pressure--systolic pressure relationship was a sigmoid-like curve. In conclusion, highest SV's may be achieved at low SV states, using CCS volumes about 400 ml and the highest CCS pressure that may cause a linear aortic pressure increase.
Authors:
D A Sideris; D N Chrysos; A Pras; L K Michalis; G K Maliaras; S D Moulopoulos
Related Documents :
21851345 - Anaesthesia for 1131 patients undergoing proximal femoral fracture repair: a retrospect...
22033055 - Physiologic determinants of prolonged mechanical ventilation in patients after major su...
9486295 - Cardiac afferents play the dominant role in renal nerve inhibition elicited by volume e...
21966215 - Association between vitamin d and pressure ulcers in older ambulatory adults: results o...
971685 - The physiology and pathophysiology of the ureter.
7602605 - The rapid expression of myocardial hsp 70 mrna and the heat shock 70 kda protein can be...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Life support systems : the journal of the European Society for Artificial Organs     Volume:  3 Suppl 1     ISSN:  0261-989X     ISO Abbreviation:  Life Support Syst     Publication Date:  1985  
Date Detail:
Created Date:  1987-11-30     Completed Date:  1987-11-30     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  8219681     Medline TA:  Life Support Syst     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  259-63     Citation Subset:  IM    
Affiliation:
Department of Clinical Therapeutics, Athens University, Alexandra Hospital, Greece.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Blood Pressure
Dogs
Evaluation Studies as Topic
Intra-Aortic Balloon Pumping / instrumentation*
Stroke Volume

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The spindle pump--a new concept for a nonpulsatile bloodpump.
Next Document:  Effect of haemodialysis on coronary occlusion-reperfusion ventricular fibrillation.