Predictive factors for pacemaker requirement after transcatheter aortic valve implantation.
Jump to Full Text
MedLine Citation:
PMID:  23035864     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
ABSTRACT: BACKGROUND: Transcatheter aortic valve implantation (TAVI) has been established as a treatment option for inoperable patients with symptomatic aortic valve stenosis. However, patients suffer frequently from conduction disturbances after TAVI. METHODS: Baseline, procedural as well as surface and intracardiac ECG parameters were evaluated for patients treated with TAVI and a comparison between patients requiring pacemaker with those not suffering from relevant conduction disorders were done. RESULTS: TAVI was successfully in all patients (n=45). Baseline surface and intracardiac ECG recording revealed longer PQ (197.1+/-51.2 msec versus 154.1+/-32.1 msec; p<0.001), longer AH (153.6+/-43.4 msec versus 116.1+/-31.2 msec; p<0.001) and HV interval (81.7+/-17.8 msec versus 56.8+/-8.5 msec; p<0.001) in patients with need for a pacemaker (n=23) versus control group (n=22); furthermore, 7-day follow-up analysis showed a higher prevalence of new left bundle branch block (LBBB) (87.0% versus 31.9%; p<0.001). Multivariate analysis revealed that only new LBBB, QRS duration >120 msec and a PQ interval >200 msec immediately (within 60 minutes) after implantation of the aortic valve were predictors for high-grade (type II second-degree and third-degree) AV block. Other clinical parameters as well as baseline electrocardiographic parameters had no impact on critical conduction delay. CONCLUSION: Cardiac conduction disturbances are common after TAVI. The need for pacing after TAVI is predictable by surface ECG evaluation immediately (within 60 minutes) after the procedure.
Authors:
Ibrahim Akin; Stephan Kische; Lylia Paranskaya; Henrik Schneider; Tim C Rehders; Ulrich Trautwein; Gökmen R Turan; Dietmar Bänsch; Olga Thiele; Dimitar Divchev; Ilkay Bozdag-Turan; Jasmin Ortak; Gunther Kundt; Christoph A Nienaber; Hüseyin Ince
Related Documents :
9632334 - An experimental investigation of the safe distance for internal orbital dissection.
23072254 - Sperm with large nuclear vacuoles and semen quality in the evaluation of male infertility.
21730254 - Effect of biomimetic remineralization on the dynamic nanomechanical properties of denti...
23253844 - Evidence for survival of implants placed into infected sites is limited.
20386664 - Cytotoxic effects of 2-bromopropane on embryonic development in mouse blastocysts.
11745514 - Long-term effects of four extraction media on the fluoride release from four polyacid-m...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-10-4
Journal Detail:
Title:  BMC cardiovascular disorders     Volume:  12     ISSN:  1471-2261     ISO Abbreviation:  BMC Cardiovasc Disord     Publication Date:  2012 Oct 
Date Detail:
Created Date:  2012-10-5     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  100968539     Medline TA:  BMC Cardiovasc Disord     Country:  -    
Other Details:
Languages:  ENG     Pagination:  87     Citation Subset:  -    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): BMC Cardiovasc Disord
Journal ID (iso-abbrev): BMC Cardiovasc Disord
ISSN: 1471-2261
Publisher: BioMed Central
Article Information
Download PDF
Copyright ©2012 Akin et al.; licensee BioMed Central Ltd.
open-access:
Received Day: 7 Month: 7 Year: 2012
Accepted Day: 28 Month: 9 Year: 2012
collection publication date: Year: 2012
Electronic publication date: Day: 4 Month: 10 Year: 2012
Volume: 12First Page: 87 Last Page: 87
PubMed Id: 23035864
ID: 3509406
Publisher Id: 1471-2261-12-87
DOI: 10.1186/1471-2261-12-87

Predictive factors for pacemaker requirement after transcatheter aortic valve implantation
Ibrahim Akin1 Email: ibrahim.akin@med.uni-rostock.de
Stephan Kische1 Email: skische@uni-rostock.de
Lylia Paranskaya1 Email: lparanksaya@uni-rostock.de
Henrik Schneider1 Email: hschneider@uni-rostock.de
Tim C Rehders1 Email: tcrehders@uni-rostock.de
Ulrich Trautwein1 Email: utrautwein@uni-rostock.de
Gökmen Turan1 Email: grturan@uni-rostock.de
Dietmar Bänsch1 Email: dbaensch@uni-rostock.de
Olga Thiele1 Email: othiele@uni-rostock.de
Dimitar Divchev1 Email: ddivchev@uni-rostock.de
Ilkay Bozdag-Turan1 Email: ibturan@uni-rostock.de
Jasmin Ortak1 Email: jortak@uni-rostock.de
Gunther Kundt2 Email: gkundt@uni-rostock.de
Christoph A Nienaber1 Email: canienaber@uni-rostock.de
Hüseyin Ince1 Email: hince@uni-rostock.de
1Heart Center Rostock, Department of Internal Medicine I, University Hospital Rostock, Rostock School of Medicine, Ernst-Heydemann-Str. 6, Rostock, 18057, Germany
2Institute for Biostatistics and Information in Medicine and Ageing Research, University Hospital Rostock, Rostock School of Medicine, Ernst-Heydemann-Str. 6, Rostock, 18057, Germany

Background

Transcatheter aortic valve implantation (TAVI) has emerged as an alternative to open surgery in patients with symptomatic aortic valve stenosis considered at high risk or unsuitable for thoracotomy [1,2]. Atrioventricular (AV) block with subsequent pacemaker requirement was described in 6% of cases after surgical aortic valve replacement, but varies after TAVI between 5.7% and 42.5%, while new left bundle branch block (LBBB) occurs in up to 50-70% [2-6]. Moreover, early prediction of pacemaker requirement after TAVI would be of potential benefit by shortening both postoperative monitoring and hospitalization. We compared both clinical parameters as well as surface and intracardiac ECG recordings prior and after TAVI and analyzed differences in patients eventually requiring permanent pacing after TAVI for relevant conduction delay.


Methods
Patients

In our retrospective analysis, between January 2007 and January 2008 45 patients without previously known pacemaker who underwent TAVI using the third-generation percutaneous self-expanding CoreValve prosthesis (Medtronic Inc; Minneapolis, USA) were identified for this analysis. The criteria for inclusion and exclusion to the TAVI procedure have been described elsewhere [1-4]. In brief, patients were included with echocardiographic measurements of severe native valvular stenosis and an area < 1 cm2, or < 0.6 cm2/m2 regardless of adjunct valvular regurgitation; a diameter of basal orifice of the stenosed valve between 20 and 27 mm; and a diameter at the sinutubular junction ≤ 43 mm. Most importantly, all patients were considered unfit for open surgery with an EuroSCORE ≥ 20% [7]. TAVI was suggested in agreement between a cardiac surgeon and both, a clinical and interventional cardiologist; patient`s or referring physician`s preference was not relevant [8,9]. Treatment strategy was in compliance with the Helsinki Declaration. Our local ethics committee approved this treatment strategy. Patients gave informed consent prior TAVI. Pacemaker implantation at follow-up was considered indicated in case of complete AV-block, type II second-degree AV block, and in presence of new LBBB in combination with HV prolongation ≥ 75 msec. We compared patients requiring a new pacemaker within 7 days with those not requiring any pacemaker and no significant conduction delay after TAVI.

Procedure

Details of the implantation technique have been described elsewhere [1-4]. In brief, all procedures were performed in an interventional hybrid suite under general anaesthesia to assure stable hemodynamic and minimize patient movement. TAVI was performed via femoral access under fluoroscopy; the aortic valve was initially dilated using a standard valvuloplasty balloon with a nominal diameter similar to the aortic valve and followed by CoreValve (Medtronic Inc.) insertion [1,3-6]. In each patient, prior to TAVI and aside a right ventricular bipolar pacing lead (Pacel™, St. Jude Medical, St Paul, MN, USA), a 6F quadripole electrode catheter with ring electrodes (5-mm interpole distance) (Webster D™, Biosense Webster, Diamond, USA) was introduced and advanced to the His bundle. A 6F quadripolar electrode catheter (Soloist™, Medtronic, Minneapolis, MN, USA) was advanced to the right atrium and another to the right ventricle to record a bipolar electrogram and for programmed atrial and ventricular stimulation. The access sites for all electrophysiologic electrode catheters were the femoral veins. With such instrumentation the sinus node recovery time (SNRT), corrected sinus node recovery time (c-SNRT), antegrade and retrograde effective refractory period (ERP) of the AV node, as well as intracardiac conduction times (atrium to His and His to ventricle time; AH and HV interval) were assessed. The rational to measure SNRT was the observations of sinus node arrest in a single patient receiving a CoreValve prior this study. Thus, we wanted to avoid oversee sinus node pathology in this elderly patients suffering from a high comorbidity index. All measurements were done on an Axiom Sensis™ (Siemens, Erlangen, Germany) electrophysiology workstation prior to the valve implantation, immediately after valve implantation and at a follow-up of 7 days. The AV nodal ERP was measured by introducing a single extrastimulus (S2) after a drive train of 8 stimuli at a fixed rate (S1) (600ms), at which time the S1-S2 interval is decreased until the S2 impulse does not conduct to the His bundle. To assess conduction disorders, patients were attached to uninterrupted ECG monitoring using the Philips monitoring system (IntelliVue™, Best, The Netherlands) that is installed at our ICU/IMC unit. All patients were prophylactically given a temporary pacemaker via the existing femoral venous access; with VVI mode the active pacing was 60 bpm for at least 24 hours.

Statistical methods

All data were stored and analysed using SPSS statistical package 17.0 (Chicago, Illinois, USA). The statistics computed included mean and standard deviations of continuous variables and are presented as means ± SD, frequencies and relative frequencies of categorical factors. The Fisher exact test and chi2-test were used to compare proportions. A normal distribution of differences was confirmed by the Kolmogorov-Smirnov test; in presence of normally distributed data the t-test was used while with non-normal distribution the Mann–Whitney U-test was applied. The logistic regression model was used to assess the interdependence of high-grade AV block (type II second-degree and third-degree) at 7 day follow-up from prognostic factors. First, univariate analyses were performed to reveal unadjusted significant associations between prognostic variables and high-grade AV block. Thereafter, variables yielding p-values ≤ 0.25 with univariate analysis were entered in the multivariate model to assess adjusted associations between outcomes and covariates which were univariate and of borderline significance. All p-values resulted from two-sided statistical tests and ≤ 0.05 was considered to be significant.


Results

The analysis included 18 men and 27 women at a mean age of 81.5±6.8years with severe aortic valve stenosis and no previously known pacemaker. All patients had qualified for TAVI following recent recommendations [8,9]. We compared patients requiring new pacemaker implantation during hospital stay (n=23) with those not suffering from relevant conduction delays (n=22) in the attempt to elucidate predictors for pacing. Smokers (34.8% versus 18.2%; p=0.024) as well as patients suffering from angina, with ischemic heart disease, previous stroke or peripheral arterial disease were more frequent in the pacemaker group. Moreover, logistic EuroSCORE operative mortality estimate was higher in patients requiring a pacemaker compared to the non-pacemaker group (24.0±14.6% versus 18.4±19.2%; p=0.034). Echocardiographic measurements on aortic stenosis severity and left ventricular function as well as intraoperative data revealed similar findings in both groups (Table 1).

TAVI was successfully performed in all patients with a mean procedural time of 107.7±27.1 min and 104.9±33.3 min (p=0.725); length of stay in intensive care unit and in-hospital stay was similar in both groups. Six patients suffered from intraprocedural circulatory depression and required transient intravenous pressors; 1 patient required 1 DC shock for ventricular fibrillation during wire navigation of the calcified aortic stenosis. Post-interventional aortography revealed similar grades of aortic insufficiency (Table 2).

The incidence of conduction delay requiring pacemaker implantation during 7-day follow-up after TAVI was 51.1% (n=23) due to complete AV block in 13.3% (n=6), type II second-degree AV block in 8.9% (n=4) and due to combined new LBBB with prolongation of HV interval ≥ 75 msec in 28.8% (n=13) (Table 3) (Figure 1). Baseline surface and intracardiac ECG recording revealed longer PQ interval of 197.1±51.2 msec versus 154.1±32.1 msec (p<0.001) caused by longer AH and HV intervals in the pacemaker group (Figure 2). Evaluation within 60 minutes after TAVI and at 7-day post TAVI, which was done in all patients, revealed a higher prevalence of new LBBB in pacemaker group with 87.0% versus 31.9% (p<0.001), as well as longer intervals for PQ, AH and HV interval (p<0.001). Similarly, QRS duration and antegrade and retrograde ERP were higher in pacemaker group (p<0.001) without any differences in SNRT or rates of right bundle branch block. Of 22 patients suffering from first-degree AV block 18 demonstrated prolonged HV interval (13 patients with a prolongation to ≥ 75 msec and 5 with a prolongation to < 75 msec), while 8 patients revealed a prolongation of AH interval without progression to complete heart block. Similarly, 5 patients with prolongation of HV interval to < 75 msec received no pacemaker (Table 3).

A multivariate analysis to identify predictors for high-grade AV block (type II second-degree and third-degree) resulting in pacemaker requirement included baseline clinical, electrocardiographic and echocardiographic parameters. With evolution to high-grade AV block as a hard endpoint in this model the univariate analysis revealed new LBBB immediately (within 60 minutes) after TAVI (OR 7.67; 95% CI 1.40-41.94; p=0.01), as well as PQ interval > 200 msec (OR 14.00; 95% CI 1.51-130.01; p=0.02) and QRS duration > 120 msec (OR 13.80; 95% CI 1.535-118.69; p=0.019) as predictors for high-grade AV block during hospital stay. Similarly, the multivariate model revealed new LBBB immediately (within 60 minutes) after TAVI (OR 24.85; 95% CI 1.57-392.57; p=0.023), as well as PQ interval > 200 msec (OR 11.37; 95% CI 1.138-97.620; p=0.02) and QRS > 120 msec (OR 14.28; 95% CI 1.50-135.88; p=0.021) as predictors for high-grade AV block. Other clinical parameters as well as baseline electrocardiographic parameters (e.g. AH interval > 100ms, HV interval > 75ms, QRS > 120ms and PQ > 200ms) had no impact to predict critical conduction delay (Tables 1, 2, 3).


Discussion

TAVI is established as an option with reduced morbidity and mortality in high risk symptomatic patients unfit for open surgery [10-14]. Considering the proximity of aortic valve annulus to conduction system, both open surgery and TAVI may impact on both AV node and His bundle conduction.

While complete AV block was reported in 5.7%, new LBBB occurred in 18% with an association to complete AV block, syncope, and sudden cardiac arrest at long term after open surgery [15-17]. Risk factors for complete AV block after surgical aortic valve replacement include previous aortic regurgitation, myocardial infarction, pulmonary hypertension, and postoperative electrolyte imbalance [16,17], among ECG criteria right bundle branch block (RBBB) on surface ECG was the strongest predictor for pacemaker requirement [16,17].

Although previous investigations report changes in surface-ECG after TAVI [2,10-14], our group was the first to describe intracardiac conduction abnormalities for better discrimination of new ECG changes on surface-ECG, and to predict critical conduction delays [18]. In our series, complete AV block was seen in 13.3%, while 8.9% suffered from type II second-degree AV block; thus, 22.2% of patients developed an indication for permanent pacemaker implantation corroborating previous findings [10-14,19-23]. Additionally, 28.9% patients received a pacemaker for new LBBB combined with a marked prolongation of the HV interval to ≥ 75 msec for primary prevention. Such indication may be debatable, but due to the lack of data and due to the novelty of TAVI in elderly patients with several comorbidities preventive pacemaker use appears justified [24].

Jilaihawi et al. reported first that pacemaker requirement after TAVI correlates to left axis deviation at baseline, LBBB, baseline thickness of the native non-coronary cusp and to diastolic interventricular septal dimension > 17 mm [13]. Similarly, Piazza et al. revealed no prosthesis-related LBBB when the proximal end of the valve frame was positioned < 6.7 mm from the lower edge of the non-coronary cusp [10]. Degenerative calcification of the aortic and mitral annulus is probably a diffuse process, in which the cardiac conduction system is often involved and making it vulnerable to injury when exposed to mechanical compression by the nitinol frame of the CoreValve, which seems to completely expand within the first 7–10 days [14]. Differences to surgical aortic valve replacement might be due to the different techniques. In surgical approach the valve is replaced by another. Thus, the amount of conduction damage is predictable because the local trauma is nearly the same in all patients. However, in TAVI the amount of local damage is dependent of local calcification, the height of implantation in LVOT, the extend of trauma during index-procedure (balloon valvuloplasty, balloon-to-aortic annulus relation, post-TAVI dilatation) and from further aortic annulus geometry.

Our intracardiac measurements revealed that occurrence of first-degree AV block were predominantly due to prolongation of HV interval, which might be prognostically relevant [24]. We implanted pacemaker in patients with new LBBB and a HV prolongation to ≥ 75 msec in a somewhat liberal fashion in this patients with high comorbidity index. In contrast, our multivariate analysis revealed that only a PQ duration > 200 msec, a LBBB and a QRS duration > 120 msec immediately (within 60 minutes) after CoreValve implantation seem to predict critical AV conduction delay. Other baseline clinical and ECG parameters had no impact. The occurrence of above ECG findings soon after TAVI may reflect the extent of trauma from the procedure. Interestingly, the exact determination of both the amount of valve calcification and the height of implantation turned out to be non-reproducible although both parameters have been claimed to impact on conduction physiology [10,13]. For example, the Edwards SAPIEN valve, shorter and less likely to extend into the left ventricular outflow tract, is obviously associated with a lower rate of complete AV block (0-6%) [25,26]. As demonstrated by our results, we believe that regardless of favourable anatomy only the extent of trauma predict the occurrence of critical conduction delay after TAVI. According to our multivariate analysis HV prolongation ≥ 75 msec is not a predictor for pacemaker requirement. We have implanted since 2007 nearly 400 aortic valves. Our pacemaker rate was initially about 45%. This was due to the novelty of this technique and lack of information regarding the true indication for pacing. However, with further analysis of patients and improving implantation techniques (e.g. high implantation technique, no further post-TAVI balloon dilatation) our pacing rate decreased to 9-10%. Based on our data we decided not to perform routinely intracardiac measurements. The presence of a new left bundle branch block in case of normal AV conduction on surface ECG in not an indication for pacing. For clinical routine, a surface 12-lead ECG immediately (within 60 minutes) after the TAVI procedure may indicate the need for pacemaker without incremental information from intracardiac ECG recordings, the quantification of aortic valve calcification, septal thickness or height of implantation.


Conclusion

TAVI is feasible and safe in patients unfit for open surgical aortic valve replacement. Due to the proximity of the implanted valve structure to the septum, conduction disturbances are common after TAVI requiring close surveillance at least for 7 days after TAVI. The occurrence of complete and type II second-degree AV block is predictable by a new LBBB, a PQ duration > 200 msec and QRS duration > 120 msec soon after TAVI and reflect the extent of trauma to the conduction system. Yet, the occurrence of new LBBB combined with first-degree AV block, reflecting prolongation of the HV interval, is frequent and should be subjected to further study.

Study limitation

The present study suffers from the limitation of being a retrospective analysis. Additionally, the low number of patients may influence the results. However, our study showed that there might be no further information from intracardiac measurements as the ECG changes predominantly present as LBBB, first grade and complete AV block. The differentiation of AV block in supra-His and infra-His is not essential in these frailty patients with high comorbidity index. Thus, there will always be a somewhat “liberal” indication for pacemaker implantation. To date, there is a need for further larger scale prospective trials as well as registries with longer follow-up for predicting factors for pacemaker requirement and defining the role and duration of ECG changes.


Abbreviations

AS: Aortic stenosis; AVA: Aortic valve area; AVR: Aortic valve replacement; c-SNRT: Corrected sinus node recovery time; ECG: Electrocardiography; ERP: Effective refractory period; ICU: Intensive care unit; IMC: Intermediate care; LBBB: Left bundle branch block; NYHA: New York Heart Association; PCI: Percutaneous coronary intervention; SNRT: Sinus node recovery time; TAVI: Transcatheter aortic valve implantation.


Competing interests

The authors declare that they have no competing interests.


Authors’ contribution

IA, SK, LP, HS, TCR, UT, RGT, DB, OT, DD, IBT, JO, HI, CAN participate in treating the patients during intervention / ICU and acquisition of data. IA, HI and CAN wrote the manuscript. IA and GK performed the statistical analyses. All authors read and approved the final manuscript.


Pre-publication history

The pre-publication history for this paper can be accessed here:

http://www.biomedcentral.com/1471-2261/12/87/prepub


References
Makkar RR,Fontana GP,Jilaihawi H,Kapadia S,Pichard AD,Douglas PS,Thourani VH,Babaliaros VC,Webb JG,Herrmann HC,Bavaria JE,Kodali S,Brown DL,Bowers B,Dewey TM,Svensson LG,Tuzcu M,Moses JW,Williams MR,Siegel RJ,Akin JJ,Anderson WN,Pocock S,Leon MB,PARTNER Trial InvestigatorsTranscatheter aortic-valve replacement for inoperable severe aortic stenosisN Engl J MedYear: 20123661696170410.1056/NEJMoa120227722443478
Kodali SK,Williams MR,Smith CR,Svensson LG,Webb JG,Makkar RR,Fontana GP,Dewey TM,Thourani VH,Pichard AD,Fischbein M,Szeto WY,Lim S,Greason KL,Teirstein PS,Malaisrie SC,Douglas PS,Hahn RT,Whisenant B,Zajarias A,Wang D,Akin JJ,Anderson WN,Leon MB,PARTNER Trial InvestigatorsN Engl J MedYear: 20123661686169510.1056/NEJMoa120038422443479
Zahn R,Gerckens U,Grube E,Linke A,Sievert H,Eggebrecht H,Hambrecht R,Sack S,Hauptmann KE,Richardt G,Figulla HR,Senges J,German Transcatheter Aortic Valve Interventions-Registry InvestigatorsTranscatheter aortic valve implantation: first results from a multi-centre real-world registryEur Heart JYear: 20113219820410.1093/eurheartj/ehq33920864486
Grube E,Schuler G,Buellesfeld L,Gerckens U,Linke A,Wenaweser P,Sauren B,Mohr FW,Walther T,Zickmann B,Iversen S,Felderhoff T,Cartier R,Bonan R,Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcomeJ Am Coll CardiolYear: 200750697610.1016/j.jacc.2007.04.04717601548
Webb JG,Chandavimol M,Thompson CR,Ricci DR,Carere RG,Munt BI,Buller CE,Pasupati S,Lichtenstein S,Percutaneous aortic valve implantation retrograde from the femoral arteryCirculationYear: 200611284285016461813
Bates MG,Matthews IG,Fazal IA,Turley AJ,Postoperative permanent pacemaker implantation in patients undergoing trans-catheter aortic valve implantation: what is the incidence and are there any predicting factors?Interact Cardiovasc Thorac SurgYear: 20011224325321098508
Roques F,Nashef SA,Michel P,EuroSCORE Project GroupRisk factors and outcome in European cardiac surgery: analysis of the EuroSCORE multinational database of 19030 patientsEur J Cardiothorac SurgYear: 19991581682210.1016/S1010-7940(99)00106-210431864
Vahanian A,Alfieri O,Al-Attar N,Antunes M,Bax J,Cormier B,Cribrier A,De Jaegere P,Fournial G,Kappetein AP,Kovac J,Ludgate S,Maisano F,Moat N,Mohr F,Nataf P,Piérard L,Pomar JL,Schofer J,Tornos P,Tuzcu M,van Hout B,Von Segesser LK,Walther T,European Association of Cardio-Thoracic Surgery; European Society of Cardiology; European Association of Percutaneous Cardiovascular InterventionsTranscatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Intervention (EAPCI)Eur Heart JYear: 2008291463147010.1093/eurheartj/ehn18318474941
Leon MB,Piazza N,Nikolsky E,Blackstone EH,Cutlip DE,Kappetein AP,Krucoff MW,Mack M,Mehran R,Miller C,Morel MA,Petersen J,Popma JJ,Takkenberg JJ,Vahanian A,van Es GA,Vranckx P,Webb JG,Windecker S,Serruys PW,Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the Valve Academic Research ConsortiumEur Heart JYear: 20113220521710.1093/eurheartj/ehq40621216739
Piazza N,Onuma Y,Jesserun E,Kint PP,Maugenest AM,Anderson RH,de Jaegere PP,Serruys PW,Early and persistent intraventricular conduction abnormalities and requirements for pacemaking after percutaneous replacement of the aortic valveJACC Cardiovasc IntervYear: 2008131031610.1016/j.jcin.2008.04.00719463319
Baan J Jr,Yong ZY,Koch KT,Henriques JP,Bouma BJ,Vis MM,Cocchieri R,Piek JJ,de Mol BA,Factors associated with cardiac conduction disorders and permanent pacemaker implantation after percutaneous aortic valve implantation with the CoreValve prosthesisAm Heart JYear: 201015949750310.1016/j.ahj.2009.12.00920211315
Calvi V,Puzzangara E,Pruiti GP,Conti S,Di Grazia A,Ussia GP,Capodanno D,Tamburino C,Early conduction disorders following percutaneous aortic valve replacementPacing Clin ElectrophysiolYear: 200932Suppl 1S12613019250075
Jilaihawi H,Chin D,Vasa-Nicotera M,Jeilan M,Spyt T,Ng GA,Bence J,Logtens E,Kovac J,Predictors for permanent pacemaker requirement after transcatheter aortic valve implantation with the CoreValve bioprosthesisAm Heart JYear: 200915786086610.1016/j.ahj.2009.02.01619376312
Piazza N,Nuis RJ,Tzikas A,Otten A,Onuma Y,Garcia-Garcia H,Schultz C,van Domburg R,van Es GA,van Geuns R,de Jaegere P,Serruys PW,Persistent conduction abnormalities and requirements for pacemaking six months after transcatheter aortic valve implantationEuroInterventionYear: 2010647548410.4244/EIJ30V6I4A8020884435
El-Khally Z,Thibault B,Stainloae C,Theroux P,Dubuc M,Roy D,Guerra P,Macle L,Talajic M,Prognostic significance of newly accuired bundle branch block after aortic valve replacementAm J CardiolYear: 2004941008101110.1016/j.amjcard.2004.06.05515476613
Marchenese K,Schenk EA,Atrioventricular conduction system lesion following cardiac valve replacementCirculationYear: 197245-46Suppl. IIII 188
Koplan BA,Stevenson WG,Epstein LM,Aranki SF,Maisel WH,Development and validation of a simple risk score to predict the need for permanent pacing after cardiac valve surgeryJ Am Coll CardiolYear: 20034179580110.1016/S0735-1097(02)02926-112628725
Akin I,Kische S,Schneider H,Liebold A,Ortak J,Bänsch D,Rehders TC,Thiele O,Schneider R,Kundt G,Krenz H,Chatterjee T,Nienaber CA,Ince H,Surface and intracardiac ECG for discriminating conduction disorders after CoreValve implantationClin Res CardiolYear: 2012535736422179507
Fraccaro C,Buja G,Tarantini G,Gasparetto V,Leoni L,Razzolini R,Corrado D,Bonato R,Basso C,Thiene G,Gerosa G,Isabella G,Iliceto S,Napodano M,Incidence, predictors, and outcome of conduction disorders after transcatheter self-expandable aortic valve implantationAm J CardiolYear: 201110774775410.1016/j.amjcard.2010.10.05421247519
Roten L,Wenaweser P,Delacretaz E,Hellige G,Stortecky S,Tanner H,Pilgrim T,Kadner A,Eberle B,Zwahlen M,Carrel T,Meier B,Windecker S,Incidence and predictors of atrioventricular conduction impairment after transcatheter aortic valve implantationAm J CardiolYear: 20101061473148010.1016/j.amjcard.2010.07.01221059439
Ferreira ND,Caeiro D,Adão L,Oliveira M,Gonçalves H,Ribeiro J,Teixeira M,Albuquerque A,Primo J,Braga P,Simões L,Ribeiro VG,Incidence and predictors of permanent pacemaker requirement after transcatheter aortic valve implantation with self-expanding bioprosthesisPacing Clin ElectrophysiolYear: 2010331364137210.1111/j.1540-8159.2010.02870.x20723083
Khawaja MZ,Rajani R,Cook A,Khavandi A,Moynagh A,Chowdhary S,Spence MS,Brown S,Khan SQ,Walker N,Trivedi U,Hutchinson N,De Belder AJ,Moat N,Blackman DJ,Levy RD,Manoharan G,Roberts D,Khogali SS,Crean P,Brecker SJ,Baumbach A,Mullen M,Laborde JC,Hildick-Smith D,Permanent pacemaker insertion after CoreValve transcatheter aortic valve implantation: incidence and contributiong factors (the UK CoreValve Collaborative)CirculationYear: 201112395196010.1161/CIRCULATIONAHA.109.92715221339482
Buellesfeld L,Stortecky S,Heg D,Hausen S,Mueller R,Wenaweser P,Pilgrim T,Gloekler S,Khattab AA,Huber C,Carrel T,Eberle B,Meier B,Boekstegers P,Jüni P,Gerckens U,Grube E,Windecker S,Impact of permanent pacemaker implantation on clinical outcome among patients undergoing transcatheter aortic valve implantationJ Am Coll CardiolYear: 20126049350110.1016/j.jacc.2012.03.05422726632
Epstein AE,DiMarco JP,Ellenbogen KA,Estes NA 3rd,Freedman RA,Gettes LS,Gillinov AM,Gregoratos G,Hammill SC,Hayes DL,Hlatky MA,Newby LK,Page RL,Schoenfeld MH,Silka MJ,Stevenson LW,Sweeney MO,American College of Cardiology/American Heart Association Task Force on Practice; American Association for Thoracic Surgery; Society of Thoracic SurgeonsACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) developed in collaboration with the Association for Thoracic Surgery and Society of Thoracic SurgeonsJ Am Coll CardiolYear: 200851e16210.1016/j.jacc.2008.02.03218498951
Erkapic D,Kim WK,Weber M,Möllmann H,Berkowitsch A,Zaltsberg S,Pajitnev DJ,Rixe J,Neumann T,Kuniss M,Sperzel J,Hamm CW,Pitschner HF,Electrocardiographic and further predictors for permanent pacemaker requirement after transcatheter aortic valve implantationEuropaceYear: 2010121188119010.1093/europace/euq09420353965
Walther T,Falk V,Kempfert J,Borger MA,Fassl J,Chu MW,Schuler G,Mohr FW,Transapical minimally invasive aortic valve implantation; the intitial 50 patientsEur J Cardiothorac SurgYear: 20083398398810.1016/j.ejcts.2008.01.04618294859

Figures

[Figure ID: F1]
Figure 1 

Rate of LBBB and AV block in pacemaker and non pacemaker group prior and 7 days after TAVI.



[Figure ID: F2]
Figure 2 

AH and HV conduction time in pacemaker and non pacemaker group prior and 7 days after TAVI.



Tables
[TableWrap ID: T1] Table 1 

Baseline characteristics of study population stratified in pacemaker (PM) and non pacemaker group (N-PM)


  PM (n=23) N-PM (n=22) p
Clinical parameters
Male, n (%)
10 (43.5)
8 (36.4)
0.763
Age (yrs)
80.7 ± 5.94
81.55 ± 7.90
0.684
Body mass index (kg/m2)
26.9 ± 3.9
26.8 ± 4.0
0.883
Hypertension, n (%)
22 (95.7)
19 (86.4)
0.346
Smoker, n (%)
8 (34.8)
4 (18.2)
0.024
Diabetes mellitus, n (%)
10 (43.5)
7 (31.8)
0.542
Creatinine (μmol/l)
151.2 ± 157.4
114.3 ± 51.6
0.372
Renal insufficiency (creatinine level > 1.5mg/dl), n (%)
14 (60.9)
11 (50)
0.554
Chronic obstructive pulmonary disease, n (%)
4 (17.1)
4 (18.2)
0.855
New York Heart Association functional class
I
0 (0)
0 (0)
0.263
II
0 (0)
2 (9.1)
 
III
15 (65.2)
15 (68.2)
 
IV
8 (34.8)
5 (22.7)
 
Logistic EuroSCORE (%)
24.0 ± 14.6
18.4 ± 19.2
0.034
Dyspnoe, n (%)
13 (56.6)
12 (54.6)
0.628
Angina, n (%)
11 (47.5)
6 (27.3)
0.022
Syncope, n (%)
3 (13.0)
8 (36.4)
0.091
Pulmonal artery pressure (mmHg)
42.8 ± 17.7
37.6 ± 19.9
0.375
Porcelain aorta, n (%)
1 (4.3)
3 (13.6)
0.346
Prior cardiac decompensation, n (%)
13 (56.5)
14 (63.7)
0.763
Ischemic heart disease, n (%)
20 (87.0)
15 (68.2)
0.017
Previous stroke, n (%)
10 (43.5)
3 (13.6)
0.047
Previous CABG, n (%)
1 (4.3)
2 (9.1)
0.838
Peripheral vessel disease, n (%)
5 (21.7)
1 (4.6)
0.002
Echocardiographic parameters
Mean aortic valve area (cm2)
0.8 ± 0.2
0.7 ± 0.2
0.108
Left ventricular ejection fraction (%)
46.1 ± 10.7
50.2 ± 12.5
0.224
Peak pressure gradient (mmHg)
90.6 ± 24.1
86.6 ± 27.9
0.641
Mean pressure gradient (mmHg)
57.7 ± 15.3
55.4 ± 15.6
0.674
Aortic annulus dimension (mm)
23.30 ± 5.32
22.36 ± 6.85
0.141
Aortic bulbus dimension (mm)
29.70 ± 5.86
28.61 ± 7.46
0.893
Interventricular septal dimension (mm)
14.17 ± 3.82
13.55 ± 2.46
0.627
Aortic regurgitation grade ≥ I, n (%)
19 (82.6)
20 (90.9)
0.692
Mitral insufficiency ≥ grade I, n (%) 22 (95.7) 19 (86.4) 0.247

[TableWrap ID: T2] Table 2 

Intraoperative data of pacemaker (PM) and non pacemaker group (N-PM)


  PM N-PM p
Procedural success, n (%)
23 (100)
22 (100)
0.999
Conversion to surgical AVR, n (%)
0 (0)
0 (0)
0.999
Intraprocedural circulatory depression, n (%)
4 (17.4)
2 (9.1)
0.635
Catecholamine therapy, n (%)
4 (17.4)
2 (9.1)
0.714
Resuscitation, n (%)
1 (4.4)
0 (0)
0.553
Defibrillation, n (%)
1 (4.4)
0 (0)
0.612
Vascular access site complication, n (%)
5 (21.8)
4 (18.2)
0.367
Contrast agent (ml)
113.3 ± 36.4
125.5 ± 60.8
0.732
Fluoroscopy time (min)
25.7 ± 1.8
23.9 ± 2.6
0.572
Procedure time (min)
107.7 ± 27.1
104.9 ± 33.3
0.725
CoreValve, n (%)
26mm
10 (43.5)
12 (54.6)
0.112
29mm
13 (56.6)
10 (45.5)
0.386
Pre TAVI, n (%)
23 (100)
22 (100)
0.999
Post TAVI valvuloplasty, n (%)
10 (43.5)
10 (45.5)
0.566
Number of inflations, n (%)
1.08 ± 0.75
1.14 ± 0.94
0.335
Balloon diameter (mm)
21.87 ± 3.84
21.27 ± 4.44
0.432
Balloon length (mm)
55.22 ± 26.81
52.73 ± 30.27
0.613
Angiographic Aortic insufficiency (grade)
1.22 ± 0.48
1.10 ± 0.62
0.725
ICU stay (days)
2.9 ± 3.2
2.5 ± 1.8
0.686
Hospital stay (days) 13.4 ± 11.3 12.0 ± 6.5 0.237

[TableWrap ID: T3] Table 3 

Electrocardiographic characteristics


  Prior TAVI After TAVI 7-day follow-up
Rhythm
Sinus
18 (78.3)
18 (78.3)
14 (63.7)
20 (91.0)
20 (91.0)
19 (82.7)*
Atrial fibrillation
5 (21.8)
5 (21.8)
5 (21.8)
2 (9.1)
2 (9.1)
5 (9.1)
Heart rate
72.7 ± 12.6
68.5 ± 13.8
66.0 ± 16.9
67.5 ± 10.5
63.2 ± 14.4
65.7 ± 13.1
Surface ECG
PQ interval
197.1 ± 51.2
230.42 ± 40.94
243.05 ± 34.59
154.1 ± 32.1*
167.37 ± 37.05*
168.56 ± 30.41*
QRS width
97.0 ± 15.0
139.04 ± 31.86
162.96 ± 33.59
95.3 ± 14.4
110.27 ± 18.38*
120.00 ± 24.17*
QT interval
388.2 ± 29.3
409.91 ± 28.67
408.52 ± 23.22
398.6 ± 32.8
407.36 ± 32.99
405.14 ± 32.03
Left-BBB
0 (0)
16 (69.6)
20 (87.0)
1 (4.5)
4 (18.2)*
7 (31.9)*
Right-BBB
1 (4.35)
0 (0)
0 (0)
1 (4.5)
1 (4.6)
1 (4.6)
First-degree block
5 (21.8)
8 (34.8)
18 (78.3)
3 (13.7) *
3 (13.7)*
4 (18.2)*
Second-degree block #
0 (0)
4 (17.4)
4 (17.4)
0 (0)
0 (0)*
0 (0)*
Complete block
0 (0)
4 (17.4)
6 (26.1)
0 (0)
0 (0)*
0 (0)*
Intracardiac measurement
SNRT
1086.8 ± 171.2
1093.89 ± 154.58
1122.11 ± 154.47
1045.6 ± 166.9
1050.56 ± 163.40
1070.59 ± 154.13
c-SNRT
487.0 ± 54.3
492.94 ± 43.88
498.13 ± 42.72
462.0 ± 68.4
491.12 ± 42.04
488.59 ± 42.48
AH interval
135.5 ± 44.7
139.32 ± 36.62
153.61 ± 43.40
105.8 ± 31.4*
114.50 ± 38.55*
116.06 ± 31.23*
HV interval
58.5 ± 12.5
74.00 ± 13.51
81.68 ± 17.76
48.7 ± 6.9*
55.27 ± 9.20*
56.76 ± 8.45*
Antegrade AV ERP
410.0 ± 42.6
470.00 ± 118.57
469.81 ± 45.75
394.4 ± 32.6
407.78 ± 32.64*
414.71 ± 31.45*
Retrograde AV ERP 442.6 ± 38.0
470.56 ± 45.10
494.00 ± 59.33
420.0 ± 34.3 436.67 ± 35.48* 441.76 ± 36.27*

* p-value <0.05.

italic : N-PM-group.

normal: PM-group.

# 2nd degree (type 2) AV-block.



Article Categories:
  • Research Article

Keywords: TAVI, AV block, Left bundle branch block, Pacemaker, His bundle.