Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices.
MedLine Citation:
PMID:  8060626     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
STUDY OBJECTIVES: To determine the frequency of true and false alarms and to determine the frequency of alarm failures for various parameters when using a postanesthesia monitoring system. DESIGN: Open prospective study. SETTING: Postanesthesia care unit at a university hospital. PATIENTS: 123 ASA physical status I-IV patients who underwent general or spinal-epidural anesthesia for general, urologic, orthopedic, ophthalmologic, otolaryngologic, or plastic surgery. MEASUREMENTS AND MAIN RESULTS: Monitoring included ECG, oxygen saturation, and respiratory rate (RR) by either transthoracic impedance or nasal-oral air-flow detection. We undertook careful, continuous observation of patients and monitors, recording of true and false alarms; the latter were defined as those coinciding with a clearly observed, unrelated cause. During a mean observation period of 101 minutes per patient, the average frequency of pulse oximetry alarms was once every 8 minutes, with 77% of the alarms being false, caused by sensor displacement, motion artifacts, poor perfusion, or a combination of these factors. Apnea alarms occurred on average once every 37 minutes, with the "false" fraction being 28% and 27% for impedance and flow detection, respectively. The impedance sensor failed to detect apnea on at least 6 occasions; the flow sensor failed on 1 occasion. The coincidence of pulse oximetry and apnea alarms was small, and ECG exhibited a low alarm rate but a high fraction of false alarms. Patients receiving opioids and neuromuscular relaxants had a higher frequency of "true" apneas than patients not receiving these drugs. No clear correlation was found in regard to age distribution, type of surgery, duration of anesthesia, oxygen administration, or ASA physical status between the patient groups that exhibited many apneas or many pulse oximeter alarms and the whole study population. CONCLUSIONS: The high frequency of apnea strongly motivates the use of continuous RR monitoring, preferably by flow-sensing techniques, since both central and obstructive apneas are then detected. Further study and development is necessary before pulse oximetry can be unconditionally recommended for postanesthesia monitoring.
Authors:
L Wiklund; B Hök; K Ståhl; A Jordeby-Jönsson
Related Documents :
23375786 - Effect of low-intensity pulsed ultrasound on bone healing at osteotomy sites after fore...
24718736 - Knee flexion after total knee arthroplasty reduces blood loss.
23543346 - Comparative effects of flurbiprofen and fentanyl on natural killer cell cytotoxicity, l...
22987136 - A comparative study between amiodarone and magnesium sulfate as antiarrhythmic agents f...
3917756 - Why patient participation groups stop functioning: general practitioners' viewpoint.
16044256 - The effectiveness of montelukast for the treatment of anti-histamine-resistant chronic ...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of clinical anesthesia     Volume:  6     ISSN:  0952-8180     ISO Abbreviation:  J Clin Anesth     Publication Date:    1994 May-Jun
Date Detail:
Created Date:  1994-09-22     Completed Date:  1994-09-22     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  8812166     Medline TA:  J Clin Anesth     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  182-8     Citation Subset:  IM    
Affiliation:
Department of Anesthesiology and Intensive Care, Uppsala University Hospital, Sweden.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Aged
Aged, 80 and over
Anesthesia Recovery Period
Apnea / diagnosis
Blood Pressure Monitors
Carbon Dioxide / analysis
Child
Child, Preschool
Electric Impedance
Electrocardiography / instrumentation
Equipment Failure
Humans
Infant
Middle Aged
Monitoring, Physiologic / instrumentation*
Oximetry / instrumentation
Prospective Studies
Pulmonary Ventilation / physiology
Recovery Room*
Respiration / physiology
Chemical
Reg. No./Substance:
124-38-9/Carbon Dioxide
Comments/Corrections
Comment In:
J Clin Anesth. 1994 May-Jun;6(3):178-9   [PMID:  8060624 ]
J Clin Anesth. 1994 May-Jun;6(3):180-1   [PMID:  8060625 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Ca2+ channel regulation by a conserved beta subunit domain.
Next Document:  Transarterial brachial plexus anesthesia for hand surgery: a retrospective analysis of 346 cases.