Pneumatic Performance of the Boussignac CPAP System in Healthy Humans.
MedLine Citation:
PMID:  21333088     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
BACKGROUND: The Boussignac CPAP has been found to effectively treat acute pulmonary oedema. However, data on airway pressure in association with the Boussignac CPAP are sparse. We therefore designed this study to evaluate the stability of the Boussignac CPAP in terms of maintaining adequate inspiratory and expiratory pressure levels. We also wanted to evaluate the perceived exertion when breathing with the Boussignac CPAP. METHODS: Continuous recordings of airway pressure and airflow were made in 18 healthy volunteers during breathing with the Boussignac CPAP at 5, 7.5 and 10 cm H₂O for three sessions of 10 minutes at each CPAP level. Participants were blinded for the sequential order of the investigated CPAP levels. They terminated each session, at the respective CPAP level, by taking 10 forced breaths. RESULTS: During 10 minutes normal breathing periods, when participants breathed at 20 % of their VC with a peak expiratory airflow of 14 % of FEV₁, the maximal pressure difference was 4.0 cm H₂O between inspiration and expiration at CPAP 10 cm H₂O. Changes in airway pressure were never large enough to reduce airway pressure below zero. When taking forced breaths, expiratory volume was 38-42 % of VC and peak expiratory airflow was 49-56 % of FEV ₁. As airflow increased, both the drop in inspiratory airway pressure and the increase in expiratory airway pressure were enhanced. CONCLUSIONS: Pressure changes during breathing with CPAP are considered to be associated with increased work of breathing. The pneumatic performance of the Boussignac CPAP is adequate during normal breathing with low airflow. However, during forced breathing resulting in increased airflow, the Boussignac CPAP is unable to maintain stable airway pressure levels, possibly resulting in increased work of breathing and respiratory fatigue. Thus, the Boussignac CPAP system might be less suitable for patients breathing at a higher frequency.
Authors:
Maria Sehlin; Siv Sandkvist Törnell; Fredrik Ohberg; Göran Johansson; Ola Winsö
Related Documents :
25161298 - Complications with utilization of positive-pressure devices in a young man with duchenn...
1556428 - Complications of nasoenteral tubes. intragastric tube knotting and intragastric tube br...
20377858 - Cuff overinflation and endotracheal tube obstruction: case report and experimental study.
10699808 - Conductive hearing loss associated with pressure equalization tubes.
14697228 - Alterations in blood pressure due to chronic exposure to natural sour gas leakage conta...
7241858 - Release of prostaglandins from the passively distended wall of guinea pig small intestine.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-2-11
Journal Detail:
Title:  Respiratory care     Volume:  -     ISSN:  0020-1324     ISO Abbreviation:  -     Publication Date:  2011 Feb 
Date Detail:
Created Date:  2011-2-21     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7510357     Medline TA:  Respir Care     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Remifentanil Improves Breathing Pattern and Reduces Inspiratory Workload in Tachypneic Patients.
Next Document:  Optimizing the Respiratory Pump: Harnessing Inspiratory Resistance to Treat Systemic Hypotension.