Document Detail


Permanent pacing system malfunction due to hidden adjacent fractures of atrial and ventricular leads.
MedLine Citation:
PMID:  9080029     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Periods of failure of ventricular pacing and atrial sensing were discovered on Holter recording in a 10-year-old boy 4 years after the implantation of a dual-chamber pacemaker with endocardial bipolar leads brought out through the right atrial wall. The old generator was explanted and the old leads tested. There was considerable atrial oversensing of an electrical artefact. The pacing threshold of ventricular lead was 1.8 V compared with 0.3 V 6 days post implantation. Post-operative lateral chest X-ray disclosed fractures of both atrial and ventricular leads at the approximate site of exit from the right atriotomy. Review of the preoperative lateral film showed that the fractures had been hidden by an ECG electrode.
Authors:
L Wang; M J Davis
Related Documents :
24192049 - Cardiac involvement in geleophysic dysplasia in three siblings of a saudi family.
1080449 - Abnormal regional metabolism and mechanical function in patients with ischemic heart di...
2432579 - Rate responsive dual chamber pacing.
1553889 - Systemic and coronary hemodynamic effects of bepridil in patients with depressed left v...
8430629 - Is atrial pacing needed for determination of coronary flow reserve by parametric imaging?
10750129 - Soluble l-selectin and neutrophil derived oxidative stress after pacing induced myocard...
6141779 - Granulomatous coronary arteritis.
2964589 - Left-ventricular cavity dimensions in children with normal and dilated hearts.
3719599 - Usefulness of predischarge echocardiographic criteria in predicting complications follo...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  International journal of cardiology     Volume:  59     ISSN:  0167-5273     ISO Abbreviation:  Int. J. Cardiol.     Publication Date:  1997 Mar 
Date Detail:
Created Date:  1997-06-12     Completed Date:  1997-06-12     Revised Date:  2006-07-12    
Medline Journal Info:
Nlm Unique ID:  8200291     Medline TA:  Int J Cardiol     Country:  IRELAND    
Other Details:
Languages:  eng     Pagination:  85-7     Citation Subset:  IM    
Affiliation:
Department of Cardiology, Royal Perth Hospital, Australia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Child
Electrocardiography
Electrodes
Equipment Failure
Humans
Male
Pacemaker, Artificial*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Versatility of Inoue balloon catheter.
Next Document:  Congenital left ventricular diverticulum.