| Performance evaluation of the ADVIA 2120 hematology analyzer: an international multicenter clinical trial. | |
| | |
MedLine Citation:
|
PMID: 15790554 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
Automated cell counters are widely used in modern clinical laboratories to provide reliable, fast, and cost-effective complete blood counts (CBCs), white blood cell differentials, and reticulocyte measurements. In addition, some advanced instruments provide novel parameters, such as the hemoglobin content of reticulocytes or the percentage of hypochromic cells, and are capable of analysis of a variety of body fluids. Bayer recently introduced the ADVIA 2120 system as an automation-ready cell counter for mid- to high-volume testing in the clinical laboratory. This instrument, which builds on the established technology of the ADVIA 120 system, operates with a cyanide-free method for hemoglobin measurement, has a new user interface, and can routinely analyze biological fluid samples in addition to blood. We used 749 samples from 6 worldwide trial sites to evaluate the clinical performance of this new device. Accuracy of the ADVIA 2120 system versus its predecessor model, the ADVIA 120 system, was excellent for all CBC and white cell differential parameters and reticulocyte counts (all correlation coefficients except for basophils >0.9). Correlation of the white cell differential with the standard manual method and within-run precision of the ADVIA 2120 system also was very good. Use of the novel cyanide-free method for hemoglobin measurement had no clinically significant impact on hemoglobin results, even in patients with hemoglobinopathies. We concluded that the ADVIA 2120 system has clinically equivalent performance to the ADVIA 120 system. |
| | |
Authors:
|
Neil Harris; Josep Maria Jou; Gianluigi Devoto; J Lotz; James Pappas; David Wranovics; Mary Wilkinson; Suzanne R Fletcher; Alexander Kratz |
Related Documents
:
|
17784644 - Electron microscopic specimen preparation from low concentration of cell suspension usi... 1694214 - Use of colloidal silica (sepracell-mn) for enrichment of dendritic cells from human per... 6602774 - Studies on thymocyte subpopulations in guinea pigs. 2. three subpopulations differing i... 12054134 - Evaluation of measurements of the conductivity of quarter milk samples for the early di... 18246544 - Blocking of intracellular ros production by phytoglycoprotein (30 kda) causes anti-prol... 16206244 - Interaction of pp2a catalytic subunit with rb2/p130 is required for all-trans retinoic ... |
Publication Detail:
|
Type: Clinical Trial; Comparative Study; Journal Article; Multicenter Study; Research Support, Non-U.S. Gov't |
Journal Detail:
|
Title: Laboratory hematology : official publication of the International Society for Laboratory Hematology Volume: 11 ISSN: 1080-2924 ISO Abbreviation: Lab Hematol Publication Date: 2005 |
Date Detail:
|
Created Date: 2005-03-25 Completed Date: 2005-07-05 Revised Date: 2006-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 9615688 Medline TA: Lab Hematol Country: United States |
Other Details:
|
Languages: eng Pagination: 62-70 Citation Subset: IM |
Affiliation:
|
University of Florida/Shands Hospital Core Laboratory, Gainesville, Florida, USA. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Automation Blood Cell Count / methods* Hematology / methods* Humans Leukocyte Count Lymphocyte Count Reticulocyte Count Sensitivity and Specificity Software |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: The ADVIA 2120 hematology system: flow cytometry-based analysis of blood and body fluids in the rout...
Next Document: Effects of the blood-collection tube material and long-term storage on platelet activation parameter...