Performance and error analysis of automated external defibrillator use in the out-of-hospital setting.
MedLine Citation:
PMID:  11524645     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
STUDY OBJECTIVE: We determined whether automated external defibrillators (AEDs) can meet the American Heart Association performance criteria to detect and shock unstable cardiac rhythms (ventricular fibrillation [VF], ventricular tachycardia [VT]) in the setting of an out-of-hospital cardiac arrest. METHODS: AED performance was reviewed for cardiac arrests occurring between January 1, 1995, and December 31, 1997. After every cardiac arrest, data regarding each rhythm analyzed and subsequent response (shock or no shock) were downloaded from the AED memory module. The study paramedic and study physician independently reviewed each case and interpreted cardiac rhythms from downloaded AED data. The emergency medical services medical director resolved all discrepancies in a blinded manner. All cases of out-of-hospital cardiac arrest in which an AED was turned on and a rhythm analyzed were included. The primary objective was the correct identification and defibrillation of VF or VT. Sensitivity, specificity, and predictive values with 95% confidence intervals (CIs) were calculated. Sources of error in AED rhythm management are also described. RESULTS: A total of 3,448 AED rhythms were available for interpretation. Sensitivity and specificity for appropriate AED management of a shockable (VF or VT) rhythm were 81.0% (95% CI 77.9% to 83.8%) and 99.9% (95% CI 99.7% to 100%), respectively. Positive and negative predictive values were 99.6% (95% CI 98.7% to 99.9%) and 95.5% (95% CI 94.7% to 96.2%), respectively. There were 132 errors associated with AED management. Two errors resulted in delivery of an inappropriate shock. In the remaining 130 errors, a shockable rhythm was not shocked. Fifty-five (42.3%) errors were AED dependent, 70 (53.9%) were operator dependent, and 5 (3.9%) were unclassified. CONCLUSION: The AED had high specificity and moderately high sensitivity in detecting and shocking unstable cardiac rhythms in the out-of-hospital setting. Few cardiac rhythms were mismanaged by the AED. Elimination of operator-dependent errors could increase AED sensitivity.
Authors:
R D Macdonald; J M Swanson; J L Mottley; C Weinstein
Related Documents :
3359585 - Energy, current, and success in defibrillation and cardioversion: clinical studies usin...
8161395 - Imipramine-provoked paradoxical pheochromocytoma crisis: a case of cardiogenic shock.
9406175 - Cardiac depression and cellular injury in hemorrhagic shock and reinfusion: role of fre...
20328285 - Shock following myocardial infarction: a clinical survey of 140 cases.
6319935 - Reduced tolerance to digitalis-induced arrhythmias caused by coronary-flow alterations ...
16159805 - Extracellular matrix scaffold for cardiac repair.
Publication Detail:
Type:  Evaluation Studies; Journal Article    
Journal Detail:
Title:  Annals of emergency medicine     Volume:  38     ISSN:  0196-0644     ISO Abbreviation:  Ann Emerg Med     Publication Date:  2001 Sep 
Date Detail:
Created Date:  2001-08-28     Completed Date:  2001-09-27     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  8002646     Medline TA:  Ann Emerg Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  262-7     Citation Subset:  AIM; IM    
Affiliation:
Department of Emergency Medicine, Boston Medical Center and Boston University School of Medicine, Boston, MA, USA. rmacdonald@gov.mb.ca
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Boston
Cardiopulmonary Resuscitation
Electric Countershock / instrumentation*
Emergency Medical Services*
Equipment Failure Analysis*
Heart Arrest / diagnosis,  mortality,  therapy
Humans
Sensitivity and Specificity
Survival Analysis
Tachycardia, Ventricular / diagnosis,  mortality,  therapy*
Ventricular Fibrillation / diagnosis,  mortality,  therapy*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Out-of-hospital defibrillation with automated external defibrillators: postshock analysis should be ...
Next Document:  Can paramedics using guidelines accurately triage patients?