Document Detail


Performance characteristics of a novel radioactive isotope detection and notification system designed for use in hospitals.
MedLine Citation:
PMID:  21451311     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
Recently, University of Pittsburgh Medical Center (UPMC) Cancer Centers has installed an Emergency Department Notification System (EDNS) in one of its hospitals. This system, manufactured by Thermo Fisher Scientific (Thermo Fisher Scientific, Inc., 81 Wyman Street, Waltham, MA 02454), was designed to discriminate non-medical radioactive isotopes from medical radioactive isotopes routinely used in nuclear medicine and radiation treatments. It is modular in nature and consists of four NaI(Tl) scintillation detectors, a 512 channels multi-channel analyzer, a system controller, and a database-monitoring server. A series of tests were carried out to evaluate the performance characteristics of this system using a variety of radioactive sources of varying activities. These included measurements of minimum detectable activity, detector response distance to various source activities, detector response to different speeds of a moving radioisotope, and single and multiple radioisotope identification and classification. Measured results show that the system is capable of identifying radioactive sources of nominal activity 0.13 MBq (3.5 μCi) and higher in a relatively short period of time (<11.1 s). The database-monitoring server could send an alarm signal to appropriate personnel when the analysis of the results indicated the presence of a non-medical or threat radioisotope. The present paper reports these results.
Authors:
Jong Oh Kim; M Saiful Huq; Josef Novotny; Greg Bednarz; Richard Palatine; Michael Reilly; Mark Izadbakhsh; Paul Paris; Joel S Greenberger
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Health physics     Volume:  100 Suppl 2     ISSN:  1538-5159     ISO Abbreviation:  Health Phys     Publication Date:  2011 May 
Date Detail:
Created Date:  2011-03-31     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  2985093R     Medline TA:  Health Phys     Country:  United States    
Other Details:
Languages:  eng     Pagination:  S71-8     Citation Subset:  IM    
Affiliation:
* University of Pittsburgh Cancer Institute, Pittsburgh, PA; † Thermo Fisher Scientific Inc., Waltham, MA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Improved Confidence Intervals when the Sample is Counted an Integer Times Longer Than the Blank.
Next Document:  Quantification of various factors influencing the precision of thermoluminescent detector calibratio...