Performance of an automated external defibrillator during simulated rotor-wing critical care transports.
MedLine Citation:
PMID:  21236548     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE: This study aimed to evaluate whether an automated external defibrillator (AED) was accurate enough to analyze the heart rhythm during a simulated rotor wing critical care transport. We hypothesized that AED analysis of the simulated rhythms during a helicopter flight would result in significant errors (i.e., inappropriate shocks, analysis delay). METHODS: Three commercial AEDs were tested for analyzing the heart rhythm in a helicopter using a manikin and a human volunteer. Ventricular fibrillation (VF), sinus rhythm, and asystole were simulated by using an arrhythmia simulator of the manikin. The intervals from analysis to shock recommendation were collected on a stationary and in-motion helicopter. Sensitivity and specificity of three AEDs were also calculated. Vibration intensities were measured with a digital vibration meter placed on the chest of the manikin/human volunteer both on the stretcher and on the floor of the helicopter. RESULTS: All AEDs correctly recommended shock delivery for the cardiac rhythms of the manikin. Sensitivity for VF was 100.0% (95% CI 91.2-100.0) and specificity for sinus rhythm and asystole were 100.0% (95% CI 91.2-100.0). Although the recorded ECG rhythms of the volunteer in an in-motion helicopter showed baseline artifacts, all AEDs analyzed the cardiac rhythm of the volunteer correctly and did not recommend shock delivery. On the floor of the helicopter, the median measured vibration intensity was 6.6m/s(2) (IQR 5.5-7.7m/s(2)) with significantly less vibrations transmitted to the manikin/human volunteer chest (manikin median 3.1m/s(2), IQR 2.2-4.0m/s(2); human volunteer median 0.95m/s(2), IQR 0.65-1.25m/s(2)). CONCLUSION: This study suggested that current AEDs could analyze the heart rhythm correctly during simulated helicopter transport. Further studies using an animal model would be needed before applying to patients.
Authors:
Sang Mo Je; Je Sung You; Tae Nyoung Chung; Yoo Seok Park; Sung Phil Chung; In Cheol Park
Related Documents :
3915878 - Expectations and limitations of endomyocardial biopsy.
18369528 - Clinical strategies for amyloid a amyloidosis secondary to rheumatoid arthritis.
8362718 - New approach to the estimation of the extent of myocardial fibrosis in patients with di...
8205668 - Clinical and prognostic significance of detection of enteroviral rna in the myocardium ...
3045498 - Anatomy and quantification of myocardial cell death.
9833138 - Force-frequency relation in patients with left ventricular hypertrophy and failure.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-1-12
Journal Detail:
Title:  Resuscitation     Volume:  -     ISSN:  1873-1570     ISO Abbreviation:  -     Publication Date:  2011 Jan 
Date Detail:
Created Date:  2011-1-17     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0332173     Medline TA:  Resuscitation     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2011. Published by Elsevier Ireland Ltd.
Affiliation:
Department of Emergency Medicine, Yonsei University College of Medicine, 250 Seongsanno (134 Sinchon-dong), Seodaemun-gu, 120-752 Seoul, Republic of Korea.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Therapeutic hypothermia attenuates acute lung injury in paraquat intoxication in rats.
Next Document:  Measuring human mast cell-mediated cytotoxicity against human tumour cells: three-colour flow cytome...