Document Detail


Pediatric compartment syndrome caused by intravenous infiltration.
MedLine Citation:
PMID:  21301289     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The small diameter and fragility of infant veins increases the risk of intravenous (IV) fluid extravasation. Modern pump delivery systems, designed with pressure sensors to detect and respond to flow obstructions in the IV line, are in wide use to limit the amount of fluid that reaches the extravascular space. In some instances, the pump safety systems can fail to detect rising pressure and may lead to compartment syndrome.
METHODS: Retrospective chart review of patients treated for extremity compartment syndrome as a result of an IV infiltration.
RESULTS: Three infants (mean age, 9 months) were identified. Each patient had a large volume of IV fluid delivered into the subcutaneous tissue through a pump infusion system. In all cases, the sensor failed to alarm and to disable the system, and the problem was identified only after routine nursing evaluation of the IV site. Findings included the following: tissue blanching, decreased capillary refill, and severely restricted active/passive motion of the affected extremity. Peripheral pulses were present in all patients. Compartment pressure elevation (>30 mm Hg) was confirmed prior to operative intervention by a slit catheter system. Fasciotomy was required in the hand and forearm of 2 patients, and of the thigh, leg, and foot in 1 patient. Each infant recovered full use of the extremity.
CONCLUSIONS: The reason that these pumps failed to detect the dangerously elevated pressures is unclear, but may be related to a combination of fluid dynamics and pump engineering. Routine nursing evaluation remains the gold standard for detection of these events.
Authors:
Simon G Talbot; Gary F Rogers
Related Documents :
22103719 - Bh3 under pressure - leaving the molecular diborane motif.
21974179 - Statistical analysis of a characteristic shock formation distance for high-amplitude no...
10963629 - Theoretical simulation of oxygen transport to brain by networks of microvessels: effect...
8093829 - The use of dopexamine hydrochloride to increase oxygen delivery perioperatively.
22215589 - Spiral chain o₄ form of dense oxygen.
6981449 - Continuous monitoring of mixed venous oxygen saturation in cardiac surgery.
8567049 - Endothelin-1 and its receptor in hypertrophic cardiomyopathy.
7677389 - Adrenergic and imidazoline receptor-mediated responses to uk-14,304-18 (brimonidine) in...
6341219 - The role of renin substrate in hypertension.
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Annals of plastic surgery     Volume:  67     ISSN:  1536-3708     ISO Abbreviation:  Ann Plast Surg     Publication Date:  2011 Nov 
Date Detail:
Created Date:  2011-10-17     Completed Date:  2012-03-05     Revised Date:  2012-10-16    
Medline Journal Info:
Nlm Unique ID:  7805336     Medline TA:  Ann Plast Surg     Country:  United States    
Other Details:
Languages:  eng     Pagination:  531-3     Citation Subset:  IM    
Affiliation:
Department of Plastic Surgery, Children's Hospital Boston, 300 Longwood Avenue, Boston, MA 02215, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Compartment Syndromes / etiology*
Extravasation of Diagnostic and Therapeutic Materials / complications*
Humans
Infant
Male
Retrospective Studies
Veins
Comments/Corrections
Comment In:
Ann Plast Surg. 2012 Sep;69(3):340   [PMID:  22722644 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Three-Year Experience Using Near Infrared Spectroscopy Tissue Oximetry Monitoring of Free-Tissue Tra...
Next Document:  Suicide by Burning: Epidemiological and Clinical Profile.