Document Detail


Patients with an ICD can safely resume work in industrial facilities following simple screening for electromagnetic interference.
MedLine Citation:
PMID:  12877699     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Patients with ICDs are commonly advised to quit industrial jobs because of concerns that strong electromagnetic fields operating in the industrial environment might interfere with ICD functions. This study was done to assess interactions between industrial equipment and ICDs, and to devise a simple low risk screening protocol. We studied 18 patients carrying nine different ICD models who were met at their workplace by a clinical technician and were asked to walk through their workplace and perform typical duties while sensing status was monitored by listening to the ICD's beeper. All devices were interrogated at the completion of testing. At follow-up, patients were contacted by phone and were asked about employment status and history of ICD discharges or syncope. One hundred eighty-four contacts with 114 types of industrial equipment in 13 different industrial facilities (including 31 contacts with arc welding machines) were monitored. Interference with ICD's function occurred in only one contact (0.5%), when ICD therapy was temporarily suspended while a worker was attaching a huge electromagnet to a crane. At follow-up 46.0 +/- 6.0 months after testing, 7 patients (41%) are still holding the same job, 7 have retired because of reasons unrelated to their ICD, and 3 patients were transferred to a nonindustrial job. None of the patients had either an ICD shock or syncope during work. The use of a simple screening procedure can safely identify sources of electromagnetic interference that may affect ICD operation, and can predict long-term safety of working in an industrial workplace for ICD patients.
Authors:
Osnat Gurevitz; Richard I Fogel; Mark E Herner; Ross Sample; Adam S Strickberger; Emile G Daoud; Fred Morady; Eric N Prystowsky
Related Documents :
1699399 - Reproducibility in circadian rhythm of ventricular premature complexes.
8595569 - High resolution electrocardiographic abnormalities as possible indexes of electric inst...
3933319 - Spectrum of antiarrhythmic response to encainide.
10725289 - Linear ablation lesions for control of unmappable ventricular tachycardia in patients w...
12798829 - Eligibility for biventricular pacing in patients with an implantable cardioverter defib...
16959689 - A new clinical model in pulmonary embolism and its correlation with v/p scan results.
6793659 - Chronic intestinal pseudo-obstruction.
8551319 - The contribution of single photon emission tomography to the clinical differentiation o...
1091659 - Excess intravascular coagulation complicating low cardiac output.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  26     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  2003 Aug 
Date Detail:
Created Date:  2003-07-24     Completed Date:  2003-12-16     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1675-8     Citation Subset:  IM    
Affiliation:
Indiana Heart Institute, Indianapolis, Indiana, USA. ossigur@hotmail.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Defibrillators, Implantable*
Electromagnetic Fields*
Equipment Safety
Female
Humans
Male
Middle Aged
Occupational Exposure*
Tachycardia, Ventricular / therapy
Workplace