Document Detail


Patient-prosthesis mismatch may be irrelevant after aortic valve replacement with the 19-mm Perimount pericardial bioprosthesis in patients aged 65 years or older.
MedLine Citation:
PMID:  18071849     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The prevalence of patient-prosthesis mismatch (PPM) and its influence on clinical midterm results were examined in elderly patients whose activity was supposed to be less than that of younger patients. We evaluated valve function and the effects of PPM on the midterm results of the 19-mm Carpentier-Edwards Perimount (CEP) pericardial aortic valve in patients aged 65 years or older. Between August 1996 and May 2005, 51 patients underwent aortic valve replacement with the 19-mm CEP valve. The mean follow-up was 2.4 +/- 1.8 years, involving a total of 134.4 patient-years. The mean age and body surface area at operation were 74.0 +/- 5.0 years and 1.41 +/- 0.14 m(2). There were two (3.9%) operative deaths. Three patients (5.9%) underwent enlargement of their small aortic annuli. The actuarial survival rate at 8 years, including operative mortality, averaged 90.2% +/- 4.7%. The freedom from thromboembolism, reoperation, and valve-related mortality averaged 75.0% +/- 21.7%, 97.8% +/- 2.2%, and 95.3% +/- 3.2%, respectively, at 8 years. High preoperative peak and mean transvalvular pressure gradients were significantly improved after the operation (peak, 93 +/- 35 versus 28 +/- 12 mmHg; mean, 58 +/- 19 versus 17 +/- 7 mmHg, respectively; P < 0.01). The mean left ventricular mass index was reduced from 192 +/- 44 to 142 +/- 46 g/m(2) at late follow-up (P < 0.01). The prevalence of PPM was low (17.6%) when an indexed effective orifice area of less than 0.85 cm(2)/m(2) was taken as the definition of PPM. The clinical results, postoperative pressure gradients, and reduction in left ventricular mass index were not different between the PPM and no-PPM groups. The 19-mm CEP valve produced satisfactory midterm clinical outcomes in patients aged 65 years or older whose activity was supposed to be less than that of younger patients, regardless of the presence or absence of PPM. Moderate PPM was rare and it did not adversely impact on the midterm results. The application of annulus enlargement could be limited to the small number of patients for whom the 19-mm CEP valves are not able to be inserted.
Authors:
Yoshimasa Sakamoto; Kazuhiro Hashimoto; Hiroshi Okuyama; Tatsuumi Sasaki; Hiromitsu Takakura; Katsuhisa Onoguchi
Related Documents :
20837909 - Impact of early surgery on embolic events in patients with infective endocarditis.
6463629 - Mechanical heart valves in children.
10986059 - Osteonecrosis as a complication of treating acute lymphoblastic leukemia in children: a...
17410499 - Initial experience with the medtronic advantage valve prosthesis in the mitral position...
9930419 - Prosthetic replacement of the tricuspid valve: biological or mechanical?
2709859 - The outcome of surgical treatment of hypertrophic obstructive cardiomyopathy. experienc...
2703719 - Assessment and prediction in leg-length discrepancy.
21035869 - A randomized controlled trial of alleviated positioning after small macular hole surgery.
22382139 - Cryoballoon in atrial fibrillation ablation.
Publication Detail:
Type:  Journal Article     Date:  2007-12-20
Journal Detail:
Title:  Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs     Volume:  10     ISSN:  1434-7229     ISO Abbreviation:  J Artif Organs     Publication Date:  2007  
Date Detail:
Created Date:  2007-12-11     Completed Date:  2008-05-13     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9815648     Medline TA:  J Artif Organs     Country:  Japan    
Other Details:
Languages:  eng     Pagination:  206-11     Citation Subset:  IM    
Affiliation:
Department of Cardiovascular Surgery, Jikei University School of Medicine, 3-25-8 Nishishinbashi, Minato-ku, Tokyo 105-8461, Japan. yysakamoto@aol.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Aortic Valve*
Aortic Valve Insufficiency / surgery*
Aortic Valve Stenosis / surgery*
Bioprosthesis*
Chi-Square Distribution
Echocardiography
Female
Follow-Up Studies
Heart Valve Prosthesis*
Humans
Male
Postoperative Complications
Prosthesis Design