| Partial inhibition of ongoing antitachycardia pacing sequence due to T-wave oversensing. | |
| | |
MedLine Citation:
|
PMID: 18507556 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
We describe a case of partial inhibition of ongoing antitachycardia pacing scheme in an implantable cardioverter defibrillator (ICD) recipient with heart failure of ischemic etiology. The cause of inhibition was related to intermittent oversensing of paced T wave and sensing of intrinsic ventricular activation. In our effort to solve this problem, we reprogrammed the paced ventricular blanking period to a higher level. |
| | |
Authors:
|
Athanasios G Manolis; Dimitrios G Chatzis; Kostas Kouvelas; Zenon S Kyriakides |
Related Documents
:
|
7943706 - The oesophageal and precordial stethoscope transducer as a monitoring and teaching aid. 447196 - Problems in the nomenclature of the univentricular heart. 15606986 - A first step towards translating evidence into practice: heart failure in a community p... 2614866 - Reduced cardiac selenium content in the acquired immunodeficiency syndrome. 11925796 - Quality handbook in radiotherapy. brachytherapy: personal experience. 20873926 - Goals and everyday problem solving: manipulating goal preferences in young and older ad... |
Publication Detail:
|
Type: Case Reports; Journal Article |
Journal Detail:
|
Title: Pacing and clinical electrophysiology : PACE Volume: 31 ISSN: 1540-8159 ISO Abbreviation: Pacing Clin Electrophysiol Publication Date: 2008 Jun |
Date Detail:
|
Created Date: 2008-05-29 Completed Date: 2008-06-27 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 7803944 Medline TA: Pacing Clin Electrophysiol Country: United States |
Other Details:
|
Languages: eng Pagination: 780-1 Citation Subset: IM |
Affiliation:
|
Second Cardiology Department, Hellenic Red Cross General Hospital, Athens, Greece. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Aged Defibrillators, Implantable* Electrocardiography / methods* Equipment Failure* Heart Failure / prevention & control* Humans Male Pacemaker, Artificial* Tachycardia, Ventricular / prevention & control* |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Precipitation of ventricular fibrillation by intravenous diltiazem and metoprolol in a young patient...
Next Document: Short PR interval in the absence of preexcitation: a characteristic finding in a patient with Fabry ...