| Pacemaker failure induced by radiotherapy. | |
| | |
MedLine Citation:
|
PMID: 7513850 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
It is well documented that radiation therapy can have significant transitory and permanent effects on the function of cardiac pacemakers. In this study 18 multiprogrammable pacemakers were tested to establish any pattern by which these pacemakers are affected. The results confirm previous predictions on how pacemakers are damaged. The damage could be divided essentially into three types: (1) temporary change to interference or safety mode pacing lasting for the duration of the irradiation only; (2) change to interference mode pacing--from which recovery may occur after reprogramming the pacemaker; and (3) severe damage; in this case the pacemaker stops generating pulses. Recovery from this may occur a long time after the end of the treatment, but this recovery is mostly incomplete and the pacemaker cannot be used reliably thereafter. Therefore it is essential that patients with implanted pacemakers undergoing radiation therapy, should be monitored closely during the course of the treatment and for a few weeks thereafter. |
| | |
Authors:
|
S K Souliman; J Christie |
Publication Detail:
|
Type: Journal Article |
Journal Detail:
|
Title: Pacing and clinical electrophysiology : PACE Volume: 17 ISSN: 0147-8389 ISO Abbreviation: Pacing Clin Electrophysiol Publication Date: 1994 Mar |
Date Detail:
|
Created Date: 1994-06-06 Completed Date: 1994-06-06 Revised Date: 2004-11-17 |
Medline Journal Info:
|
Nlm Unique ID: 7803944 Medline TA: Pacing Clin Electrophysiol Country: UNITED STATES |
Other Details:
|
Languages: eng Pagination: 270-3 Citation Subset: IM |
Affiliation:
|
Department of Clinical Physics and Bio-Engineering, Beatson Oncology Centre, Belvidere Hospital, Glasgow, Scotland, United Kingdom. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Electromagnetic Fields Equipment Design Equipment Failure Humans Materials Testing Pacemaker, Artificial* Radiotherapy* Radiotherapy Dosage |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Codon reading scheme in Mycoplasma pneumoniae revealed by the analysis of the complete set of tRNA g...