Oxygen enrichment of bag-valve-mask units during positive-pressure ventilation: a comparison of various techniques.
MedLine Citation:
PMID:  3278659     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Bag-mask devices are used frequently to provide patients with positive-pressure-assisted ventilation. To increase the percentage of oxygen delivered (FDO2) from the bag, supplemental oxygen must be provided by way of an oxygen inlet nipple attached to the unit. Using ten medical volunteers and a test lung with oxygen analyzer, we studied the effect of several variables on the FDO2 and determined the most effective reservoir that would provide the highest consistent FDO2 from the ventilating port of the bag. An FDO2 of 1.00 was consistently provided by the 2.5-L bag reservoir and a demand-valve set-up attached to the reservoir port of the ventilating bag. Bag refill time significantly affected the FDO2 when no reservoir or corrugated tube reservoirs were used. Corrugated tube reservoirs were found to be more sensitive to all variations in ventilatory technique and to oxygen flow rates. From our findings we recommend that corrugated tube reservoirs not be used for oxygen supplementation, as they are sensitive to variations in ventilatory technique and cannot alert clinicians to problems with oxygen flow. While both the 2.5-L bag reservoir and demand-valve provide a consistent FDO2 of 1.00, the demand valve has the advantage of audible filling of the ventilating bag as well as being compact and independent of ventilatory technique.
Authors:
T P Campbell; R D Stewart; R M Kaplan; R V DeMichiei; R Morton
Related Documents :
3665009 - The effect of cardioplegic oxygenation on the correlation between the linearized frank-...
171249 - Carbon monoxide-induced localized toxic anoxia in the rat brain cortex.
16889379 - Numerical simulation of oxygen delivery to muscle tissue in the presence of hemoglobin-...
1095039 - A comparison of methods for the measurement of cardiac output and blood oxygen content.
1303459 - Pao2, sao2 and what constitutes adequate oxygenation in the premature newborn.
9012109 - Comparisons of sevoflurane, isoflurane, and halothane anesthesia in spontaneously breat...
12200639 - Endogenous nitric oxide regulates right coronary blood flow during acute pulmonary hype...
3558229 - Density dependence of maximal flow is lung volume dependent during bronchoconstriction.
6685599 - In vitro analysis of thermal transport in coronary sinus thermodilution catheters.
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Annals of emergency medicine     Volume:  17     ISSN:  0196-0644     ISO Abbreviation:  Ann Emerg Med     Publication Date:  1988 Mar 
Date Detail:
Created Date:  1988-03-28     Completed Date:  1988-03-28     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  8002646     Medline TA:  Ann Emerg Med     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  232-5     Citation Subset:  AIM; IM    
Affiliation:
Division of Emergency Medicine, University of Pittsburgh School of Medicine, Pennsylvania.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Equipment Design
Humans
Lung / metabolism,  physiology
Lung Compliance
Oxygen Consumption
Oxygen Inhalation Therapy / methods*
Positive-Pressure Respiration / instrumentation*,  methods