Overall accuracy of the BpTRU--an automated electronic blood pressure device.
MedLine Citation:
PMID:  15021078     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The objective of this report is to combine the data from an earlier adult study with the data from a paediatric study in order to determine the overall accuracy of the BpTRU (BPM-100 model) as compared to the recognized standard, auscultatory mercury sphygmomanometer. DESIGN: The individual blood pressure points recorded for both adult and paediatric studies were compared directly to its corresponding observer reference measurements from data collected and stored from the two separate studies. There were 255 sets of readings in the adult study and 162 sets from the paediatric study, which were combined to make 417 pairs of blood pressure readings for this study. METHODS: The overall observer standard reference mean for the 417 measurements was calculated and the difference between this and the overall mean BPM-100 was calculated with SD and ranges. Measurements within 5, 10 and 15 mmHg agreement were expressed as percentages. RESULTS: A total of 121 subjects were included for this study (85 from the adult study and 36 from the paediatric study). From these, 417 paired measurements were recorded. The mean difference between the BpTRU and the reference standard systolic blood pressure (BP) was 0.47+/-5.40 mmHg with 89.2% measurements within 5 mmHg, 96.4% within 10 mmHg and 99.3% within 15 mmHg. The mean difference between the BpTRU and reference diastolic BP was -2.12+/-5.93 mmHg with 81.1% within 5 mmHg, 92.1% within 10 mmHg and 97.6% within 15 mmHg. CONCLUSION: The BpTRU has been shown to be an accurate non-invasive blood pressure monitoring device in the general population over a wide range of ages (3-83 years). This combined study meets all requirements of the Association of Advancement of Medical Instrumentation and achieved a grade 'A' in the BHS protocol.
Authors:
Gurdial S Mattu; Balraj S Heran; James M Wright
Related Documents :
11007598 - Description and validation of a novel liquid metal-free device for venous congestion pl...
2597558 - Development of an intravascular lung assist device.
12055408 - Task force i: methodological aspects of blood pressure measurement.
4023898 - An electronic technique for compartment pressure measurement using the wick catheter.
11591618 - Vasoconstrictor effect of the angiotensin-converting enzyme-resistant, chymase-specific...
17421018 - Feasibility and clinical impact of transcatheter closure of interatrial communications ...
Publication Detail:
Type:  Comparative Study; Evaluation Studies; Journal Article    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  9     ISSN:  1359-5237     ISO Abbreviation:  Blood Press Monit     Publication Date:  2004 Feb 
Date Detail:
Created Date:  2004-03-15     Completed Date:  2004-10-19     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  47-52     Citation Subset:  IM    
Affiliation:
Department of Pharmacology & Therapeutics, University of British Columbia, Vancouver, BC, Canada.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Age Factors
Automation
Blood Pressure
Blood Pressure Determination / instrumentation*,  standards
Blood Pressure Monitors / standards*
Child
Child, Preschool
Humans
Reproducibility of Results
Sphygmomanometers / standards

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Comparison of the automated non-invasive oscillometric blood pressure monitor (BpTRU) with the auscu...
Next Document:  Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH I...