Optimizing the AV delay in DDD pacemaker patients with high degree AV block: mitral valve Doppler versus impedance cardiography.
MedLine Citation:
PMID:  9358487     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
In DDD-pacemaker patients with high degree AV block, Doppler echocardiography of transmitral blood flow can be used to find the individually optimal AV delay (AVO) for left heart AV synchronization. This study tried to validate a Doppler method (ECHO) recently proposed to optimize left ventricular filling by comparing it to stroke volume data derived from impedance cardiography (ICG). It should be further elucidated if optimizing the AV delay (AVD) by means of this method is superior to fixed AVD settings and which differential AVD (pace-sense-offset) should be programmed for atrially triggered (ATP) and AV sequential (AVP) pacing, respectively. AVO as measured in 53 patients showed a linear correlation between ECHO and ICG for both ATP (r = 0.66, P < 0.00001) and AVP (r = 0.53; P < 0.005). The mean deviation in AVO between ECHO and ICG was +/- 26 ms (ATP) and +/- 30 ms (AVP), respectively, with a tendency to longer AVDs with the Doppler method. ECHO limitations could mainly be attributed to: (1) restrictions of AVD programming options (which may be compensated for by slight modification of the proposal); and (2) to pathophysiological mechanisms that alter mitral valve dynamics. Optimization of the AVD by Doppler produced a stroke volume that was significantly higher (19%) than with a fixed AVD (150 ms in ATP; 200 ms in AVP). There was a wide scatter in pace-sense-offsets between-7 and 134 ms, which was reflected by both methods. It is concluded that AVO determinations by ECHO are valid provided that methodological pitfalls and limitations caused by the disease are recognized. Tailoring AVD with respect to diastolic filling improves systolic function and is superior to nominal AVD settings. Fixed differential AVDs as offered by some manufacturers are far from being physiological. Thus modern pulse generators should offer free programmability over a wide range of AV delays.
Authors:
M Kindermann; G Fröhlig; T Doerr; H Schieffer
Related Documents :
21622637 - Pigmented villonodular synovitis of the ankle: radiologic characteristics.
8517387 - Cine magnetic resonance imaging and color doppler flow mapping in infants and children ...
19909927 - Quantification of functional mitral regurgitation by real-time 3d echocardiography: com...
17490237 - Pr01 "a picture speaks a thousand words"- the use of digital photography and the intern...
11719677 - Patient's assessment of discomfort during mr arthrography of the shoulder.
9430827 - Imaging the long head of biceps tendon--a pictorial essay emphasizing magnetic resonance.
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  20     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  1997 Oct 
Date Detail:
Created Date:  1997-12-19     Completed Date:  1997-12-19     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  2453-62     Citation Subset:  IM    
Affiliation:
Universitätskliniken des Saarlandes, Homburg/Saar, Germany.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Atrioventricular Node / physiopathology*
Blood Flow Velocity
Cardiography, Impedance*
Echocardiography, Doppler*
Electrocardiography
Female
Heart Block / physiopathology,  therapy*
Humans
Male
Middle Aged
Mitral Valve / physiopathology*,  ultrasonography
Pacemaker, Artificial*
Stroke Volume
Ventricular Function, Left

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A comparison of transvenous atrial defibrillation of acute and chronic atrial fibrillation and the e...
Next Document:  Effect of ventricular pacing on coronary blood flow in patients with normal coronary arteries.